25 July 2022 Detection of hypoxia by near-infrared spectroscopy and pulse oximetry: a comparative study
Amanda Cheung, Lorna Tu, Andrew Macnab, Brian K. Kwon, Babak Shadgan
Author Affiliations +
Abstract

Significance: Pulse oximetry is widely used in clinical practice to monitor changes in arterial oxygen saturation (SpO2). However, decreases in SpO2 can be delayed relative to the actual clinical event, and near-infrared spectroscopy (NIRS) may detect alterations in oxygenation earlier than pulse oximetry, as shown in previous cerebral oxygenation monitoring studies.

Aim: We aim to compare the response of transcutaneous muscle NIRS measures of the tissue saturation index with pulse oximetry SpO2 during hypoxia.

Approach: Episodes of acute hypoxia were induced in nine anesthetized Yucatan miniature pigs. A standard pulse oximeter was attached to the ear of the animal, and a transcutaneous NIRS sensor was placed on the hind limb muscle. Hypoxia was induced by detaching the ventilator from the animal and reattaching it once the pulse oximeter reported 70% SpO2.

Results: Twenty-four episodes of acute hypoxia were analyzed. Upon the start of hypoxia, the transcutaneous NIRS measures changed in 5.3 ± 0.4 s, whereas the pulse oximetry measures changed in 14.9 ± 1.0 s (p < 0.0001).

Conclusions: Transcutaneous muscle NIRS can detect the effects of hypoxia significantly sooner than pulse oximetry in the Yucatan miniature pig. A transcutaneous NIRS sensor may be used as an earlier detector of oxygen saturation changes in the clinical setting than the standard pulse oximeter.

1.

Introduction

Pulse oximetry is a simple, portable, and noninvasive tool widely used in clinical practice to evaluate the oxygenation status of a patient by monitoring arterial oxygen saturation (SpO2).1,2 Each pulse oximeter unit consists of a monitor and a sensor. The monitor component controls light signal generation. The sensor component consists of one light-emitting diode (LED) that uses two red (600 to 750 nm) and near-infrared (NIR) (850 to 1000 nm) wavelengths to penetrate the skin and tissue, a photodetector (PD) to detect the light absorption, and a photon-counting processor.3 The optical technology is based on photoplethysmography (PPG), which is used to measure blood volume changes in the microvascular tissue bed. PPG detects the increase in light absorption during the systolic increase in arterial blood volume.4 Pulse oximeters measure the difference in light absorption of oxygenated (O2Hb) and deoxygenated hemoglobin (HHb) in pulsatile arterial blood to estimate the percentage of oxygen saturation in the arterial circulation.57

In situations in which a patient has poor peripheral perfusion, a pulse oximeter can be less reliable due to the pulse wave being inadequate for measurement.7 Similarly, pulse oximeters may lose their accuracy in patients with cardiac arrhythmia and will not measure at all following cardiac arrest when peripheral pulses are absent.8 Hypothermia, hypotension, or cold extremities may also contribute to the failure of an oximeter to register a signal.9 Movement artifacts are another factor that can affect the reading of SpO2.10,11 Abnormal hemoglobin levels as a result of the presence of carboxyhemoglobin or methemoglobin have also resulted in errors in SpO2 readings in both animal studies and clinical case reports.1215 Other factors can also affect the reliability of pulse oximetry, such as fluorescent lighting and nail varnish.1618 In addition, SpO2 is an estimate of arterial oxygen saturation (SaO2) derived by arterial blood gas analysis with measures of SpO2 demonstrating a typical difference of 2% from SaO2.1921 These differences are reported in subjects with SaO2 levels over 80%, so it is important to consider that the performance of pulse oximeters is less reliable when SaO2 is <80%.8,20

Several studies have reported that decreases in SpO2 detected by pulse oximetry may lag behind the actual clinical event.6,7,22,23 We have also observed, when using different pulse oximeter systems in past animal experiments, that the onset of changes in SpO2 in response to acutely induced hypoxia can occur after significant delays. Being able to detect a critical cardiorespiratory event earlier can provide additional time for therapeutic interventions able to improve immediate care and outcome. For example, SpO2 is monitored in sedative dentistry to detect potential airway obstruction, and identifying hypoxia even seconds earlier would be vital and critical to restoring airway patency.24

Previous studies have demonstrated that cerebral oxygenation monitoring using near-infrared spectroscopy (NIRS) can detect alterations in oxygenation earlier than pulse oximetry.2529 NIRS is a noninvasive optical technology used to measure tissue oxygenation and hemodynamics in various tissues, such as the brain, skeletal muscle, and spinal cord.3033 Using similar physics principles to pulse oximetry, NIRS generates NIR light signals that penetrate tissue and measure the amount of light absorbed by tissue chromophores.34 With the use of a single light source with at least two wavelengths in the NIR spectrum (650 to 1000 nm) and one PD, NIRS is able to monitor changes in tissue oxygen delivery and consumption by measuring O2Hb, HHb, and the oxygenated–deoxygenated hemoglobin difference (Hbdiff=O2HbHHb), a relative measure of tissue oxygenation.3538 In a multi-distance spatially resolved (SR) configuration (one PD with multiple light sources) or a single-distance configuration (one PD with one light source emitting multiple wavelengths), the light sources and PDs are adjacent to each other, which allows for calculation of the differential pathlength factor, the average distance a photon travels between the source and detector through tissue.39 This factor is incorporated into an algorithm derived from the modified Beer–Lambert law, which allows for an absolute measure of tissue oxygenation to be derived, providing relative changes in tissue oxygenation expressed as the tissue saturation index (TSI).3941

The NIRS measures of tissue oxygen saturation are derived from a combination of blood in the arteries (25%), capillaries (5%), and veins (70%).42,43 The main difference between pulse oximetry and NIRS in terms of the calculations of oxygen saturation is the tissue being sampled. Pulse oximetry calculates the percentage of O2Hb in arterial blood, whereas NIRS calculates the percentage of O2Hb in both arterial and venous blood compartments.21

NIRS oxygenation monitoring is an already established approach in the management of traumatic brain injury to minimize hypoxic and ischemic brain damage.44 Several studies have reported reduced mortality and improved clinical outcomes from clinical decisions and diagnoses driven by NIRS monitoring.4547 Patients benefiting from this approach include children on cardiopulmonary bypass,48,49 older adults during abdominal and orthopedic surgery who showed reduced postoperative cognitive impairment,50 and patients undergoing coronary artery bypass grafting with significantly shorter lengths of stay in the intensive care unit and postoperative hospitalization.51 Cerebral ischemia caused by technical issues with cannulation or cardiopulmonary bypass can also be detected in real time via NIRS monitoring.50,52 These studies emphasize the benefit and importance of NIRS monitoring as a means of maintaining an adequate oxygen supply or rapidly restoring interrupted oxygen delivery. As NIRS does not require pulsatile blood flow, it can be a valuable tool in situations in which pulse oximetry has limited applicability, e.g., in periods of low blood flow occurring in hypotensive patients, during cardiopulmonary bypass, and when the circulation is nonpulsatile, such as cardiac arrest. To our knowledge, there are no reports to date comparing transcutaneous NIRS monitoring to pulse oximetry during acute hypoxia, as previous comparative studies focused on cerebral NIRS measurements.2529 Transcutaneous NIRS monitoring of tissue hypoxia provides a better alternative to cerebral NIRS monitoring as it uses a simple and small sensor that can be easily placed on any part of the body over the skin. Cerebral NIRS systems require a large sensor with a 3- to 4-cm interoptode distance (IOD) because a higher depth of penetration is needed to access the brain cortex, whereas a transcutaneous NIRS system can use a smaller sensor with a 1-cm IOD. It would be highly advantageous when limited locations are available for monitoring.

This study aimed to compare transcutaneous muscle NIRS and pulse oximetry in detecting the onset of hypoxia. Specifically, we compared the time of onset between changes in hind limb muscle oxygenation measured with NIRS to changes in arterial oxygen saturation measured with pulse oximetry in anesthetized Yucatan miniature pigs during induced acute hypoxia. The NIRS sensor was placed on the thigh muscle of the animal, a standard and common region for NIRS measurements. At the same time, the pulse oximeter was attached to the ear of the animal, a standard placement for pulse oximetry. Because the application of pulse oximetry on the hind limb is not standard practice and we could not place our NIRS sensor on the ear, we did not compare NIRS and pulse oximeter readings from the same region. Still, we compared two standard and applicable regions between the two different techniques.

2.

Material and Methods

All animal protocols and procedures performed in this study were approved by the Animal Care Committee of the University of British Columbia (UBC) and were compliant with the policies of the Canadian Council of Animal Care, the U.S. Army Medical Research and Materiel Command (USAMRMC), and the Animal Care and Use Review Office (ACURO). The anesthesia/analgesia protocols were established by the UBC Center for Comparative Medicine.

2.1.

Porcine Model and Transcutaneous NIRS Sensor

Nine female Yucatan minipigs weighing 25 to 31 kg from previous studies that employed NIRS for investigating spinal cord hemodynamics after spinal cord injury (SCI) were analyzed.33 Animals were prepared for surgery, intubated, and anesthetized as previously described.53,54 In brief, animals were anesthetized with intramuscular telazol (5 mg/kg) and xylazine (1 mg/kg) and then endotracheally intubated. Propofol (2 mg/kg), fentanyl (10  μg/kg), and ketamine (11 mg/kg) were used for anesthesia induction, and propofol (8 mg/kg/h), fentanyl (12  μg/kg/h), and ketamine (11 mg/kg/h) were used for anesthesia maintenance through a continuous rate infusion. Animals were mechanically ventilated with a ventilator rate of 12 breaths/min and a tidal volume of 12 to 15 mL/kg with 1.4 L (70%) nitrogen and 0.6 L (30%) of oxygen (Veterinary Anesthesia Ventilator model 2002, Hallowell EMC, Pittsfield, Massachusetts). Standard monitoring of the animals was performed throughout the procedure, including monitoring of the respiratory rate, blood pressure, end-tidal carbon dioxide, heart rate, and oxygen saturation.

A standard pulse oximeter system was attached to the ear of the animal to measure arterial oxygen saturation and heart rate at a frequency of 0.25 Hz (Rad-5 Masimo, California). A transcutaneous NIRS sensor (PortaMon, Artinis Medical Systems, Elst, The Netherlands) was placed on the biceps femoris muscle belly of the hind limb of the animal. The NIRS system uses a continuous-wave multi-distance SR NIRS configuration (Fig. 1) and an NIRS algorithm based on the modified Beer–Lambert law to measure changes in skeletal muscle O2Hb and HHb concentrations and calculate TSI% at a sampling rate of 10 Hz.55 The sensor has three LEDs with two wavelengths (760 and 850 nm) and one photodiode with ambient light protection. The distances between the emitters and the PD are 30, 35, and 40 mm, and the sensor size is 83.8×42.9×17.2  mm.

Fig. 1

A multi-distance SR NIRS sensor.

JBO_27_7_077001_f001.png

The software of the NIRS sensor (Oxysoft, Artinis Medical Systems, Elst, The Netherlands) uses an SR spectroscopy algorithm that measures the slope of the optical density change as a function of multiple distances. This method enables the calculation of scaled absorption coefficient measures, which can allow for measuring values of chromophore ratios and calculating tissue oxygen saturation index in percentage.39

Episodes of acute hypoxia were performed in each animal. Hypoxia was induced by disconnecting the animal from the ventilator and reattaching it once SpO2 reached up to 70% SpO2, as identified by the pulse oximeter. The target of 70% SpO2 was selected to induce a severe hypoxic response related to the spinal cord hemodynamics study that was primarily investigated in animals.33

2.2.

Data Analysis

The first detectable change in the signals was calculated by determining when the value of SpO2 and TSI first changed from their respective baseline values after induction of hypoxia. For the pulse oximeter, the time needed for the SpO2 value to change on the monitor from its baseline value once hypoxia was induced was noted. For the NIRS signal, the data trace was analyzed, and the time point in which the TSI value changed from its baseline value was noted. Comparisons between the time when the first detectable change was seen in the NIRS and pulse oximetry measurements were calculated using the Student’s t-test, with the level of significance set at p<0.05. This comparison was also calculated after downsampling the NIRS data from 10 to 0.25 Hz to compare the NIRS and pulse oximetry data when sampling at the same frequency. Data were analyzed using GraphPad Prism 9.0.0 (GraphPad Software, La Jolla, California). Data are presented as mean ± standard error of the mean.

3.

Results

We analyzed 24 episodes of acute hypoxia in nine female Yucatan miniature pigs. Hypoxia was induced and allowed to proceed with anticipated drops in SpO2 up to 70%.

3.1.

Comparison of SpO2 and TSI Detection Times During Hypoxia

For all 24 hypoxic episodes combined, the first detectable change in NIRS TSI (sampling at 10 Hz) occurred in 5.3±0.4  s, and the first detectable change in SpO2 occurred in 14.9±1.0s (p<0.0001) [Fig. 2(a)]. The NIRS data were then downsampled to 0.25 Hz to compare the NIRS data with the same pulse oximetry data sampling frequency. The first detectable change in NIRS TSI (sampling at 0.25 Hz) occurred in 6.9±0.3  s compared with the first detectable change in SpO2 occurring in 14.9±1.0s (p<0.0001) [Fig. 2(b)]. In all cases, the transcutaneous NIRS sensor detected changes in oxygen saturation earlier than the pulse oximeter.

Fig. 2

Comparison of the detection times for the first detectable change in oxygen saturation measured by NIRS TSI sampling at (a) 10 Hz and (b) 0.25 Hz with pulse oximetry arterial oxygen saturation (SpO2) sampling at 0.25 Hz upon the onset of acute hypoxia.

JBO_27_7_077001_f002.png

3.2.

Changes in TSI Magnitude During Hypoxia

During acute hypoxia, the TSI decreased on average by 10.5±1% from a baseline of 70.6±7.6%. A representative graph showing the onset of changes between SpO2 (sampling at 0.25 Hz) and TSI (sampling at 10 Hz) is shown in Fig. 3. The pulse oximeter and transcutaneous NIRS sensor showed a similar pattern of change during acute hypoxia, depicted by a sharp decrease in their respective oxygenation measurement and followed by an immediate increase as the animal recovered.

Fig. 3

Changes in arterial oxygen saturation (SpO2) and TSI during acute hypoxia in one animal. The dotted black line at A and D represents when the ventilator was (A) detached and (D) reattached. The dotted red line at B represents when the first change in TSI was detected. The dotted blue line at C represents when the first change in SpO2 was detected. Red arrows show respiration cycles detected by the NIRS sensor; note the absence of respiratory cycles during hypoxia induction.

JBO_27_7_077001_f003.png

3.3.

Effect of Ventilation on NIRS Signals

The transcutaneous NIRS sensor recorded regular O2Hb and TSI oscillations with a frequency of 0.2 Hz during the experiment. The rhythm of regular oscillations in all animals that were connected to the ventilator was similar, and the respiratory rate was set at 12 breaths/min. The oscillations disappeared during the hypoxia episodes (Fig. 3).

4.

Discussion

In this study, we demonstrated that NIRS-derived TSI% responds in 5.3±0.4s when sampling at 10 Hz and 6.9±0.3s when sampling at 0.25 Hz (the same sampling frequency as the pulse oximeter) upon the start of acute hypoxia, whereas the pulse oximeter SpO2 responds in 14.9±1.0s. Our results show that transcutaneous NIRS can detect the effects of hypoxia earlier than pulse oximetry and transcutaneous NIRS can detect and monitor respiratory cycles in real time.

Pulse oximetry measurements may lag behind a change in the patient’s oxygenation status and present a delay in detecting hypoxic events. These delays may be a result of irregular pulse volume or rhythm, which can slow down the instrument’s response, or a result of computation delay due to the signal averaging algorithms in the device.56 Individual pulse oximeters have varying specifications and hence different operating characteristics, and the signal averaging required to derive the percentage value displayed always takes several seconds, and in some instruments, it can even exceed 20 s.7,20 Due to such a signal lag, there is a period during which the actual arterial blood oxygen saturation starts to fall in real time before a measurement is completed and displayed by the pulse oximeter. It is important for this limitation to be recognized and, in particular, for clinicians and nurses to be aware of this time delay between the beginning of potentially critical hypoxia and the reflection of the event in its SpO2 reading. Here, NIRS is advantageous as it can monitor respiration and detect hypoxia sooner than pulse oximetry by operating at higher frequencies.

Previous studies centered on cerebral NIRS oxygenation measurements in comparison with pulse oximetry during hypoxic events. In 2007, Tobias reported that hypercyanotic spells during anesthesia resulted in a decrease in cerebral oxygenation [regional oxygen saturation (rSO2)] monitored by NIRS 15 to 30 s prior to any change in pulse oximetry.25 In 2008, Tobias found that a 5% decrease in rSO2 in 10 patients occurred 52 s earlier than a similar decrease in oxygenation measured by pulse oximetry.26 Studies in pediatric patients have shown a delay of SpO2 compared to NIRS rSO2 during hypoxia1012 and that rSO2 decreased about 40 s earlier than SpO2 during hypoxia generated by paused mechanical ventilation.27 In this study, we found that, similar to cerebral oxygenation, muscle oxygenation monitored by NIRS decreased about 10 s before any changes were detected by pulse oximetry upon the start of acute hypoxia. NIRS TSI responded in 5.3±0.4  s whereas the pulse oximeter responded in 14.9±1.0  s. Similarly, with the NIRS and pulse oximeter monitoring at the same frequency (0.25 Hz), NIRS detects changes in oxygenation about 8 s earlier than pulse oximetry, with NIRS TSI% responding in 6.9±0.3  s and the pulse oximeter taking 14.9±1.0  s to show a drop in SpO2.

We recognize limitations in what we report, especially because we are comparing different technologies applied to two different anatomical locations. The ear is a common place for transmission pulse oximetry, whereas the hind limb is a common region for NIRS monitoring. We are unable to monitor SpO2 on the hind limb using conventional pulse oximeters, and applying conventional NIRS sensors such as the Artinis PortaMon used in this study or even the smaller Artinis PortaLite-mini (size: 40×18×5  mm) on the ear lobe is not possible. Therefore, we used two different standard regions to compare the sensitivity of these two technologies that monitor transcutaneous oxygenation changes to detect the onset of acute hypoxia. In addition, we acknowledge that the oxygen saturation we measured in the ear with pulse oximetry was a different location from what we measured in the hind limb with NIRS, but this regional difference did not by itself explain the differences in the measurements that we report in this study. Arterial oxygen saturation is not tissue-dependent as it indicates the amount of oxygen within the arterial stream that supplies each organ and tissue. However, tissue oxygen saturation is a combination of arterial and venous streams at the microvascular level that can be different in different regions based on their local metabolism and hemodynamics. In this context, Tobias suggested that the different hypoxia detection times between NIRS and pulse oximetry are related to the vascular compartments that the technologies are evaluating. During hypoxia, the partial pressure of oxygen decreases at an equal rate in venous and arterial blood; however, given the lower venous partial pressure of oxygen, it decreases more rapidly on the oxyhemoglobin dissociation curve, and a decrease in NIRS tissue oxygenation would occur first.26 Therefore, the difference in detection times is due to the sensitivity to different vascular compartments, in which pulse oximetry monitors arterial oxygen saturation and NIRS reflects changes in arterial, capillary, and venous saturation. We recognize that different sampling frequencies between the NIRS sensor and pulse oximeter may also contribute to the observed time-delay differences. The time difference between 10 and 0.25 Hz sampling rates for the NIRS data was 1.6±0.2  s. In addition, the animals that we used for this study were part of a larger study exploring the local effects of SCI on cord hemodynamics, and the episodes of acute hypoxia were studied both prior to the cord injury occurring and 7 days after the animals had experienced an acute traumatic SCI. There is also the potential limitation that there is no absolute baseline TSI value; hence each patient’s resting value serves as their baseline. However, both sensors monitor transcutaneous oxygenation changes and provide a comparable measurement to allow us to assess whether NIRS is more sensitive than pulse oximetry in detecting oxygenation changes.

A clinical setting where our findings that transcutaneous NIRS data have advantages over pulse oximetry is relevant in sedative dentistry, in which hypoxia is a critical problem.57 With the use of conscious sedation techniques, it is standard to monitor SpO2 to detect potential oral cavity or airway obstruction as this is a clinical situation that can become critical in seconds if its onset is not recognized and patency of the airway is not restored.24 With the ability of NIRS to detect changes in oxygenation earlier than pulse oximetry and monitor patients regardless of blood pressure fluctuation and variations in peripheral perfusion, the advantages of monitoring oxygenation using NIRS would appear to provide additional safety for patients during dental procedures requiring conscious sedation. Furthermore, the ability to monitor the respiratory rate and rhythm in addition to tissue oxygen saturation highlights potential applications of transcutaneous NIRS for monitoring cardiorespiratory function at the bedside.15

Overall, our study provides new evidence of the ability of transcutaneous muscle NIRS to detect the effects of hypoxia earlier than pulse oximetry. Hence, the use of a transcutaneous muscle NIRS sensor in many clinical situations may be a more sensitive and earlier detector of systemic hypoxia than the standard pulse oximeter.

Disclosures

The authors have no relevant financial interests in this manuscript and no potential conflicts of interest to disclose.

Acknowledgments

The authors gratefully acknowledge the technical expertise and contributions of Dr. Behnam Molavi in the design and development of the NIRS prototypes. The authors also gratefully acknowledge the staff at the UBC Center for Comparative Medicine (CCM), who support these complex in vivo experiments and who provide care for the animals involved in the studies. This study was supported by a Translational Research Award from the U.S. Department of Defense, Spinal Cord Injury Research Program (SCIRP), SC150178 and an International Collaboration on Repair Discoveries (ICORD) Seed Grant from the Blusson Integrated Cures Partnership. A. C. is supported by a Vanier Canada Graduate Scholarship from the Canadian Institutes of Health Research. B. K. K. is the Canada Research Chair in Spinal Cord Injury and the Dvorak Chair in Spine Trauma. B. S. holds a scholar award from the Michael Smith Foundation for Health Research.

References

1. 

A. Shah and K. H. Shelley, “Is pulse oximetry an essential tool or just another distraction? The role of the pulse oximeter in modern anesthesia care,” J. Clin. Monit. Comput., 27 235 –242 (2013). https://doi.org/10.1007/s10877-013-9428-7 Google Scholar

2. 

A. F. Merry, “Role of anesthesiologists in WHO safe surgery programs,” Int. Anesthesiol. Clin., 48 (2), 137 –150 (2010). https://doi.org/10.1097/AIA.0b013e3181cd1615 IACLAV 0020-5907 Google Scholar

3. 

W. G. Zijlstra, A. Buursma and W. P. Meeuwsen-van der Roest, “Absorption spectra of human fetal and adult oxyhemoglobin, de-oxyhemoglobin, carboxyhemoglobin, and methemoglobin,” Clin. Chem., 37 (9), 1633 –1638 (1991). https://doi.org/10.1093/clinchem/37.9.1633 Google Scholar

4. 

I. Yoshiya, Y. Shimada and K. Tanaka, “Spectrophotometric monitoring of arterial oxygen saturation in the fingertip,” Med. Biol. Eng. Comput., 18 27 –32 (1980). https://doi.org/10.1007/BF02442476 MBECDY 0140-0118 Google Scholar

5. 

M. Nitzan, A. Romem and R. Koppel, “Pulse oximetry: fundamentals and technology update,” Med. Devices Evid. Res., 7 231 –239 (2014). https://doi.org/10.2147/MDER.S47319 Google Scholar

6. 

J. Allen, “Photoplethysmography and its application in clinical physiological measurement,” Physiol. Meas., 28 R1 –R39 (2007). https://doi.org/10.1088/0967-3334/28/3/R01 PMEAE3 0967-3334 Google Scholar

7. 

S. DeMeulenaere, “Pulse oximetry: uses and limitations,” J. Nurse Pract., 3 (5), 312 –317 (2007). https://doi.org/10.1016/j.nurpra.2007.02.021 JANPEB 1041-2972 Google Scholar

8. 

B. G. Nickerson, C. Sarkisian and K. Tremper, “Bias and precision of pulse oximeters and arterial oximeters,” Chest, 93 (3), 515 –517 (1988). https://doi.org/10.1378/chest.93.3.515 CHETBF 0012-3692 Google Scholar

9. 

A. J. Macnab, G. Baker-Brown and E. E. Anderson, “Oximetry in children recovering from deep hypothermia for cardiac surgery,” Crit. Care Med., 18 (10), 1066 –1069 (1990). https://doi.org/10.1097/00003246-199010000-00003 CCMDC7 0090-3493 Google Scholar

10. 

C. F. Poets and D. P. Southall, “Noninvasive monitoring of oxygenation in infants and children: practical considerations and areas of concern,” Pediatrics, 93 (5), 737 –746 (1994). https://doi.org/10.1542/peds.93.5.737 PEDIAU 0031-4005 Google Scholar

11. 

C. F. Poets and V. A. Stebbens, “Detection of movement artifact in recorded pulse oximeter saturation,” Eur. J. Pediatr., 156 808 –811 (1997). https://doi.org/10.1007/s004310050719 Google Scholar

12. 

S. J. Barker and K. K. Tremper, “The effect of carbon monoxide inhalation on pulse oximetry and transcutaneous PO2,” Anesthesiology, 66 (5), 677 –679 (1987). https://doi.org/10.1097/00000542-198705000-00014 ANESAV 0003-3022 Google Scholar

13. 

S. J. Barker, K. K. Tremper and J. Hyatt, “Effects of methemoglobinemia on pulse oximetry and mixed venous oximetry,” Anesthesiology, 70 (1), 112 –117 (1989). https://doi.org/10.1097/00000542-198901000-00021 ANESAV 0003-3022 Google Scholar

14. 

M. F. Watcha, M. T. Connor and A. V. Hing, “Pulse oximetry in methemoglobinemia,” Am. J. Dis. Child., 143 (7), 845 –847 (1989). https://doi.org/10.1001/archpedi.1989.02150190095030 AJDCAI 0002-922X Google Scholar

15. 

N. B. Hampson, “Pulse oximetry in severe carbon monoxide poisoning,” Chest, 114 (4), 1036 –1041 (1998). https://doi.org/10.1378/chest.114.4.1036 CHETBF 0012-3692 Google Scholar

16. 

M. M. ChanM. M. Chan and E. D. Chan, “What is the effect of fingernail polish on pulse oximetry?,” Chest, 123 (6), 2163 –2164 (2003). https://doi.org/10.1378/chest.123.6.2163 CHETBF 0012-3692 Google Scholar

17. 

C. J. Cote et al., “The effect of nail polish on pulse oximetry,” Anesth. Analg., 67 (7), 683 –686 (1988). https://doi.org/10.1213/00000539-198807000-00013 Google Scholar

18. 

A. S. Rubin, “Nail polish color can affect pulse oximeter saturation,” Anesthesiology, 68 825 (1988). https://doi.org/10.1097/00000542-198805000-00049 ANESAV 0003-3022 Google Scholar

19. 

J. E. Sinex, “Pulse oximetry: principles and limitations,” Am. J. Emerg. Med., 17 (1), 59 –66 (1999). https://doi.org/10.1016/S0735-6757(99)90019-0 Google Scholar

20. 

S. Fouzas, K. N. Priftis and M. B. Anthracopoulos, “Pulse oximetry in pediatric practice,” Pediatrics, 128 (4), 740 –752 (2011). https://doi.org/10.1542/peds.2011-0271 PEDIAU 0031-4005 Google Scholar

21. 

A. Jubran, “Pulse oximetry,” Intensive Care Med., 30 (11), 2017 –2020 (2004). https://doi.org/10.1007/s00134-004-2399-x ICMED9 0342-4642 Google Scholar

22. 

N. S. Trivedi et al., “Effects of motion, ambient light, and hypoperfusion on pulse oximeter function,” J. Clin. Anesth., 9 (3), 179 –183 (1997). https://doi.org/10.1016/S0952-8180(97)00039-1 JCLBE7 Google Scholar

23. 

A. Cheung et al., “Detecting oxygenation changes after hypoxia: pulse oximetry vs. near-infrared spectroscopy,” Proc. SPIE, 11638 116380K (2021). https://doi.org/10.1117/12.2583306 PSISDG 0277-786X Google Scholar

24. 

S. Wilson, “Conscious sedation and pulse oximetry: false alarms?,” Pediatr. Dent., 12 (4), 228 –232 (1990). PEDEDL Google Scholar

25. 

J. D. Tobias, “Cerebral oximetry monitoring provides early warning of hypercyanotic spells in an infant with tetralogy of Fallot,” J. Intensive Care Med., 22 (2), 118 –120 (2007). https://doi.org/10.1177/0885066606297966 Google Scholar

26. 

J. D. Tobias, “Cerebral oximetry monitoring with near infrared spectroscopy detects alterations in oxygenation before pulse oximetry,” J. Intensive Care Med., 23 (6), 384 –388 (2008). https://doi.org/10.1177/0885066608324380 Google Scholar

27. 

Y. Lu et al., “Comparing the response of pulse oximetry and regional cerebral oxygen saturation to hypoxia in preschool children,” Exp. Ther. Med., 19 (1), 353 –358 (2020). https://doi.org/10.3892/etm.2019.8199 Google Scholar

28. 

P. E. Séguéla et al., “Ductal closure and near-infrared spectroscopy for regional oxygenation monitoring in ductus-dependent congenital heart disease,” Arch. Pediatr., 22 (8), 857 –860 (2015). https://doi.org/10.1016/j.arcped.2015.04.020 Google Scholar

29. 

I. C. Tanidir et al., “Near infrared spectroscopy monitoring in the pediatric cardiac catheterization laboratory,” Artif. Organs, 38 (10), 838 –844 (2014). https://doi.org/10.1111/aor.12256 ARORD7 0160-564X Google Scholar

30. 

A. J. Macnab et al., “NIRS monitoring of brain and spinal cord—detection of adverse intraoperative events,” J. Spectrosc., 17 (2–3), 483 –490 (2003). https://doi.org/10.1155/2003/607654 Google Scholar

31. 

T. Hamaoka et al., “The use of muscle near-infrared spectroscopy in sport, health and medical sciences: recent developments,” Philos. Trans. R. Soc. A Math. Phys. Eng. Sci., 369 (1955), 4591 –4604 (2011). https://doi.org/10.1098/rsta.2011.0298 Google Scholar

32. 

B. Shadgan et al., “Optical assessment of spinal cord tissue oxygenation using a miniaturized near infrared spectroscopy sensor,” J. Neurotrauma, 36 (21), 3034 –3043 (2019). https://doi.org/10.1089/neu.2018.6208 JNEUE4 0897-7151 Google Scholar

33. 

A. Cheung et al., “Continuous optical monitoring of spinal cord oxygenation and hemodynamics during the first seven days post-injury in a porcine model of acute spinal cord injury,” J. Neurotrauma, 37 (21), 2292 –2301 (2020). https://doi.org/10.1089/neu.2020.7086 JNEUE4 0897-7151 Google Scholar

34. 

F. Jobsis, “Noninvasive, infrared monitoring of cerebral and myocardial oxygen sufficiency and circulatory parameters,” Science, 198 (4323), 1264 –1267 (1977). https://doi.org/10.1126/science.929199 Google Scholar

35. 

S. J. Matcher et al., “Performance comparison of several published tissue near-infrared spectroscopy algorithms,” Anal. Biochem., 227 54 –68 (1995). https://doi.org/10.1006/abio.1995.1252 ANBCA2 0003-2697 Google Scholar

36. 

A. Ferraris, M. Jacquet-Lagrèze and J. L. Fellahi, “Four-wavelength near-infrared peripheral oximetry in cardiac surgery patients: a comparison between EQUANOX and O3,” J. Clin. Monit. Comput., 32 253 –259 (2018). https://doi.org/10.1007/s10877-017-0025-z Google Scholar

37. 

D. Wyser et al., “Wearable and modular functional near-infrared spectroscopy instrument with multi-distance measurements at four wavelengths,” Neurophotonics, 4 041413 (2017). https://doi.org/10.1117/1.NPh.4.4.041413 Google Scholar

38. 

I. Tachtsidis et al., “Relationship between brain tissue haemodynamics, oxygenation and metabolism in the healthy human adult brain during hyperoxia and hypercapnea,” Adv. Exp. Med. Biol., 645 315 –320 (2009). https://doi.org/10.1007/978-0-387-85998-9_47 AEMBAP 0065-2598 Google Scholar

39. 

S. Suzuki et al., “Tissue oxygenation monitor using NIR spatially resolved spectroscopy,” Proc. SPIE, 3597 582 –592 (1999). https://doi.org/10.1117/12.356862 PSISDG 0277-786X Google Scholar

40. 

I. Hagino et al., “Tissue oxygenation index is a useful monitor of histologic and neurologic outcome after cardiopulmonary bypass in piglets,” J. Thorac. Cardiovasc. Surg., 130 384 –392 (2005). https://doi.org/10.1016/j.jtcvs.2005.02.058 JTCSAQ 0022-5223 Google Scholar

41. 

Z. Kovacsova et al., “Investigation of confounding factors in measuring tissue saturation with NIRS spatially resolved spectroscopy,” Adv. Exp. Med. Biol., 1072 307 –312 (2018). https://doi.org/10.1007/978-3-319-91287-5_49 AEMBAP 0065-2598 Google Scholar

42. 

H. M. Watzman et al., “Arterial and venous contributions to near-infrared cerebral oximetry,” Anesthesiology, 93 947 –953 (2000). https://doi.org/10.1097/00000542-200010000-00012 ANESAV 0003-3022 Google Scholar

43. 

P. Lemmers, G. Naulaers, F. van Bell, “Clinical applications of near-infrared spectroscopy in neonates,” Hemodynamics and Cardiology: Neonatology Questions and Controversies, 173 –183 2nd ed.Elsevier(2012). Google Scholar

44. 

S. Bratton et al., “Guidelines for the management of severe traumatic brain injury. X. Brain oxygen monitoring and thresholds,” J. Neurotrauma, 24 ((Suppl. 1), S65 –S70 (2007). https://doi.org/10.1089/neu.2007.9986 JNEUE4 0897-7151 Google Scholar

45. 

K. L. Kiening et al., “Monitoring of cerebral oxygenation in patients with severe head injuries: brain tissue PO2 versus jugular vein oxygen saturation,” J. Neurosurg., 85 751 –757 (1996). https://doi.org/10.3171/jns.1996.85.5.0751 JONSAC 0022-3085 Google Scholar

46. 

J. L. Pascual et al., “Reduced brain tissue oxygen in traumatic brain injury: are most commonly used interventions successful?,” J. Trauma - Inj. Infect. Crit. Care, 70 535 –546 (2011). https://doi.org/10.1097/TA.0b013e31820b59de Google Scholar

47. 

M. F. Stiefel et al., “Reduced mortality rate in patients with severe traumatic brain injury treated with brain tissue oxygen monitoring,” J. Neurosurg., 103 805 –811 (2005). https://doi.org/10.3171/jns.2005.103.5.0805 JONSAC 0022-3085 Google Scholar

48. 

C. D. Kurth, J. M. Steven and S. C. Nicolson, “Cerebral oxygenation during pediatric cardiac surgery using deep hypothermic circulatory arrest,” Anesthesiology, 82 (1), 74 –82 (1995). https://doi.org/10.1097/00000542-199501000-00011 ANESAV 0003-3022 Google Scholar

49. 

E. H. Austin et al., “Benefit of neurophysiologic monitoring for pediatric cardiac surgery,” J. Thorac. Cardiovasc. Surg., 114 (5), 707 –717 (1997). https://doi.org/10.1016/S0022-5223(97)70074-6 JTCSAQ 0022-5223 Google Scholar

50. 

R. J. Ing et al., “Detection of unintentional partial superior vena cava occlusion during a bidirectional cavopulmonary anastomosis,” J. Cardiothorac. Vasc. Anesth., 18 (4), 472 –474 (2004). https://doi.org/10.1053/j.jvca.2004.05.006 JCVAEK 1053-0770 Google Scholar

51. 

J. M. Murkin et al., “Monitoring brain oxygen saturation during coronary bypass surgery: a randomized, prospective study,” Anesth. Analg., 104 (1), 51 –58 (2007). https://doi.org/10.1213/01.ane.0000246814.29362.f4 Google Scholar

52. 

T. Sakamoto et al., “Cerebral ischemia caused by obstructed superior vena cava cannula is detected by near-infrared spectroscopy,” J. Cardiothorac. Vasc. Anesth., 18 (3), 293 –303 (2004). https://doi.org/10.1053/j.jvca.2004.03.008 JCVAEK 1053-0770 Google Scholar

53. 

F. Streijger et al., “Changes in pressure, hemodynamics, and metabolism within the spinal cord during the first 7 days after injury using a porcine model,” J. Neurotrauma, 34 (24), 3336 –3350 (2017). https://doi.org/10.1089/neu.2017.5034 JNEUE4 0897-7151 Google Scholar

54. 

K. T. Kim et al., “Differences in morphometric measures of the uninjured porcine spinal cord and dural sac predict histological and behavioral outcomes after traumatic spinal cord injury,” J. Neurotrauma, 36 (21), 3005 –3017 (2019). https://doi.org/10.1089/neu.2018.5930 JNEUE4 0897-7151 Google Scholar

55. 

P. V. Zephaniah and J. G. Kim, “Recent functional near infrared spectroscopy based brain computer interface systems: developments, applications and challenges,” Biomed. Eng. Lett., 4 (3), 223 –230 (2014). https://doi.org/10.1007/s13534-014-0156-9 Google Scholar

56. 

J. T. B. Moyle, “Uses and abuses of pulse oximetry,” Arch. Dis. Child., 74 77 –80 (1996). https://doi.org/10.1136/adc.74.1.77 ADCHAK 0003-9888 Google Scholar

57. 

S. R. Gandy, “The use of pulse oximetry in dentistry,” J. Am. Dent. Assoc., 126 (9), 1274 –1278 (1995). https://doi.org/10.14219/jada.archive.1995.0363 Google Scholar

Biography

Amanda Cheung received her PhD in neuroscience at the University of British Columbia in 2022, where she was a Vanier Scholar and Killam Doctoral Scholar. She received her BSc degree (Hons) in Neuroscience and Physiology from the University of Toronto in 2017. She current research interests include near-infrared spectroscopy, spinal cord injury, and hemodynamic management. She has been a member of SPIE since 2018.

Lorna Tu is a PhD student in medical physics at the University of British Columbia, where she received her BSc degree (Hons) in biophysics in 2020. Her current research interests include medical imaging, radiomics, and machine learning. Lorna received an SPIE Scholarship in 2020.

Andrew Macnab is a clinician, distinguished scholar, and award-winning researcher and global health project director. He has a reputation for innovation and excellence and has led the application of several new technologies in medicine, including pioneering the use of near-infrared spectroscopy in urology and spinal cord research. He has > 50 years of experience developing solutions to problems in clinical care; several of his initiatives provided the foundation for everyday healthcare technologies now relied on worldwide.

Brian K. Kwon is a surgeon-scientist with advanced training in spine surgery and also a PhD in neuroscience. He is a professor in the Department of Orthopaedics at the University of British Columbia. As a neuroscientist and principal investigator at ICORD, he runs an active basic/translational research laboratory. As an attending orthopedic spine surgeon at Vancouver General Hospital, his practice is focused on the management of adult spine trauma and spinal cord injuries.

Babak Shadgan is a medical doctor specializing in Sports Medicine with a PhD in clinical biophotonics. He is an MSFHR Scholar, an assistant professor in the Department of Orthopaedics at the University of British Columbia (UBC), an associate faculty member at the UBC School of Biomedical Engineering, and a principal investigator at the International Collaboration on Repair Discoveries (ICORD), where he is directing the Implantable Biosensing Laboratory. He is an SPIE Fellow Member.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Amanda Cheung, Lorna Tu, Andrew Macnab, Brian K. Kwon, and Babak Shadgan "Detection of hypoxia by near-infrared spectroscopy and pulse oximetry: a comparative study," Journal of Biomedical Optics 27(7), 077001 (25 July 2022). https://doi.org/10.1117/1.JBO.27.7.077001