JAPANESE CIRCULATION JOURNAL
Online ISSN : 1347-4839
Print ISSN : 0047-1828
ISSN-L : 0047-1828
Clinical Study
An Enhanced Method for Measuring Cardiac Output Using Doppler Color Flow Echocardiography
Hiromi SeoMasakazu YamagishiSyed Azilzul HaqueA.K.M. MohibullahAbdullah Al-Shafi MajumderSatoshi NakataniOi Ling KwanAnthony N. DeMariaKunio Miyatake
Author information
JOURNAL FREE ACCESS

1997 Volume 61 Issue 11 Pages 905-911

Details
Abstract

An enhanced method for determining cardiac output using Doppler color flow imaging techniques to measure mitral orifice diameter was developed and validated in an experimental model and in clinical patients. In an in vitro circuit model, color jet width correlated well with actual orifice dimension from 12 to 24 mm (r=0.99). In the clinical application, mitral valve area was calculated as a ×b×π/4 where a and b represent the width of the color flow stream in the mitral orifice just distal to the annulus in apical long-axis (short-diameter) and 4-chamber (90° rotated, Iong-diameter) views, respectively. Cardiac output was then computed as the product of mitral valve area and time-velocity integral of transmitral flow from the same site. Cardiac output was also measured by thermodilution and conventional echocardiographic methods using diameters and time-velocity integrals from the left ventricular outflow tract. In 30 patients with nonvalvular heart disease, cardiac output measured by thermodilution ranged from 3.40 to 8.40 L/min. Cardiac output was determined in 28 of 30 patients (93%) by the Doppler color flow imaging technique; it ranged from 3.00 to 8.36 L/min and correlated well with thermodilution: y=0.90x+0.63, r=0.91. Cardiac output was determined in 24 of 30 patients by the conventional left ventricular outflow method (80%). The cardiac output measured by the conventional method correlated less closely with thermodilution (r=0.84), although there was no statistical difference in correlation coefficiencies between the 2 methods. These results indicate that the Doppler color flow imaging technique can be used to enhance the determination of cardiac output by echocardiography, particularly when the conventioal method has resulted in technically inadequate recordings. (Jpn Circ J 1997; 61: 905 - 911)

Content from these authors
© 1997 THE JAPANESE CIRCULATION SOCIETY
Previous article Next article
feedback
Top