Numerical Analysis of Hydrodynamic Flow in Microfluidic Biochip for Single-Cell Trapping Application

1
Department of Control and Mechatronic Engineering, Faculty of Electrical Engineering, Universiti Teknologi Malaysia, Skudai, Johor 81310, Malaysia
2
Institute of Ibnu Sina, Universiti Teknologi Malaysia, Skudai, Johor 81310, Malaysia
*
Author to whom correspondence should be addressed.
Int. J. Mol. Sci. 2015, 16(11), 26770-26785; https://0-doi-org.brum.beds.ac.uk/10.3390/ijms161125987
Submission received: 30 May 2015 / Revised: 29 July 2015 / Accepted: 5 August 2015 / Published: 9 November 2015
(This article belongs to the Special Issue Single Cell Analysis in Biotechnology and Systems Biology)

Abstract

:
Single-cell analysis has become the interest of a wide range of biological and biomedical engineering research. It could provide precise information on individual cells, leading to important knowledge regarding human diseases. To perform single-cell analysis, it is crucial to isolate the individual cells before further manipulation is carried out. Recently, microfluidic biochips have been widely used for cell trapping and single cell analysis, such as mechanical and electrical detection. This work focuses on developing a finite element simulation model of single-cell trapping system for any types of cells or particles based on the hydrodynamic flow resistance (Rh) manipulations in the main channel and trap channel to achieve successful trapping. Analysis is carried out using finite element ABAQUS-FEA™ software. A guideline to design and optimize single-cell trapping model is proposed and the example of a thorough optimization analysis is carried out using a yeast cell model. The results show the finite element model is able to trap a single cell inside the fluidic environment. Fluid’s velocity profile and streamline plots for successful and unsuccessful single yeast cell trapping are presented according to the hydrodynamic concept. The single-cell trapping model can be a significant important guideline in designing a new chip for biomedical applications.

Graphical Abstract

1. Introduction

Lab on a Chip (LOC) and Micro Total Analysis Systems (μTAS) have attracted researchers’ attention in the areas of biotechnology and biomedical engineering. The rise in interest is due to the utilization of these devices in a broad range of biological and biomedical application areas including genomics, enzymatic analysis, disease diagnosis, cell treatment, drug screening, single-cell analysis, and drug delivery. In cellular biology, single-cell analysis refers to the study of individual cells isolated from tissues in multi-cellular organisms. Conventionally, cell analyses are conducted with large populations of cells and data measurement can only represent the average values summed over the responses of many cells. Therefore, single-cell analysis is important to obtain more precise information and to reveal the properties of individual cells and cell-to-cell differences [1].
In order to perform single-cell analysis in microfluidic devices, trapping of a single cell is necessary. A variety of techniques have been employed to trap an individual cell. For example, microwell-based [2,3,4,5,6], dielectrophoresis-based [7,8,9,10,11], and hydrodynamic-based [12,13,14,15,16,17,18,19,20,21,22,23,24] microfluidic devices for single-cell trapping have been developed in response to an increasing demand for simple yet reliable tools for high-throughput cell manipulation at the single cell level. In microwell-based platforms, a precise geometry design is required to achieve a high trapping efficiency [4]. Dielectrophoresis-based cell trapping applied a non-uniform AC field to manipulate polarized particles in suspension and is an effective technique to efficiently manipulate a single cell. However, it appears to damage the trapped cells, thus affecting the cell proliferation. Hydrodynamic trapping uses the altered fluidic resistance created by microstructures on a fluid path, such as sieve-like traps [23,24,25] or small trapping sites [12,13,14,15,16,17,26,27], to control the movement of cells in a microchannel. For straight or serpentine-shaped channels with trapping sites, the fluidic resistances of these channels are carefully calculated so that the fluid and cells in the main channel will preferentially flow into the trapping sites when they are empty, but bypass them when they are occupied with a cell. The main challenge in hydrodynamic trapping is that it requires a precise microfluidic control of multiple streams. Further investigation and optimization of cells’ trapping efficiencies are still required [20].
The concept of hydrodynamic trapping for small trapping sites was originally proposed by Tan and Takeuchi [26]. However, a proof of concept is performed by experimental work only and no prior optimization of the microfluidic design through simulation works has been reported. From our point of view, this could probably involve high fabrication costs and it might be time consuming to find the optimized geometry through devices fabricated by trial and error. Therefore, our work is focused on developing the single-cell trapping model to produce a finite element simulation system that could be used to optimize a channel’s geometry for any type of cells or particles. The model is developed based on hydrodynamic flow resistance (Rh) manipulation in the main channel and trap channel to achieve successful trapping. This study provides a proof of concept demonstration for a cell positioning platform to trap single cells and a guideline for designing and optimizing single-cell trapping channel is proposed. The example of a thorough optimization study is presented using a 5-μm yeast cell model. Microchannels’ geometrical size optimization is carried out by manipulating the geometry of the trap channel, trap hole, and main channel. Numerical simulations are conducted to evaluate the cells’ trapping efficiencies for a variety of geometrical parameters. Fluid’s velocity profile and streamline plots are studied to explain the fluid’s stream direction according to the hydrodynamic principles. The single-cell trapping system is dependent on the cell’s size, as different cells require different optimized trapping channel sizes, trap hole’s sizes, and main channel lengths (LMain). Therefore, it is important for us to optimize the channel’s geometry before fabricating the real device to reduce time and fabrication costs.

2. Results and Discussion

2.1. Verification of Hydrodynamic Trapping Concept

The purpose of this finite element analysis is to verify the hydrodynamic trapping concept in the proposed model and to perform geometry optimization for efficient single-cell trapping. According to the hydrodynamic trapping concept proposed by Tan and Takeuchi [26], single-cell/particle trapping is achievable when the flow rate of trap channel to main channel (QTrap/QMain) ratio is above 1. To verify the concept, the cell trapping model with trap hole’s width (WHole) 2.0 μm is used to study the appropriate flow resistance of main channel to trap channel (RhMain/RhTrap) ratio. Main channel’s length (LMain) is manipulated to create an RhMain/RhTrap ratio ranging from 1 to 6. Increasing the RhMain/RhTrap ratio is proportional with the increase in the main channel’s (loop path) length. A yeast cell model is successfully trapped when a RhMain/RhTrap ratio of 3.5 or higher is used (Figure 1C,D). Furthermore, results show that an RhMain/RhTrap ratio ranging between 1.0 and 3.0 caused the cell to bypass the trap channel (Figure 1A,B).
From the simulation result, an RhMain/RhTrap ratio of 3.5 or above is found to be able to trap single cells via the hydrodynamic trapping concept. To further verify the principle of the hydrodynamic trapping, the fluid’s velocity inside the main channel and trap channel before and after trapping is analyzed. The velocity of the fluid at two points is analyzed (Figure 2A) to represent the fluid’s velocity before and after trapping for a cell trapping model with an RhMain/RhTrap ratio of 3.5 or 4.5 (Figure 2B). The fluid’s velocity in the main channel before cell trapping is found to be lower than the velocity in the trap channel (Figure 2B). However, after the cell is trapped inside the trap channel, the fluid’s velocity inside the trap channel decreases instantly and the fluid’s velocity at the main channel increases dramatically. This finding supports the principle of hydrodynamic trapping in which when the trapping side is empty, the trap channel will have lower flow resistance compared to the bypass channel (main channel). When the velocity of fluid in the trap channel is higher, it leads to a lower hydrodynamic resistance in the trapping site, which creates a trapping stream that will direct cells into the trap channel. When a cell has been trapped inside the trap channel, it blocks the trap hole and drastically decreases the fluid’s velocity inside the trap channel. The direction of fluid flow diverges from the trap channel to the loop path (main channel). Therefore, subsequent cells will be directed to the loop path. The simulation results are found to be in good agreement with the reported experimental results.
Figure 1. Cell trapping results at simulation time of 28 s for cell trapping model with WHole of 2 μm for different RhMain/RhTrap ratio of (A) 1.5; (B) 2.5; (C) 3.5; and (D) 4.5.
Figure 1. Cell trapping results at simulation time of 28 s for cell trapping model with WHole of 2 μm for different RhMain/RhTrap ratio of (A) 1.5; (B) 2.5; (C) 3.5; and (D) 4.5.
Ijms 16 25987 g001
Figure 2. (A) Points representing velocity of fluid in the trap channel (left side) and main channel (right side); (B) graph representing velocity of fluid in trap channel and main channel for cell trapping model with RhMain/RhTrap ratio of 3.5 or 4.5 before and after cell trapping.
Figure 2. (A) Points representing velocity of fluid in the trap channel (left side) and main channel (right side); (B) graph representing velocity of fluid in trap channel and main channel for cell trapping model with RhMain/RhTrap ratio of 3.5 or 4.5 before and after cell trapping.
Ijms 16 25987 g002

2.2. Effects of RhMain/RhTrap Ratio in Cell Trapping

Subsequent simulation is carried out to study the effects of the RhMain/RhTrap ratio on cell trapping using a model with trap hole widths of 1.0 or 1.5 μm. The main channel’s length has to be manipulated to comply with the desired RhMain/RhTrap ratio. Similar trapping behavior is obtained when the WHole is decreased to 1.5 μm. The hydrodynamic concept works accordingly and the yeast cell is able to be directed towards the trap channel by the fluid stream when the RhMain/RhTrap ratio is 3.5 and above. However, for models with a trap holes width of 1.0 μm, a cell is not able to be trapped even though the RhMain/RhTrap ratio is above 3.5. The cell is found to not be moving to the trap channel and bypasses it (data not shown). This result shows that a WHole of 1 μm is not suitable for the specified trap channel dimension (7 μm width, height, and length). The design fails to follow the hydrodynamic trapping concept, probably due to the small trap hole (<1/5 of trap channel’s width (WTrap)). The small WHole probably cause a very low fluid velocity distribution and produce low pressure drop that unable to capture cells into the trap channel. [28]. A simulation study performed by Khalili et al. [28] showed the same trend of results when a very small WHole/WTrap is used (<1/5). For designing a single-cell trapping channel, we suggest for the WHole to be more than 1/5 of WTrap for a uniform HChannel. Table 1 summarizes the single-cell trapping model’s ability for different WHole, HChannel, and trap channel’s length (LTrap), and various RhMain/RhTrap ratios.
Table 1. Cell trapping results for different single-cell trapping model with different sizes of Whole, Hchan, and LTrap and various RhMain/RhTrap.
Table 1. Cell trapping results for different single-cell trapping model with different sizes of Whole, Hchan, and LTrap and various RhMain/RhTrap.
Ratio of RhMain/RhTrapAbility to Trap Cells
WHole: 1.0 μmWHole: 1.5 μmWHole: 2.0 μmWHole: 2.5 μmWHole: 3.0 μmWHole: 3.5 μmHChan: 6.0 μmHChan: 7.0 μmHChan: 8.0 μmHChan: 9.0 μmLTrap: 3.0 μmLTrap: 5.0 μmLTrap: 7.0 μmLTrap: 9.0 μm
1.0nononononononononononononono
1.5nononononononononononononono
2.0nononononononononononononono
2.5nononononononononononononono
3.0nononononononononononononono
3.5noyesyesyesyesyesyesyesyesyesyesyesyesyes
4.5noyesyesyesyesyesyesyesyesyesyesyesyesyes
5.0noyesyesyesyesyesyesyesyesyesyesyesyesyes
5.5noyesyesyesyesyesyesyesyesyesyesyesyesyes
6.0noyesyesyesyesyesyesyesyesyesyesyesyesyes
Fixed geometry for WHole optimization: HChan: 7 μm, WTrap: 7 μm, LTrap: 7 μm, LHole: 1 μm; fixed geometry for HChannel optimization: WTrap: 7 μm, LTrap: 7 μm, WHole: 2 μm, LHole: 1 μm; fixed geometry for LTrap optimization: HChan: 7 μm, WTrap: 7 μm, LTrap: 7 μm, WHole: 2 μm.
The fluid velocity profile and velocity streamline field of the cell trapping model are analyzed to understand the hydrodynamic trapping mechanism. Fluid velocity streamlines present the direction the fluid streams are heading, while velocity profiles represent the velocity value in the channel by the contour color. The velocity streamlines produced by the cell trapping model with an RhMain/RhTrap ratio below 3.5 (Figure 3A,B), are found to be not fully directed to the trap channel and the portions of the streamlines that passed through the trap channel are directed to the loop. The produced fluid streams unable to direct the cell into the trap channel. This finding is in agreement with the fluid’s velocity distribution produced by the same model (Figure 4A,B). Results show that the main channel’s (loop path) fluid velocity for the single-cell trapping model with an RhMain/RhTrap ratio of 1.5 and 2.5 is higher compared to the trap channel’s fluid velocity. Therefore the main stream will direct the yeast cell to flow into the main channel’s path and bypass the trap channel.
In contrast with the cell trapping model with an RhMain/RhTrap ratio of 3.5 and above (Figure 3C,D), the streamlines profiles show the fluid flow diverging from the main channel to the trap channel and directed towards the trap channel. For models with an RhMain/RhTrap ratio of 3.5 or 4.5 (Figure 4C,D), the fluid’s velocity distribution from the trap hole to the trap channel is higher compared to the fluid’s velocity in the main channel. These results show that the trap channel produces lower hydrodynamic resistance than the main channel and the mainstream will direct the yeast cell into the trap channel. Both models with an RhMain/RhTrap ratio of 3.5 and 4.5 produce almost similar fluid velocity patterns that will produce appropriate pressure drop for the cell to be trapped.
Figure 3. Velocity streamlines before cell trapping (top view) for cell trapping model with WHole of 2 μm for different RhMain/RhTrap ratios of (A) 1.5; (B) 2.5; (C) 3.5; and (D) 4.5. V represents the fluid’s velocity in μms−1.
Figure 3. Velocity streamlines before cell trapping (top view) for cell trapping model with WHole of 2 μm for different RhMain/RhTrap ratios of (A) 1.5; (B) 2.5; (C) 3.5; and (D) 4.5. V represents the fluid’s velocity in μms−1.
Ijms 16 25987 g003
Figure 4. Velocity of fluid before cell trapping for single-cell trapping model with trapping hole width of 2 μm for RhMain/RhTrap ratios of (A) 1.5; (B) 2.5; (C) 3.5; and (D) 4.5. V represents the fluid’s velocity in μms−1.
Figure 4. Velocity of fluid before cell trapping for single-cell trapping model with trapping hole width of 2 μm for RhMain/RhTrap ratios of (A) 1.5; (B) 2.5; (C) 3.5; and (D) 4.5. V represents the fluid’s velocity in μms−1.
Ijms 16 25987 g004
The hydrodynamic trapping concept is found to be ineffective for a cell trapping model with a WHole of 1.0 μm. Subsequently, the LMain has been increased to obtain an RhMain/RhTrap ratio between 3.5 and 6.0; however, the cell trapping is not successful. The fluid velocity streamlines produced by this model show different profiles compared to the streamlines produced by models with WHole of 1.5 μm and 2.0 μm (Figure 5). The streamlines profile for the model shows that the flow direction is not fully focused into the trap channel but diverted to both the trap channel and the loop path directions (Figure 5A). The behavior of the fluid before cell bypass trap channel represents same trend of velocity profile and streamlines as models with unsuccessful trapping (Figure 3A,B). From the simulation results, the minimum main channel length needed to perform successful trapping is the length which produces an RhMain/RhTrap ratio of 3.5 (with the exception of the model with WHole of 1.0 μm).
Both cell trapping models with a trap hole width of 1.5 μm or 2.0 μm are found to be able to trap the yeast cell model with almost similar velocity profile. However, there are variations in the complete cell trapping time (time when the cell touches the surface of the trap channel) between different RhMain/RhTrap ratios. A higher ratio requires a shorter time for the trapping process compared to a lower ratio. The graph in Figure 6 shows the results of trapping time for cell trapping models with Whole of 1.5 and 2.0 μm for RhMain/RhTrap ranging from 3.5 to 6.0. From the graph, it is evident that the trapping time decreases with increasing RhMain/RhTrap. This is probably due to the higher RhMain/RhTrap ratio being able to perform velocity distribution in a shorter time compared to the lower RhMain/RhTrap. A greater RhMain/RhTrap ratio could provide a lower hydrodynamic resistance in the trap channel and could transfer the fluid at a faster rate. The velocity distribution produces different pressure from the main channel to the trap hole, making the flow resistance inside the trap channel lower than the main channel. Therefore, together with the fluid, cells will flow to the lower flow resistance area and be trapped. A bigger Whole value is able to produce shorter trapping time compared to the smaller height. Analyses are conducted for Whole of 2.5, 3.0, and 3.5 μm and similar results are obtained where the cell is able to be trapped with RhMain/RhTrap ratio of 3.5 and above (refer Table 1).
Figure 5. Velocity streamlines for the cell trapping model (top view) with RhMain/RhTrap ratio of 3.5 for model with trap hole width of (A) 1.0 μm; (B) 1.5 μm; or (C) 2.0 μm. V represents the fluid’s velocity in μms−1.
Figure 5. Velocity streamlines for the cell trapping model (top view) with RhMain/RhTrap ratio of 3.5 for model with trap hole width of (A) 1.0 μm; (B) 1.5 μm; or (C) 2.0 μm. V represents the fluid’s velocity in μms−1.
Ijms 16 25987 g005
Figure 6. Cell trapping time for model with different RhMain/RhTrap ranging from 3.5 to 6.0 for single-cell trapping model with trapping hole widths of 1.5 and 2.0 μm.
Figure 6. Cell trapping time for model with different RhMain/RhTrap ranging from 3.5 to 6.0 for single-cell trapping model with trapping hole widths of 1.5 and 2.0 μm.
Ijms 16 25987 g006

2.3. Optimization of Trap Channel’s Length

After investigating the effects of RhMain/RhTrap ratio for the single cell trapping model, the efficiency of the single-cell trapping is enhanced by optimizing the trap channel’s length (LTrap) (refer Figure 2A). Using yeast cell and four different LTrap, the behavior of cell trapping is observed. A model with LTrap of 3 μm is able to trap single cells; however, after a cell is trapped, both of the paths to the loop and the outlet will eventually be blocked, causing clogging of subsequent cells at the main channel and thus preventing the smooth movement of cells towards the outlet (Figure 7A). Therefore the length is not suitable for efficient cell trapping. From the results, a model with LTrap of 5 μm is found to be the most suitable length to trap a 5-μm yeast cell as it could allow the subsequent cell to flow to the loop and heading to the outlets (Figure 7B). For a model with LTrap of 7 or 9 μm, results show that two cells are able to enter the trap channel during cell trapping (Figure 7C,D). The aim for the cell trapping model development is to trap a single cell; therefore the LTrap of 7 and 9 μm are not suitable for efficient single-cell trapping.
Figure 7. Cell trapping results for the optimization of different LTrap values: (A) 3 μm; (B) 5 μm; (C) 7 μm; and (D) 9 μm.
Figure 7. Cell trapping results for the optimization of different LTrap values: (A) 3 μm; (B) 5 μm; (C) 7 μm; and (D) 9 μm.
Ijms 16 25987 g007
Figure 8. Cell trapping results at simulation time of 34s for cell trapping model with the same trap hole size (2, 7 and 7 μm of W, L and H, respectively) for an RhMain/RhTrap ratio of 3.5 at three different positions: (A) model A; (B) model B; and (C) model C. (Left) side view; (right) top view of the model.
Figure 8. Cell trapping results at simulation time of 34s for cell trapping model with the same trap hole size (2, 7 and 7 μm of W, L and H, respectively) for an RhMain/RhTrap ratio of 3.5 at three different positions: (A) model A; (B) model B; and (C) model C. (Left) side view; (right) top view of the model.
Ijms 16 25987 g008

2.4. Effects of Different Trap Hole Positions

The final analysis is carried out to study the effects of the trap hole’s position on the cell trapping. Analysis is carried out using a cell trapping model with an RhMain/RhTrap ratio of 3.5 and trap hole dimensions of 1, 7, and 2 μm in length, height, and width, respectively. Three different trap hole positions with similar dimensions are analyzed as illustrated in Figure 12, namely models A–C. Cell trapping results demonstrate that all of the models are able to trap cells with an RhMain/RhTrap ratio of 3.5 (Figure 8). The streamlines velocity fields produced by the models before cell trapping are focused towards the trap channel. The streamlines show that the fluid stream produced is fully directed toward the trap channel with a similar pattern. The only difference in the streamlines pattern between the models is the position of streams towards the trap hole (Figure 9B,D). The differences could be observed by viewing the streamlines at three different views (Figure 9A). For model A, the streamlines’ focusing could be clearly seen from the top view, where the streamlines’ direction focused on the center of the trap channel (Figure 9A(i)). For models B and C, the streamline focusing could be clearly differentiated by viewing from the front and side (Figure 9C,D(i,ii)), where model B streamlines are focused at the base of the channel while model C are focused at the middle of the trap channel. The findings suggest that a cell trapping model with similar RhMain/RhTrap ratio produces similar trapping behavior despite the different trap hole’s positions.
Figure 9. Streamline velocity for three different cell trapping models with the same trap hole size (2 μm, 7 μm, and 7 μm of W, L, and H, respectively) for an RhMain/RhTrap ratio of 3.5 at three different positions: (A) model A; (B) model B; and (C) model C. (Left) side view; (right) top view (top) and front view (bottom) of the model. V represents the fluid’s velocity in μms−1.
Figure 9. Streamline velocity for three different cell trapping models with the same trap hole size (2 μm, 7 μm, and 7 μm of W, L, and H, respectively) for an RhMain/RhTrap ratio of 3.5 at three different positions: (A) model A; (B) model B; and (C) model C. (Left) side view; (right) top view (top) and front view (bottom) of the model. V represents the fluid’s velocity in μms−1.
Ijms 16 25987 g009

3. The Concept of the Model

The hydrodynamic trapping concept can be summarized as follows: (a) the trapping channel has a lower Rh than the by-passing channel when a trapping site is empty, and will make the particles/cells flow into the trapping stream and directed into the trap; (b) when a bead/cell is trapped, it will act as a plug and will increase the Rh along the trap channel drastically; and (c) the main flow will change from the trap channel to the by-pass channel (main channel) and the next particles/cells will be directed to the by-pass stream, passing by the filled trapping site [29]. Figure 10 shows a schematic explanation of the hydrodynamic trapping concept with RhTrap and RhMain representing the flow resistance of trap channel and main channel, respectively. The yellow circle denotes a yeast cell that needs to be trapped.
Figure 10. Simple schematic of single-cell trapping channel with the hydrodynamic resistance.
Figure 10. Simple schematic of single-cell trapping channel with the hydrodynamic resistance.
Ijms 16 25987 g010
The Darcy-Weisbach equation is used to determine the pressure drop or pressure difference in a microchannel and solve the continuity and momentum equations for the Hagen-Poiseuille flow problem. From the Hagen–Poiseuille equation, the flow rate (Q) can be defined by the following equation:
P = Q × R h = Q × ( C μ L P 2 A 3 )
where ∆P is the pressure drop, Rh is the flow resistance of the rectangular channels, C is a constant that depends on the aspect ratio (ratio between height and width of the channel), μ is the fluid’s viscosity, and L, P, and A are the length, perimeter, and cross-sectional area of the channel, respectively.
From Equation (1), by approximating that the pressure drop across the trap channel and the main channel are the same (∆PTrap = ∆PMain), the flow rate ratio (QTrap/Qmain) or flow resistance ratio (RhMain/RhTrap) between the trap channel and the main channel can be given as follows [30]:
Q Trap Q Main = R h Main R h Trap = ( C Main C Trap )   ( L Main L Trap ) ( P Main P Trap ) 2 ( A Trap A Main ) 3
By using a relationship of A = W × H and P = 2 (W + H), where W and H are the width and height of the channel, respectively, Equation (2) can be defined as:
Q Trap Q Main = R h Main R h Trap = ( C Main C Trap ) ( L Main L Trap ) ( W Main + H Main W Trap + H Trap ) 2 ( W Trap H Trap W Main H Main ) 3
From Equations (2) and (3), it is noted that the flow rates of the trap channel (QTrap) and the main channel (QMain) are distributed depending on the corresponding Rh. For the trap to work, the flow rate along the trap channel must be greater than that of main channel (QTrap > QMain). In other words, the flow resistance along the main channel must be greater than that of the trap channel (RhMain > RhTrap). Therefore, a single cell can be trapped by manipulating the flow resistance ratio (RhMain/RhTrap), which is determined by the geometric parameters of the channels.
A single-cell trapping model is developed to produce a finite element single-cell trapping system in which the optimization of a channel’s geometry, dependent on the desired cell size, could be performed. The geometry of the trapping channel is a variable for optimization (see Equation (3)) and subject to the size of cells and the application that will be carried out in the channel after the cells are trapped. An example of a thorough optimization study is presented in this paper using a 5-μm yeast cell model. For other cell sizes, a guideline for designing and optimizing the cell trapping channel is proposed. Firstly the diameter of the viable cells in suspension (floating cells) before cell adhesion occur (for adherent type of cells) should be determined. This is important to determine the range of suitable trap hole sizes. We suggest that the WHole to be less than one third of the cell’s size due to the ability of cells to deform and the flexibility to enter the trap hole instead of being trapped in the trap channel [30]. This could happen, especially to cells that have no cell wall such as human cells. Next, after determining WHole, the HChannel and WTrap have to be optimized. HChannel should be bigger than the diameter of the cells to reduce friction between the cell surface and the channel’s wall and to avoid cell squeezing (for applications that do not require cells to be squeezed, e.g., cell culturing, drugs treatment, and cell adhesion study). The optimization of the LTrap, is dependent on the application of cells after being trapped. Long LTrap could cause more than one cell to be trapped if cells in suspension are very near to each other. However, for long-term monitoring of cell behavior for Tetrahymena thermophila, a long trap channel is needed to avoid cell from swimming back to the main channel [30]. The trap channel’s geometry size choices are dependent on the application of the trapping platform after the cells/particles are trapped. For example, if adherence cells are used and need to be cultured inside the trapping platform, the WTrap should be bigger than the diameter of the cell (viable cells in suspension before adhesion). This is because cells need space for cell adhesion and spreading as the diameter of cells after adhesion will increase depending on culture time. In different applications, individual ciliate protozoan, Tetrahymena thermophila [30] need to be trapped and maintained in the trap channel for long-term monitoring of cell behavior. Therefore, no expansion in size is expected after the trapping process and the trap channel’s width does not require space for expansion. In summary, the geometry of channels is a variable (L, H, and W; see Equation (3)) for optimization, subject to the size of cells used and the application that will be carried out after the cells are trapped.
In this single-cell trapping model, cells are introduced into the device through the inlet with an appropriate flow rate and directed to the trap channel by optimizing the channel’s geometry. Trap hole and trap channel geometry are optimized and LMain is manipulated to produce an appropriate Rh ratio that leads to successful trapping (see Equation (3)). The excess and remaining cells will be directed out through the channel’s outlet by injecting cell’s culture medium. The appropriate channel’s geometry to trap a 5-μm single yeast cell in the specified design is studied. The finite element single-cell trapping model is focusing only on a single trap channel (see dashed box in Figure 1) for geometry optimization due to the complexity and high processing time required for the analysis.

4. Simulation Setup

The analysis is carried out using finite element ABAQUS-FEA™ analysis software, which can perform multiphysics analyses. The single-cell trapping model consists of two different parts, the Eulerian part as the fluid channel and a three-dimensional (3D) deformable part as the sphere-shaped elastic yeast cell model (Figure 11A,B). The fluid consists of two microchannels, the main channel (loop channel) and a trap channel with a rectangular trap hole placed in the center, at the edge of the trap channel. The microchannel is modeled as 3D Eulerian explicit EC3DR and an eight-node linear Eulerian brick element part assigned with water properties (density, equation of state, and viscosity). A sphere-shaped yeast cell (5 μm in diameter) is modeled as an elastic 3D standard solid deformable C3D8R and an eight-node linear brick 3D part with the yeast properties (Young’s modulus, Poisson’s ratio, and density) obtained from literature [31,32,33,34,35,36,37,38].
Figure 11C shows the assembly setup with a yeast cell positioned in the main channel, near the channel’s inlet (left). The parts are assembled to develop the finite element model for the proposed system (Figure 11C). The initial position of the cell is fixed (same distance between cell and trap channel) for all models. Interaction between cell and water is set as general contact with rough tangential behavior and the interaction between cell surface and channel’s wall is set as frictionless. The fluid channel and cell are meshed using hexahedron mesh types. Total mesh elements for the cell trapping model ranged from 10,627 to 22,485 elements. No-inflow and non-reflecting outflow Eulerian boundary conditions are applied to the channel’s wall. A constant inflow velocity of 0.5 μms−1 is applied to the inlet and atmosphere pressure is applied to the outlet of the channel.
Figure 11. Construction of the finite element model of single-cell trapping system and parts involved: (A) Eulerian part (fluid channel’s top view) LMain represents the main channel’s length and LTrap represents the trap channel’s length; (B) 3D deformable part (yeast cell model); (C) simulation’s assembly setup (cell is positioned between inlet and trap channel as initial position). WHole represents trap hole’s width.
Figure 11. Construction of the finite element model of single-cell trapping system and parts involved: (A) Eulerian part (fluid channel’s top view) LMain represents the main channel’s length and LTrap represents the trap channel’s length; (B) 3D deformable part (yeast cell model); (C) simulation’s assembly setup (cell is positioned between inlet and trap channel as initial position). WHole represents trap hole’s width.
Ijms 16 25987 g011
The simulation analysis could be divided into four parts: the verification of the hydrodynamic trapping concept, the effects of RhMain/RhTrap ratio in cell trapping, the optimization of the trap channel’s length, and the effects of the trap hole’s position. For the verification of the hydrodynamic trapping concept, a model with a trap hole’s width of 2.0 μm is used for the analysis. To study the effects of RhMain/RhTrap ratio in cell trapping, various LMain ranging from 46 to 268 μm (Figure 11A) and WHole ranging from 1.0 to 3.5 μm (Figure 2C) with fixed LHole of 1 μm are applied to obtain the appropriate RhMain/RhTrap ratio for cell trapping. The height of main channel, trap channel, and trap hole are uniform (HChannel) and were tested in the range of 6–9 μm and set to be 7 μm throughout the analysis. For trap channel length (LTrap) optimization, various trap channel lengths from 3 to 9 μm and a fixed trap channel width of 7 μm are used, with three yeast cells in the analysis. Lastly, to study the effects of trap hole’s position, three different positions for similar trap hole’s dimensions are studied to observe the ability of the model for cell trapping. Figure 12 shows the different views for the three different positions, represented by models A, B, and C.
Figure 12. The views for three cell trapping models with the same trap hole size (2, 7 and 7 μm for WHole, LHole, and HHole, respectively) at different positions.
Figure 12. The views for three cell trapping models with the same trap hole size (2, 7 and 7 μm for WHole, LHole, and HHole, respectively) at different positions.
Ijms 16 25987 g012

5. Conclusions

This study presents the finite element model of single-cell trapping inside microfluidic channel. This single-cell trapping system is constructed using Abaqus-FEA™ software. A guideline to design and optimize single-cell trapping model is proposed and the example of a thorough optimization analysis is carried out using a yeast cell model. The results show that the finite element model is able to trap a single cell inside the fluidic environment. The fluid velocity profile and streamline plots of successful and unsuccessful single yeast cell trapping are presented according to the hydrodynamic concept. This cell trapping model is able to isolate an individual yeast cell inside a fluidic environment, thus providing a platform for further single-cell mechanical or biological study. Single-cell manipulation such as chemical and biophysical treatments and also mechanical characterization could be performed inside the microfluidic channel using this system. The single-cell trapping model can be a significant important guideline in designing a new chip for biomedical applications.

Acknowledgments

The research was supported by the Ministry of Higher Education of Malaysia (grant Nos. 4L640 and 4F351), and Universiti Teknologi Malaysia (grant Nos. 02G46, 03H82, and 03H80); we thank them for funding this project and for their endless support. The authors would like to express their heartiest appreciation to Mohd Ariffanan Mohd Basri from the Faculty of Electrical Engineering, Universiti Teknologi Malaysia for his contribution of ideas and valuable discussion during the theoretical development of this study.

Author Contributions

Amelia Ahmad Khalili and Mohd Ridzuan Ahmad wrote and edited the article, respectively.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Johann, R.M. Cell Trapping in Microfluidic Chips. Anal. Bioanal. Chem. 2006, 385, 408–412. [Google Scholar] [CrossRef] [PubMed]
  2. Lee, G.-H.; Kim, S.-H.; Kang, A.; Takayama, S.; Lee, S.-H.; Park, J.Y. Deformable L-shaped microwell array for trapping pairs of heterogeneous cells. J. Micromech. Microeng. 2015, 25. [Google Scholar] [CrossRef]
  3. Sun, T.; Kovac, J.; Voldman, J. Image-based single-cell sorting via dual-photopolymerized microwell arrays. Anal. Chem. 2014, 86, 977–981. [Google Scholar] [CrossRef] [PubMed]
  4. Rettig, J.R.; Folch, A. Large-scale single-cell trapping and imaging using microwell arrays. Anal. Chem. 2005, 77, 5628–5634. [Google Scholar] [CrossRef] [PubMed]
  5. Tang, J.; Peng, R.; Ding, J. The Regulation of Stem cell differentiation by cell-cell contact on micropatterned material surfaces. Biomaterials 2010, 31, 2470–2476. [Google Scholar] [CrossRef] [PubMed]
  6. Doh, J.; Kim, M.; Krummel, M.F. Cell-laden microwells for the study of multicellularity in lymphocyte fate decisions. Biomaterials 2010, 31, 3422–3428. [Google Scholar] [CrossRef] [PubMed]
  7. Chen, N.-C.; Chen, C.-H.; Chen, M.-K.; Jang, L.-S.; Wang, M.-H. Single-cell trapping and impedance measurement utilizing dielectrophoresis in a parallel-plate microfluidic device. Sens. Actuators B Chem. 2014, 190, 570–577. [Google Scholar] [CrossRef]
  8. Sen, M.; Ino, K.; Ramon-Azcon, J.; Shiku, H.; Matsue, T. Cell pairing using a dielectrophoresis-based device with interdigitated array electrodes. Lab Chip 2013, 13, 3650–3652. [Google Scholar] [CrossRef] [PubMed]
  9. Voldman, J.; Gray, M.L.; Toner, M.; Schmidt, M.A. A Microfabrication-based dynamic array cytometer. Anal. Chem. 2002, 74, 3984–3990. [Google Scholar] [CrossRef] [PubMed]
  10. Thomas, R.S.; Morgan, H.; Green, N.G. Negative DEP traps for single cell immobilisation. Lab Chip 2009, 9, 1534–1540. [Google Scholar] [CrossRef] [PubMed]
  11. Gray, D.S.; Tan, J.L.; Voldman, J.; Chen, C.S. Dielectrophoretic registration of living cells to a microelectrode array. Biosens. Bioelectron. 2004, 19, 771–780. [Google Scholar] [CrossRef] [PubMed]
  12. Chen, Y.-C.; Allen, S.G.; Ingram, P.N.; Buckanovich, R.; Merajver, S.D.; Yoon, E. Single-cell migration chip for chemotaxis-based microfluidic selection of heterogeneous cell populations. Sci. Rep. 2015, 5, 1–13. [Google Scholar] [CrossRef] [PubMed]
  13. Jin, D.; Deng, B.; Li, J.X.; Cai, W.; Tu, L.; Chen, J.; Wu, Q.; Wang, W.H. A Microfluidic device enabling high-efficiency single cell trapping. Biomicrofluidics 2015, 9. [Google Scholar] [CrossRef] [PubMed]
  14. Benavente-Babace, A.; Gallego-Pérez, D.; Hansford, D.J.; Arana, S.; Pérez-Lorenzo, E.; Mujika, M. Single-cell trapping and selective treatment via co-flow within a microfluidic platform. Biosens. Bioelectron. 2014, 61, 298–305. [Google Scholar] [CrossRef] [PubMed]
  15. Kim, J.; Erath, J.; Rodriguez, A.; Yang, C. A High-efficiency microfluidic device for size-selective trapping and sorting. Lab Chip 2014, 14, 2480–2490. [Google Scholar] [CrossRef] [PubMed]
  16. Lee, P.J.; Hung, P.J.; Shaw, R.; Jan, L.; Lee, L.P. Microfluidic application-specific integrated device for monitoring direct cell-cell communication via gap junctions between individual cell pairs. Appl. Phys. Lett. 2005, 86. [Google Scholar] [CrossRef]
  17. Frimat, J.-P.; Becker, M.; Chiang, Y.-Y.; Marggraf, U.; Janasek, D.; Hengstler, J.G.; Franzke, J.; West, J. A microfluidic array with cellular valving for single cell co-culture. Lab Chip 2011, 11, 231–237. [Google Scholar] [CrossRef] [PubMed]
  18. Kim, H.; Lee, S.; Kim, J. Hydrodynamic trap-and-release of single particles using dual-function elastomeric valves: design, fabrication, and characterization. Microfluid. Nanofluid. 2012, 13, 835–844. [Google Scholar] [CrossRef]
  19. Arakawa, T.; Noguchi, M.; Sumitomo, K.; Yamaguchi, Y.; Shoji, S. High-throughput single-cell manipulation system for a large number of target cells. Biomicrofluidics 2011, 5. [Google Scholar] [CrossRef] [PubMed]
  20. Kobel, S.; Valero, A.; Latt, J.; Renaud, P.; Lutolf, M. Optimization of microfluidic single cell trapping for long-term on-chip culture. Lab Chip 2010, 10, 857–863. [Google Scholar] [CrossRef]
  21. Hong, S.; Pan, Q.; Lee, L.P. Single-cell level co-culture platform for intercellular communication. Integr. Biol. 2012, 4, 374–380. [Google Scholar] [CrossRef] [PubMed]
  22. Shi, W.; Qin, J.; Ye, N.; Lin, B. Droplet-based microfluidic system for individual caenorhabditis elegans assay. Lab Chip 2008, 8, 1432–1435. [Google Scholar] [CrossRef] [PubMed]
  23. Di Carlo, D.; Aghdam, N.; Lee, L.P. Single-cell enzyme concentrations, kinetics, and inhibition analysis using high-density hydrodynamic cell isolation arrays. Anal. Chem. 2006, 78, 4925–4930. [Google Scholar] [CrossRef] [PubMed]
  24. Skelley, A.M.; Kirak, O.; Suh, H.; Jaenisch, R.; Voldman, J. Microfluidic control of cell pairing and fusion. Nat. Methods 2009, 6, 147–152. [Google Scholar] [CrossRef] [PubMed]
  25. Di Carlo, D.; Wu, L.Y.; Lee, L.P. Dynamic single cell culture array. Lab Chip 2006, 6, 1445–1449. [Google Scholar] [CrossRef] [PubMed]
  26. Tan, W.-H.; Takeuchi, S. A Trap-and-release Integrated microfluidic system for dynamic microarray applications. Proc. Natl. Acad. Sci. USA 2007, 104, 1146–1151. [Google Scholar] [CrossRef] [PubMed]
  27. Chung, K.; Rivet, C.A.; Kemp, M.L.; Lu, H.; States, U. Imaging single-cell signaling dynamics with a deterministic high-density single-cell trap array. Anal. Chem. 2011, 83, 7044–7052. [Google Scholar] [CrossRef] [PubMed]
  28. Khalili, A.A.; Basri, M.A.M.; Ahmad, M.R. Simulation of single cell trapping via hydrodynamic manipulation. J. Teknol. 2014, 69, 121–126. [Google Scholar]
  29. Teshima, T.; Ishihara, H.; Iwai, K.; Adachi, A.; Takeuchi, S. A dynamic microarray device for paired bead-based analysis. Lab Chip 2010, 10, 2443–2448. [Google Scholar] [CrossRef] [PubMed]
  30. Kumano, I.; Hosoda, K.; Suzuki, H.; Hirata, K.; Yomo, T. Hydrodynamic trapping of tetrahymena thermophila for the long-term monitoring of cell behaviors. Lab Chip 2012, 12, 3451–3457. [Google Scholar] [CrossRef] [PubMed]
  31. Gervais, T.; El-Ali, J.; Günther, A.; Jensen, K.F. Flow-induced deformation of shallow microfluidic channels. Lab Chip 2006, 6, 500–507. [Google Scholar] [CrossRef] [PubMed]
  32. Bryan, A.K.; Goranov, A.; Amon, A.; Manalis, S.R. Measurement of mass, density, and volume during the cell cycle of yeast. Proc. Natl. Acad. Sci. USA 2010, 107, 999–1004. [Google Scholar] [CrossRef] [PubMed]
  33. Ahmad, M.R.; Nakajima, M.; Kojima, S.; Homma, M.; Fukuda, T. The Effects of cell sizes, environmental conditions, and growth phases on the strength of individual W303 yeast cells inside ESEM. IEEE Trans. Nanobiosci. 2008, 7, 185–193. [Google Scholar] [CrossRef] [PubMed]
  34. Smith, A.E.; Zhang, Z.; Thomas, C.R.; Moxham, K.E.; Middelberg, A.P. The Mechanical properties of saccharomyces cerevisiae. Proc. Natl. Acad. Sci. USA 2000, 97, 9871–9874. [Google Scholar] [CrossRef] [PubMed]
  35. Stenson, J.D.; Thomas, C.R.; Hartley, P. Modelling the mechanical properties of yeast cells. Chem. Eng. Sci. 2009, 64, 1892–1903. [Google Scholar] [CrossRef]
  36. Stenson, J.D.; Hartley, P.; Wang, C.; Thomas, C.R. Determining the mechanical properties of yeast cell walls. Biotechnol. Prog. 2011, 27, 505–512. [Google Scholar] [PubMed]
  37. Burg, T.P.; Godin, M.; Knudsen, S.M.; Shen, W.; Carlson, G.; Foster, J.S.; Babcock, K.; Manalis, S.R. Weighing of Biomolecules, Single Cells and Single Nanoparticles in Fluid. Nature 2007, 446, 1066–1069. [Google Scholar] [CrossRef] [PubMed]
  38. Lee, J.; Chunara, R.; Shen, W.; Payer, K.; Babcock, K.; Burg, T.P.; Manalis, S.R. Suspended microchannel resonators with piezoresistive sensors. Lab Chip 2011, 11, 645–651. [Google Scholar] [CrossRef]

Share and Cite

MDPI and ACS Style

Khalili, A.A.; Ahmad, M.R. Numerical Analysis of Hydrodynamic Flow in Microfluidic Biochip for Single-Cell Trapping Application. Int. J. Mol. Sci. 2015, 16, 26770-26785. https://0-doi-org.brum.beds.ac.uk/10.3390/ijms161125987

AMA Style

Khalili AA, Ahmad MR. Numerical Analysis of Hydrodynamic Flow in Microfluidic Biochip for Single-Cell Trapping Application. International Journal of Molecular Sciences. 2015; 16(11):26770-26785. https://0-doi-org.brum.beds.ac.uk/10.3390/ijms161125987

Chicago/Turabian Style

Khalili, Amelia Ahmad, and Mohd Ridzuan Ahmad. 2015. "Numerical Analysis of Hydrodynamic Flow in Microfluidic Biochip for Single-Cell Trapping Application" International Journal of Molecular Sciences 16, no. 11: 26770-26785. https://0-doi-org.brum.beds.ac.uk/10.3390/ijms161125987

Article Metrics

Back to TopTop