Diagnostic Accuracy of Contrast-Enhanced, Spectral Mammography (CESM) and 3T Magnetic Resonance Compared to Full-Field Digital Mammography plus Ultrasound in Breast Lesions: Results of a (Pilot) Open-Label, Single-Centre Prospective Study

1
Radiology and Diagnostic Imaging Department, IRCCS Regina Elena National Cancer Institute, Via Elio Chianesi 53, 00144 Rome, Italy
2
Division of Medical Oncology 2, IRCCS Regina Elena National Cancer Institute, Via Elio Chianesi 53, 00144 Rome, Italy
3
UOSD Clinical Trial Center, Biostatistical and Bioinformatic, Scientific Direction, San Gallicano Dermatological Institute IRCCS, Via Elio Chianesi 53, 00144 Rome, Italy
4
Division of Breast Surgery, IRCCS Regina Elena National Cancer Institute, Via Elio Chianesi 53, 00144 Rome, Italy
5
Sperimentazioni di Fase IV, IRCCS Regina Elena National Cancer Institute, 00144 Rome, Italy
*
Authors to whom correspondence should be addressed.
Submission received: 24 February 2022 / Revised: 3 March 2022 / Accepted: 4 March 2022 / Published: 7 March 2022
(This article belongs to the Special Issue Advanced Breast Cancer: From Biology to Cure)

Abstract

:

Simple Summary

The characterization of breast lesions by magnetic resonance imaging (MRI) is influenced by a high rate of false positives. Contrast-enhanced spectral mammography (CESM) is a promising modality that seems to compensate for the high costs, times and main limitations of MRI. The aim of our prospective study was to evaluate the diagnostic accuracy of CESM in comparison to 3T MRI imaging in the characterization of breast lesions. We enrolled 118 patients and histologically assessed 142 breast lesions. Patients underwent full-field digital mammography (FFDM), ultrasound (US), with CESM and MRI assessing the diagnostic accuracy of CESM. Sensitivity (Se), specificity (Sp), positive predictive value (PPV) and negative predictive value (NPV) were considered as measures of accuracy of different diagnostic procedures in predicting the nature and characteristics of the examined breast lesions.

Abstract

Introduction: To assess the diagnostic accuracy of CESM and 3T MRI compared to full-field digital mammography (FFDM), plus US, in the evaluation of advanced breast lesions. Materials and Methods: Consenting women with suspicious findings underwent FFDM, US, CESM and 3T MRI. Breast lesions were histologically assessed, with histology being the gold standard. Two experienced breast radiologists, blinded to cancer status, read the images. Diagnostic accuracy of (1) CESM as an adjunct to FFDM and US, and (2) 3T MRI as an adjunct to CESM compared to FFDM and US, was assessed. Measures of accuracy were sensitivity (Se), specificity (Sp), positive predictive value (PPV) and negative predictive value (NPV). Results: There were 118 patients included along with 142 histologically characterized lesions. K agreement values were 0.69, 0.68, 0.63 and 0.56 for concordance between the gold standard and FFDM, FFDM + US, CESM and MRI, respectively (p < 0.001, for all). K concordance for CESM was 0.81 with FFDM + US and 0.73 with MRI (p value < 0.001 for all). Conclusions: CESM may represent a valuable alternative and/or an integrating technique to MRI in the evaluation of breast cancer patients.

1. Introduction

Female breast cancer is the most common cancer diagnosed worldwide, ranking fifth among the most frequent causes of cancer death [1]. In recent years, relevant advances in breast cancer diagnostics have been achieved. Contrast-enhanced dual-energy spectral mammography (CESM) is an imaging technique based on the use of a dual-energy approach for cancer detection. In more detail, with CESM, the low-energy (LE) component of breast images renders the morphological information similar to a two-dimensional (2D) digital mammography, while the high-energy (HE) component of breast images demonstrates post-contrast enhanced mammograms by using the K-edge effect of iodine, allowing for the evaluation of tumor neovascularity [2]. Although CESM is still in the early stages of clinical use, results from the related studies appear extremely encouraging [3]. When evaluating 143 breasts of 72 women and considering histopathology to be the gold standard, Mori and colleagues observed significantly greater sensitivity, specificity and accuracy of CESM compared to conventional full-field digital mammography (FFDM) [4].
The findings concerning CESM sensitivity have been strengthened by the results of the systematic review and meta-analysis of eight eligible studies that included 920 patients with 994 lesions [5]. In addition, in a feasibility study of CESM compared to FFDM to assess residual tumors following neoadjuvant chemotherapy, the sensitivity and specificity of CESM were about 83% and 100%, respectively, compared to 50% (for both) with FFDM. Similarly, positive predictive and negative predictive values were higher for CESM [6]. Adjunctive evidence in support of the superior accuracy of CESM in evaluating tumor size compared to FFDM and ultrasound (US), both in dense and non-dense breasts, is derived from a recent study [7]. Finally, CESM has shown increased diagnostic performance compared to FFDM in lower-prevalence patient populations, such as those referred from breast cancer screening [8]. However, exclusive evidence from an extremely recent systematic review and meta-analysis, including prospective trials, was not supportive. The accuracy of CESM is now under debate [9].
MRI is an extremely accurate imaging tool in breast cancer diagnosis. Breast MRI can efficiently integrate diagnostic information to identify breast lesions not definitively diagnosed by FFDM and US. MRI sensitivity in cancer diagnosis has been estimated as ranging from 80 to 97.8%, while specificity is between 46 and 93.3%. The high rate of false positives, along with the high expenses, difficult accessibility and contraindications in patients with metallic implants and pacemakers, poses remarkable limitations to its use in breast cancer diagnostics and screening programs [10,11].
Globally considered, the above-cited literature fueled our interest in this timely topic, which culminated in the design and performance of a pilot, open-label, prospective study conceived and conducted at our facility, the Regina Elena National Cancer Institute. Our trial aimed to compare diagnostic accuracy of CESM and 3T MRI with FFDM and US to assess breast cancer lesions. In more detail, we evaluated diagnostic accuracy of CESM as adjunct to FFDM + US and MRI as adjunct to FFDM + US. In doing so, we addressed an extremely timely issue to a research agenda focused on the assessment and characterization of breast lesions.

2. Materials and Methods

This study was conceived as an open-label, single-center study carried out at the Radiology Department of the Regina Elena National Cancer Institute. The study protocol and inherent consent form were submitted to the Institutional Review Board for formal evaluation and then approval (RS 410/13). The outcome assessment was based on the evaluation of accuracy expressed in terms of sensitivity (Se), specificity (Sp), positive predictive value (PPV) and negative predictive value (NPV). Data analysis was also performed using SPSS version 20.0 (SPSS Inc., Chicago, IL, USA).
Consenting women with unresolved/suspicious findings after FFDM and US underwent a diagnostic workup that included baseline (FFDM and US) reassessment, plus CESM and 3T MRI after being informed and after having signed an informed consent form. The timing of breast imaging differed by menopausal status. In postmenopausal women, defined as a self-reported (at least) 12-month period of amenorrhea, the estimated time interval between baseline reassessment and performance of CESM was (on average) 1 week. This same time window was expected between CESM and administration of the paramagnetic contrast agent (gadolinium) and between the latter, as well as 3T MRI. In premenopausal women, any of the aforementioned procedures were performed during the second week of their menstrual cycle. The actual nature and characterization of breast lesions was histologically assessed in samples collected by core biopsy and/or surgery. Two experienced breast radiologists read the FFDM, US, CESM and 3T MRI, with a minimum 4-week washout period among different sessions for the same patient. Outcome assessors were blinded to cancer status and were able to work in consensus.
Any breast lesions detected were characterized based on data gathered from breast imaging and pathology records. Features of interest included: lesion presence and location (breast site and quadrant); finding the type (e.g., mass and architectural distortion) and size (largest diameter) from FFDM and US, CESM and 3T MRI; BIRADS classification (FFDM, CESM and 3T MRI) (12); mammographic density (FFDM); type of contrast kinetic curves (CESM and 3T MRI); and (pathologic) diagnosis assessed by cytology/histology (e.g., benign/malignant).
Patients of at least 18 years of age were eligible if presenting with unresolved/suspicious findings after FFDM plus US, e.g., breast mass, breast asymmetry and distortion. Moreover, fine-needle aspiration biopsy, core-needle biopsy and/or breast surgery had to be scheduled independently for our study purposes. Written informed consent was secured from each participating patient. Eligibility was not granted if one or more of the following was verified: prior diagnosis of breast and/or other malignancy/ies, except adequately treated non-melanoma skin cancer and/or curatively treated in situ cancer of the cervix; exclusive evidence of clusters of microcalcifications in the absence of previously mentioned findings (e.g., mass and asymmetry); history of allergy to contrast medium and/or multiple allergies; pregnancy and lactation; and breast implants or any unstable medical conditions, which could interfere with safe participation in the trial. All eligible participants repeated baseline FFDM and US in radiology. Each patient underwent CESM and 3T MRI. A detailed description of the procedures and related protocols is reported below.

2.1. CESM Examinations

CESM examinations were performed with a device developed by GE Healthcare (Chicago, IL, USA), allowing dual-energy CESM acquisitions. It uses a current full-field digital mammography system, SenoBright, or Senographe Essential. A catheter was inserted into the antecubital vein of the contralateral arm to the breast of concern. A one-shot intravenous injection of 1.5 mL/body weighted contrast agent (Visipaque 320, GE, Oslo, Norvegia) was then performed using a power injector. Two minutes after injection of the contrast agent, a bilateral mammography examination with standard compression was performed in craniocaudal (CC) and medio-lateral-oblique (MLO) views, each with a pair of low- and high-energy exposures and in a total examination time of 6 min. A combination of low-energy and high-energy images through specific image processing was automatically provided by the system to generate subtracted images with contrast agent uptake information (one in the MLO and one in the CC view of each breast). The mean, minimum and maximum X-ray doses administered in the course of FFDM and CESM are shown in Table 1.

2.2. T MR Examinations

Breast MRI scan was conducted with 3T Tomography (Discovery MR750w GE Medical Systems, Waukesha, WI, USA) and with an 8-channel surface coil to assess the patients’ bilateral breasts in a prone position.
For a morphological contrast evaluation, a T2-weighted fast spin-echo axial FSE was used (slice thickness 3 mm; gap interslice 0.3 mm; matrix 512 × 512; FOV 32–40 cm). Short time inversion recovery axial STIR sequence suppressed the adipose tissue (slice thickness 3 mm; gap interslice 0.3 mm; matrix 512 × 512; FOV 32–40 cm). The dynamic contrast evaluation was based on an axial 3D dynamic T1 sequence, weighted by 1 basal acquisition and 5 post-contrast acquisitions of Gd-DTPA of 0.2 mmol/kg at a speed of 2 mL/s followed by 20 mL of saline solution. The 3T MR examination might include a sagittal 3D FSPGR sequence with fat-saturated imaging, post-contrast injection on the suspicious breast with shimming delimitation (slice thickness 1 mm; gap interslices 0 mm; matrix 360 × 352; FOV 32–40 cm). Diffusion imaging was performed on a single-shot axial echo-planar imaging (EPI) with a diffusion gradient amplitude on 3 octagonal axels and b-values of 0 and 800 s/mm (slice thickness 4 mm; gap interslices 0 mm; matrix 128 × 128; FOV 32–40 cm)

2.3. Image Analysis for FFDM, US and CESM

Baseline FFDM and US examinations for the reading session were repeated on study entrance. For the two radiologists, readings were performed on individual workstations, loaded with all cases, and calibrated in controlled ambient lighting conditions. Prior to studying the readings, a training session was held to familiarize radiologists with CESM and the reading protocol. Iodine-enhanced CESM images were reviewed by criteria of the American College of Radiology, Reston, VA, USA [12].
To reduce bias from image recall, for each study participant, readings occurred in two different sessions separated by a 1-month washout interval. Thus, for the same patient, the first reading session included images from FFDM + US and 3T MRI, whereas CESM-related images were evaluated in the second reading session. In both sessions, the two radiologists assessed the case, localized findings, assigned BI-RADS scores and completed and saved an electronic data form.
This form included patient identification, breast density on FFDM (BI-RADS scores of 1 to 4), location of each finding, type of finding (e.g., asymmetry, mass, scar/distortion), degree of confidence in the presence of each finding (5-step scale in which 1 = very low and 5 = very high), BI-RADS classification (1 to 5, BI-RADS scores of 0 and 6 not allowed). The electronic assessment was completed following the histological characterization of samples obtained by biopsy/surgery.

2.4. Statistical Analysis

Descriptive statistics were computed for all variables of interest. Continuous variables were reported as means and standard deviations, while categorical variables were reported as frequencies and percentage values. Sensitivity (Se), Sp, PPV and NPV were chosen to report on the accuracy of the diagnostic procedures with the scope to predict the nature and characteristics of the breast lesions under examination.
Lesion characteristics (lesion presence and location, finding type and size) on images obtained throughout the techniques were considered for potential associations with pathologic features, as assessed by cytology/histology.
The agreement between diagnostic tests and the gold standard, or CESM, was estimated by using raw agreement and the Cohen’s kappa statistics [13], interpreted with the Landis and Koch classification criteria [14]. Inter-reader variability was reported in terms of kappa statistics. The inter-observer agreement, diagnostic accuracy (DA), sensitivity (Se), specificity (Sp), positive predictive value (PPV) and negative predictive value (NPV) were calculated for all the imaging techniques included in our study. A p-value < 0.05 was considered statistically significant. All statistical analyses were performed with SPSS statistical software version 20 (SPSS Inc., Chicago, IL, USA).

3. Results

Overall, 118 female patients contributed data to our analysis. As per the study protocol, all patients were assessed by FFDM + US and MRI + CESM. Patient characteristics are shown in Table 2. Mean age at study entry was 48.5 (±9.7), with 79 of 118 patients menopausal. In 88 (74.6%) patients, breast histology revealed the neoplastic nature of the lesion/s assessed. Seven patients had the nature of the lesion assessed by cytology, of which three with malignant results were addressed by biopsy. In 70 (59.3%) patients, we found evidence of monofocal lesions (Figure 1), while 18 (20.5%) showed multifocal breast cancer (Figure 2), and 8 (6.8%) had bilateral breast cancer. In five (4.2%) patients, we only assessed benign lesions. Conversely, in 24 patients (20.3%), one benign lesion could coexist with monofocal breast cancer, but in 37 patients (31.4%), a benign lesion was associated with multifocal or bilateral breast cancer. A total of 57 of 88 patients underwent mastectomy, and 11, quadrantectomy.
Breast lesions’ characteristics are reported in Table 3. BI-RADS descriptors for each mode of assessment are listed as well. We observed 105 lesions with FFDM, 137 lesions with US, 110 lesions with CESM and 108 lesions with 3T MRI. The total number of lesions assessed by cytology, biopsy or surgery was 142. In 88 of 118 patients, breast cancer was ultimately diagnosed. In 62 patients (52.5%), we observed unifocal breast cancers, while in 18 (15.2%), we observed multifocal breast cancers, and in 8 (6.8), bilateral breast cancers were found (results reported in Table 2).
The results of concordance between CESM and 3T MR and the gold standard were encouraging. Table 4 summarizes inherent findings based on a “per lesion” analysis. Kappa values were 0.69, 0.81, 0.76 and 0.68 for concordance between FFDM and the gold standard, FFDM and CESM, and FFDM and US. CESM and 3T MRI showed a K concordance with the gold standard of 0.63 and 0.56, respectively. K concordance between CESM and the other modalities showed the following results: 0.81 with FFDM + US, 0.73 with MRI; p value ≤ 0.001 for all reported results. These same tests were performed according to a “per patient” approach analysis, showing a slightly lower K concordance between the gold standard, CESM and MRI (0.52 and 0.37, with a p value equal to 0.03 and 0.018, respectively) in monofocal breast cancers and a K concordance of 0.44 with a p value of 0.057 between MRI and the gold standard for multifocal and bilateral breast cancers. These results are extensively reported in Table 5.
Results on the diagnostic accuracy are reported in Table 6. Sensitivity (Se) was 0.99 for FFDM, 0.97 for US, 0.98 for the combination of FFDM + US, 0.99 for MRI and 1 for CESM. Specificity (Sp) was 0.61 for FFDM, 0.76 for US, 0.63 for the combination of FFDM + US, 0.47 for MRI and 0.5 for CESM. The PPV was 0.94 for FFDM, 0.92 for US, 0.89 for the combination of FFDM + US, 0.88 for MRI and 0.92 for CESM. The NPV was 0.89 for FFDM, 0.88 for US, 0.90 for the combination of FFDM + US, 0.9 for MRI and 1 for CESM. The DA was 0.93 for FFDM, 0.92 for US, 0.89 for the combination of FFDM + US, 0.88 for MRI and 0.93 for CESM. Results on the inter-observer agreement are shown in Table 7. The K values were 0.53 for FFDM, and 1.0 for US, MRI and CESM, respectively.

4. Discussion

CESM is increasingly emerging as an imaging technique with a key role in the diagnostic workup of patients affected by breast cancer [2]. When compared to MRI, some major strengths of CESM are represented by faster performance modality, lower costs and better tolerance, particularly in patients with physical limitations and/or claustrophobia [15]. The increasing interest toward CESM, with some encouraging data in the literature, prompted us to perform an open-label, single-center study of 118 patients in need of a diagnostic workup.
According to our study protocol, all patients enrolled underwent baseline assessment throughout FFDM + US. Subsequently, breast lesions were also studied by CESM and MRI, with cytological/histological assessment being the gold standard. Correlation analysis was performed according to a per patient and per lesion approach. Overall, CESM showed good concordance compared to the gold standard and the other modalities. When analyzing our data according to a per lesion approach, MRI, though maintaining a significant p value, tended to show a slightly lower K concordance. This aspect is evident in the per patient analysis, while concordance is not even significant when evaluating multifocal and bilateral breast cancers. This lack of concordance on MRI can be at least partly justified by the limits of this modality in terms of low specificity, which translates into a high rate of false positive cases [16]. Our results suggest that CESM could provide a compensation to MRI by means of a minor propensity to false positive cases. This aspect has been widely studied, as even recent work confirms encouraging results in terms of better specificity rates of CESM compared to MRI [17]. In a further attempt to improve specificity, an interesting study investigated the feasibility of CESM in combination with molecular imaging in the pre-surgical evaluation window. Data showed similar visualization of index cancers, with higher specificity for CESM compared to MRI. However, undergoing multimodal pre-surgical assessment was associated with discomfort in a relevant number of patients [18].
Patients undergoing CESM are exposed to radiation. However, recent literature shows a trend toward dose reduction, including those undergoing tomosynthesis. This would undoubtedly add to CESM advantages in terms of cost, time and patient tolerance [19,20].
The study herein presented does not include investigational tasks related to the combined use of digital breast tomosynthesis (DBT) and CESM. Indeed, the use of BDT in breast cancer diagnosis has recently grown, along with the detection rate of architecture distortion (AD). CESM may provide an effective tool in the evaluation of tomosynthesis-detected architectural distortion (AD). According to recently published data, the absence of AD lesion enhancement on CESM against a background of minimal or mild background parenchymal enhancement may indicate a lower risk of breast malignancy. The negative NPV reported by the authors is remarkably high, i.e., 100% in 26 (27.7%) of the 94 lesions with no suspicious enhancement on CESM, and encourages further investigation. It is our firm intention to include the use of digital breast tomosynthesis in combination with CESM in future studies within this same research pipeline [21].
Among the hardest and most intriguing challenges of CESM, its use in women with dense breasts deserves mentioning. In a pilot study of 318 women at increased breast cancer risk, CESM and MRI were both able to detect breast cancers not previously highlighted on conventional mammography. Neither did these two techniques differ by positive predictive value and specificity to a significant extent. However, given the still limited evidence, the authors carefully encourage the consideration of CESM in the evaluation of women at increased breast cancer risk who do not meet the criteria for MRI or for whom access to MRI is limited [22].
Results on DA, Se, Sp, PPV and NPV are consistent with those from the available literature [23]. We observed quite high values of Se and lower values of Sp, thus confirming that the findings related to vascularization may mimic suspicious lesions subsequently ascertained as benign in nature [24].
The results of inter-observer agreement show a considerable difference between FFDM and US, MRI and CESM. The not negligible number of advanced breast cancers (T3 and T4) resulting in extended lesions observed on the mammogram may have significantly contributed to this result. Indeed, in extended cancers, differentiating one single large lesion from multiple small lesions can be particularly difficult. Results from the second-level breast imaging techniques with contrast media significantly add to the most appropriate evaluation of tumor extension.
Our study has some limitations. As mentioned, two experienced radiologists were involved in the evaluation of imaging studies. We did not evaluate the background parenchymal enhancement (BPE) in our study, which is of interest, as it is one of the main aspects that affects MRI and seems to be less relevant in CESM [25,26]. Along with MRI, CESM is influenced by BPE, which is a crucial aspect with which the radiologists must be familiar, in order not to miss malignant enhancement; however, while BPE is markedly influenced by menstrual cycle in MRI studies, it seems than this aspect is less relevant when performing CESM [25,26] (Figure 3). The results are not transferrable to a screening population, as only patients with unresolved findings at US and FFMD were included. An issue of further relevance relates to the undesired side effects related to the exposure to a non-ionic iodinated contrast agent (Visipaque 320) in women undergoing CESM. Indeed, CESM advantages have to be balanced against the risk of contrast agent reactions, which, although generally low and overall acceptable, are greater than those associated with gadolinium [3].
However, CESM may be an alternative for patient populations at high risk who are unable to undergo MRI [27], using its applications as a guide in performing biopsies and radiomics applications [28] in the near future. Of further note, computer-aided detection (CAD) and artificial intelligence (AI) were not included in the study design at this stage of our research. Nevertheless, we are aware of the importance of comparing and eventually integrating results from humans with those from expert systems. In future studies, CAD and AI will surely be among our operative tasks [29,30].
We would also cite in this context of critical discussion feasibility issues that emerged over the conduct of this trial, with potential implication in terms of bias in radiology. When considered globally, the performance of the totality of the imaging techniques included in our trial generated a washout interval of about 1 month, which may have consequences in terms of bias from image recall. Errors, discrepancy and bias in radiology represent highly debated research themes, with increasingly emerging evidence on the potential causes and consequences [31]. Further attention will be devoted to these tasks in future studies from our team.

5. Conclusions

Emerging evidence supports the potential role of radiomics in supporting clinicians and surgeons in decision making, though larger samples of patients need to be investigated [32]. We provide evidence in support of CESM in the clinical routine and with regard to the diagnostic workup of breast lesions. In this context, CESM may represent a valuable alternative and/or an integrating technique to MRI in the evaluation of breast cancer patients. Compared to MRI, CESM has the advantage of low cost, speed and accessibility, and being better tolerated by patients. Our results represent good-quality scientific evidence, obtained from an ad hoc conceived study, which was carried out according to a prospective design. Methodological aspects related to the study sample size were taken into account, resulting in an overall study population comparable to the other available studies. Nevertheless, further evidence is warranted to confirm our findings and address aspects related to imaging techniques’ comparability and accuracy in well-characterized subsets of patients with specific features.

Author Contributions

Conceptualization, F.R.F.; methodology, M.B.; software, F.S. and I.T.; validation, P.V. and A.V.; formal analysis, F.V., M.B., I.T. and F.S.; investigation, F.R.F., F.G. and C.B.; resources, A.V.; data curation, F.V., M.B., I.T. and F.S; writing—original draft preparation, M.B. and F.R.F.; writing—review and editing, F.V. and M.B.; visualization, F.R.F. and F.V.; supervision, M.B.; project administration, M.B.; funding acquisition, A.V. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

The study was conducted according to the guidelines of the Declaration of Helsinki, and approved by the Institutional Review Board of IRCCS Regina Elena National Cancer Institute (protocol code RS 410/13 on 18 June 2013).

Informed Consent Statement

Informed consent was obtained from all subjects involved in the study. Written informed consent has been obtained from the patients to publish this paper.

Data Availability Statement

The data presented in this study are available on request from the corresponding authors.

Acknowledgments

We thank Federico Cappuzzo, Division of Medical Oncology 2, for his collaboration and the availability of professional figures from his Division. We also thank the radiology technician Franco Rea for providing support in the performance of FFDM and CESM.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Sung, H.; Ferlay, J.; Siegel, R.L.; Laversanne, M.; Soerjomataram, I.; Jemal, A.; Bray, F. Global cancer statistics 2020: Globocan estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J. Clin. 2021, 71, 209–249. [Google Scholar] [CrossRef] [PubMed]
  2. Bhimani, C.; Matta, D.; Roth, R.G.; Liao, L.; Tinney, E.; Brill, K.; Germaine, P. Contrast-enhanced spectral mammography: Technique, indications, and clinical applications. Acad. Radiol. 2017, 24, 84–88. [Google Scholar] [CrossRef] [PubMed]
  3. James, J.J.; Tennant, S.L. Contrast-enhanced spectral mammography (CESM). Clin. Radiol. 2018, 73, 715–723. [Google Scholar] [CrossRef] [PubMed]
  4. Mori, M.; Akashi-Tanaka, S.; Suzuki, S.; Daniels, M.I.; Watanabe, C.; Hirose, M.; Nakamura, S. Diagnostic accuracy of contrast-enhanced spectral mammography in comparison to conventional full-field digital mammography in a population of women with dense breasts. Breast Cancer 2017, 24, 104–110. [Google Scholar] [CrossRef]
  5. Tagliafico, A.S.; Bignotti, B.; Rossi, F.; Signore, A.; Sormani, M.P.; Valdora, F.; Calabrese, M.; Houssami, N. Diagnostic performance of contrast-enhanced spectral mammography: Systematic review and meta-analysis. Breast 2016, 28, 13–19. [Google Scholar] [CrossRef] [PubMed]
  6. Barra, F.R.; de Souza, F.F.; Camelo, R.E.F.A.; de Oliveira Ribeiro, A.C.; Farage, L. Accuracy of contrast-enhanced spectral mammography for estimating residual tumour size after neoadjuvant chemotherapy in patients with breast cancer: A feasibility study. Radiol. Bras. 2017, 50, 224–230. [Google Scholar] [CrossRef] [Green Version]
  7. Patel, B.K.; Garza, S.A.; Eversman, S.; Lopez-Alvarez, Y.; Kosiorek, H.; Pockaj, B.A. Assessing tumour extent on contrast-enhanced spectral mammography versus full-field digital mammography and ultrasound. Clin. Imaging 2017, 46, 78–84. [Google Scholar] [CrossRef]
  8. Lobbes, M.B.; Lalji, U.; Houwers, J.; Nijssen, E.C.; Nelemans, P.J.; van Roozendaal, L.; Smidt, M.L.; Heuts, E.; Wildberger, J.E. Contrast-enhanced spectral mammography in patients referred from the breast cancer screening programme. Eur. Radiol. 2014, 24, 1668–1676. [Google Scholar] [CrossRef]
  9. Suter, M.F.; Pesapane, F.; Agazzi, G.M.; Gagliardi, T.; Nigro, O.; Bozzini, A.; Priolo, F.; Penco, S.; Cassano, E.; Chini, C.; et al. Diagnostic accuracy of contrast-enhanced spectral mammography for breast lesions: A systematic review and meta-analysis. Breast. 2020, 53, 8–17. [Google Scholar] [CrossRef]
  10. Evans, D.G.; Maxwell, A.J. MRI screening in women with a personal history of breast cancer. J. Natl. Cancer Inst. 2016, 108, 2–3. [Google Scholar] [CrossRef] [Green Version]
  11. Lima, Z.S.; Ebadi, M.R.; Amjad, G.; Younesi, L. Application of imaging technologies in breast cancer detection: A review article. Open Access Maced J. Med. Sci. 2019, 7, 838–848. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  12. D’Orsi, C.J.; Sickles, E.A.; Mendelson, E.B.; Morris, E. ACR BI-RADS® Atlas, Breast Imaging Reporting and Data System; Committee on BI-RADS American College of Radiology 1891; Preston White Drive: Reston, VA, USA, 2013. [Google Scholar]
  13. Cohen, J. Weighted kappa: Nominal scale agreement provision for scaled disagreement or partial credit. Psychol. Bull. 1968, 70, 213–220. [Google Scholar] [CrossRef] [PubMed]
  14. Landis, J.R.; Koch, G.G. The measurement of observer agreement for categorical data. Biometrics 1977, 3, 159–174. [Google Scholar] [CrossRef] [Green Version]
  15. Hobbs, M.M.; Taylor, D.B.; Buzynski, S.; Peake, R.E. Contrast enhanced spectral mammography (CESM) and contrast-enhanced MRI (CEMRI): Patient preferences and tolerance. J. Med. Imaging Radiat. Oncol. 2015, 59, 300–305. [Google Scholar] [CrossRef]
  16. Fallenberg, E.M.; Schmitzberger, F.F.; Amer, H.; Ingold-Heppner, B.; Balleyguier, C.; Diekmann, F.; Dromain, C. Contrast-enhanced spectral mammography vs. mammography and MRI—Clinical performance in a multi-reader evaluation. Eur. Radiol. 2016, 27, 2752–2764. [Google Scholar] [CrossRef]
  17. Clauser, P.; Baltzer, P.A.T.; Kapetas, P.; Hoernig, M.; Weber, M.; Leone, F.; Bernathova, M.; Helbich, T.H. Low-dose, contrast-enhanced mammography compared to contrastenhanced breast MRI: A feasibility study. J. Magn. Res. Imaging 2020, 52, 589–595. [Google Scholar] [CrossRef]
  18. Sumkin, J.H.; Berg, W.A.; Carter, G.; Bandos, A.I.; Chough, D.M.; Ganott, M.A.; Hakim, C.M.; Kelly, A.E.; Zuley, M.L.; Houshmand, G.; et al. Diagnostic performance of MRI, molecular breast imaging and contrast–enhanced mammography in women with newly diagnosed breast cancer. Radiology 2019, 293, 531–540. [Google Scholar] [CrossRef]
  19. Phillips, J.; Mihai, G.; Hassonjee, S.E.; Raj, S.D.; Palmer, M.R.; Brook, A.; Zhang, D. Comparative dose of contrast enhanced spectral mammography (CESM), digital mammography, and digital breast Tomosynthesis. Am. J. Roentgenol. 2018, 211, 839–846. [Google Scholar] [CrossRef]
  20. James, J.R.; Pavlicek, W.; Hanson, J.A.; Boltz, T.F.; Patel, B.K. Breast radiation dose with CESM compared with 2D FFDM and 3D tomosynthesis mammography. Am. J. Roentgenol. 2017, 208, 362–372. [Google Scholar] [CrossRef]
  21. Goh, Y.; Chan, C.W.; Pillay, P.; Lee, H.S.; Pan, H.; Hung, B.; Quek, S.T.; Chou, C. Architecture distortion score (ADS) in malignancy risk stratification of architecture distortion on contrast-enhanced digital mammography. Eur. Radiol. 2021, 31, 2657–2666. [Google Scholar] [CrossRef]
  22. Jochelson, M.S.; Pinker, K.; Dershaw, D.D.; Hughes, M.; Gibbons, G.F.; Rahbar, K.; Robson, M.E.; Mangino, D.A.; Goldman, D.; Moskowitz, C.S.; et al. Comparison of screening CEDM and MRI for women at increased risk for breast cancer: A pilot study. Eur. J. Radiol. 2017, 97, 37–43. [Google Scholar] [CrossRef] [PubMed]
  23. Bhavika, K.; Patel, K.; Lobbes, M.B.; Lewin, J. Contrast Enhanced Spectral Mammography: A Review. In Seminars in Ultrasound, CT and MRI; WB Saunders: Philadelphia, PA, USA, 2018. [Google Scholar]
  24. Wanqing, X.; Haiying, R.; Liyu, Z. A meta-analysis of contrast-enhanced spectral mammography versus MRI in the diagnosis of breast cancer. Thorac. Cancer 2020, 11, 1423–1432. [Google Scholar]
  25. Sogani, J.; Morris, E.A.; Kaplan, J.B.; D’Alessio, D.; Goldman, D.; Moskowitz, C.S.; Jochelson, M.S. Comparison of background parenchymal enhancement at contrast-enhanced spectral mammography and breast MR imaging. Radiology 2017, 282, 63–73. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  26. Savaridas, S.L.; Taylor, D.B.; Gunawardana, D.; Phillips, M. Could parenchymal enhancement on contrast-enhanced spectral mammography (CESM) represent a new breast cancer risk factor? Correlation with known radiology risk factors. Clin. Radiol. 2017, 72, 1085.e1–1085.e9. [Google Scholar] [CrossRef] [PubMed]
  27. Sogani, J.; Mango, V.L.; Keating, D.; Sung, J.S.; Jochelson, M.S. Contrast-enhanced mammography: Past, present, and future. Clin. Imaging 2021, 69, 269–279. [Google Scholar] [CrossRef]
  28. Marino, M.A.; Pinker, K.; Leithner, D.; Sung, J.; Avendano, D.; Morris, E.A.; Jochelson, M. Contrast-enhanced mammography and radiomics analysis for noninvasive breast cancer characterization: Initial results. Mol. Imaging Biol. 2020, 22, 780–787. [Google Scholar] [CrossRef]
  29. Chan, H.P.; Samala, R.K.; Hadjiiski, L.M. CAD and AI for breast cancer-recent development and challenges. Br. J. Radiol. 2020, 93, 20190580. [Google Scholar] [CrossRef]
  30. Patel, B.K.; Ranjbar, S.; Wu, T.; Pockaj, B.A.; Li, J.; Zhang, N.; Lobbes, M.; Zhang, B.; Mitchell, J.R. Computer-aided diagnosis of contrast-enhanced spectral mammography: A feasibility study. Eur. J. Radiol. 2018, 98, 207–213. [Google Scholar] [CrossRef]
  31. Onder, O.; Yarasir, Y.; Azizova, A. Errors, discrepancies and underlying bias in radiology with case examples: A pictorial review. Insights Imaging 2021, 12, 51. [Google Scholar] [CrossRef]
  32. La Forgia, D.; Fanizzi, A.; Campobasso, F.; Bellotti, R.; Didonna, V.; Lorusso, V.; Moschetta, M.; Massafra, R.; Tamborra, P.; Tangaro, S.; et al. Radiomic analysis in contrast-enhanced spectral mammography for predicting breast cancer histological outcome. Diagnostics 2020, 10, 708. [Google Scholar] [CrossRef]
Figure 1. Local advanced breast cancer of the upper outer quadrant of the right breast; FFDM shows an area of hyperdensity (a); US confirmed a hypoechoic lesion (b), while MRI and CESM showed more extensive disease (c,d).
Figure 1. Local advanced breast cancer of the upper outer quadrant of the right breast; FFDM shows an area of hyperdensity (a); US confirmed a hypoechoic lesion (b), while MRI and CESM showed more extensive disease (c,d).
Cancers 14 01351 g001
Figure 2. Local advanced breast cancer of the left breast. FFDM (a) does not allow for detection of the real disease extension of the lesion, in relation to breast density. MRI (b) and CESM (c) show multiple foci in relation to multicentric disease.
Figure 2. Local advanced breast cancer of the left breast. FFDM (a) does not allow for detection of the real disease extension of the lesion, in relation to breast density. MRI (b) and CESM (c) show multiple foci in relation to multicentric disease.
Cancers 14 01351 g002
Figure 3. FFDM shows an irregular opacity in the inner lower quadrant of the right breast (a); US images show an irregular hypoechoic nodule in the inner lower quadrant of the right breast, with some suspicious smaller images close to the bigger one (b). CESM images confirm an irregular contrast uptake nodule in the inner lower quadrant of the right breast (c). Compared to the MRI, the background enhancement is still present but milder (d).
Figure 3. FFDM shows an irregular opacity in the inner lower quadrant of the right breast (a); US images show an irregular hypoechoic nodule in the inner lower quadrant of the right breast, with some suspicious smaller images close to the bigger one (b). CESM images confirm an irregular contrast uptake nodule in the inner lower quadrant of the right breast (c). Compared to the MRI, the background enhancement is still present but milder (d).
Cancers 14 01351 g003
Table 1. Minimum (min), maximum (max) and mean dose exposure for FFDM and CESM.
Table 1. Minimum (min), maximum (max) and mean dose exposure for FFDM and CESM.
Dose Glandular Exposure
FFDMCESM
Min0.68 mGy1.22 mGy
Max1.32 mGy2.28 mGy
Mean dose1.04 mGy (±0.05)1.75 mGy (±0.07)
Table 2. Characteristics of the study participants (N = 118).
Table 2. Characteristics of the study participants (N = 118).
Study Participants CaracteristicsN (%)
Age at study entry (years) (mean ± SD)48.5 ± 9.7
Patients with multifocal breast cancer (*)
No70 (79.5)
Yes18 (20.5)
Patients with bilateral breast lesions
No94 (79.7)
Yes24 (20.3)
Bilateral breast cancer
No110 (93.2)
Yes8 (6.8)
Multifocal and/or bilateral breast lesions
No81 (68.6)
Yes37 (31.4)
Cytological/Histological assessment Right breastN = 66 (55.9)
No cancer27 (40.9)
Cancer39 (59.1)
Cytological/Histological evaluation Left breastN = 76 (64.4)
No cancer17 (22.4)
Cancer59 (77.6)
* Only cancer cases N = 88.
Table 3. Characteristics of the breast lesions by diagnostic test in 118 study participants.
Table 3. Characteristics of the breast lesions by diagnostic test in 118 study participants.
BI-RADSN (%)
FFDM Total lesions105
Benign1 (1.0)
Probably benign–suspect <2%14 (13.3)
Low suspect <10%21 (20.0)
Intermediate suspect 10–50%24 (22.9)
High suspect 50–95%12 (11.4)
Malignant >95%33 (31.4)
US Total lesions137
Benign5 (3.6)
Probably benign–suspect <2%31 (22.6)
Low suspect <10%12 (8.8)
Intermediate suspect 10–50%14 (10.2)
High suspect 50–95%23 (16.8)
Malignant >95%52 (38.0)
CESM Total lesions110
Probably benign–suspect <2%10 (9.1)
Low suspect <10%10 (9.1)
Intermediate suspect 10–50%17 (15.5)
High suspect 50–95%27 (24.5)
Malignant >95%46 (41.8)
MRI Total lesions108
Benign4 (3.7)
Probably benign–suspect <2%13 (12.0)
Low suspect <10%10 (9.3)
Intermediate suspect 10–50%3 (2.8)
High suspect 50–95%29 (26.8)
Malignant >95%49 (45.4)
FFDM: full-field digital mammography; CESM: contrast-enhanced spectral mammography; US: ultrasound; MRI: magnetic resonance imaging.
Table 4. Concordance analysis between imaging techniques. Per lesion approach (N: 142).
Table 4. Concordance analysis between imaging techniques. Per lesion approach (N: 142).
OverallK Qualitative Scale
Gold StandardCESM<0.01No Agreement
FFDM0.69 p < 0.0010.81 p < 0.0010.01–0.20Scarce
US0.76 p < 0.001_0.21–0.40Low
FFDM + US0.68 p < 0.0010.81 p < 0.0010.41–0.60Moderate
CESM0.63 p < 0.001_0.61–0.80High/Good
MRI0.56 p < 0.0010.73 p < 0.0010.81–1.00Excellent
Table 5. Concordance analysis between imaging techniques. Per patient approach (N: 118).
Table 5. Concordance analysis between imaging techniques. Per patient approach (N: 118).
Right BreastIn NON-Multifocal and/or Bilateral LesionsIn Multifocal and/or Bilateral Lesions
Gold StandardCESMGold StandardCESM
FFDM0.70 p < 0.0010.87 p < 0.0011.00 p < 0.0011.00 p < 0.001K Qualitative Scale
US0.70 p < 0.001_0.74 p = 0.001_<0.01No agreement
FFDM + US0.61 p < 0.0010.87 p < 0.0010.74 p = 0.0011.00 p < 0.0010.01–0.20Scarce
CESM0.52 p = 0.003_0.77 p = 0.001_0.21–0.40Low
MRI0.37 p = 0.0180.63 p = 0.0010.44 p = 0.0570.63 p = 0.0110.41–0.60Moderate
Left breastIn Non-Multifocal and/or bilateral lesionsIn Multifocal and/or bilateral lesions0.61–0.80High/Good
Gold standardCESMGold standardCESM0.81–1.00Excellent
FFDM0.62 p < 0.0010.64 p < 0.001NANA
US0.78 p < 0,001_0.86 p < 0.001_
FFDM + US0.64 p < 0.0010.72 p < 0.0010.83 p < 0.0010.65 p = 0.001
CESM0.64 p < 0.001_0.64 p = 0.002_
MRI0.71 p < 0.0010.79 p < 0.0011.00 p < 0.0011.00 p < 0.001
Table 6. Results for diagnostic accuracy (DA), sensitivity (SV), specificity (Sp), positive predictive value (PPV) and negative predictive value (VPN) for each modality and the gold standard.
Table 6. Results for diagnostic accuracy (DA), sensitivity (SV), specificity (Sp), positive predictive value (PPV) and negative predictive value (VPN) for each modality and the gold standard.
Gold StandardSeSpPPVNPVDA
PositiveNegative
FFDMPositive7750.990.610.940.890.93
Negative18
USPositive8770.970.760.920.880.92
Negative322
US + FFDMPositive91110.980.630.890.900.89
Negative219
MRIPositive74100.990.470.880.90.88
Negative19
CESMPositive87710.50.9210.93
Negative07
Table 7. Results on the inter-observer agreement calculated on the findings of the two readers.
Table 7. Results on the inter-observer agreement calculated on the findings of the two readers.
KAPPAp-Value95% CI
FFDM0.530.0240.19–0.88
US1.00<0.0011.00–1.00
MRI1.00<0.0011.00–1.00
CESM1.00<0.0011.00–1.00
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Ferranti, F.R.; Vasselli, F.; Barba, M.; Sperati, F.; Terrenato, I.; Graziano, F.; Vici, P.; Botti, C.; Vidiri, A. Diagnostic Accuracy of Contrast-Enhanced, Spectral Mammography (CESM) and 3T Magnetic Resonance Compared to Full-Field Digital Mammography plus Ultrasound in Breast Lesions: Results of a (Pilot) Open-Label, Single-Centre Prospective Study. Cancers 2022, 14, 1351. https://0-doi-org.brum.beds.ac.uk/10.3390/cancers14051351

AMA Style

Ferranti FR, Vasselli F, Barba M, Sperati F, Terrenato I, Graziano F, Vici P, Botti C, Vidiri A. Diagnostic Accuracy of Contrast-Enhanced, Spectral Mammography (CESM) and 3T Magnetic Resonance Compared to Full-Field Digital Mammography plus Ultrasound in Breast Lesions: Results of a (Pilot) Open-Label, Single-Centre Prospective Study. Cancers. 2022; 14(5):1351. https://0-doi-org.brum.beds.ac.uk/10.3390/cancers14051351

Chicago/Turabian Style

Ferranti, Francesca Romana, Federica Vasselli, Maddalena Barba, Francesca Sperati, Irene Terrenato, Franco Graziano, Patrizia Vici, Claudio Botti, and Antonello Vidiri. 2022. "Diagnostic Accuracy of Contrast-Enhanced, Spectral Mammography (CESM) and 3T Magnetic Resonance Compared to Full-Field Digital Mammography plus Ultrasound in Breast Lesions: Results of a (Pilot) Open-Label, Single-Centre Prospective Study" Cancers 14, no. 5: 1351. https://0-doi-org.brum.beds.ac.uk/10.3390/cancers14051351

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop