Next Article in Journal
Mathematical Modeling of Solutes Migration under the Conditions of Groundwater Filtration by the Model with the k-Caputo Fractional Derivative
Previous Article in Journal
Minkowski Dimension and Explicit Tube Formulas for p-Adic Fractal Strings
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Fatigue-Induced Cole Electrical Impedance Model Changes of Biceps Tissue Bioimpedance

Department of Electrical and Computer Engineering, The University of Alabama, Tuscaloosa, AL 35487, USA
*
Author to whom correspondence should be addressed.
Submission received: 13 September 2018 / Revised: 17 October 2018 / Accepted: 22 October 2018 / Published: 24 October 2018
(This article belongs to the Special Issue Fractional Behavior in Nature)

Abstract

:
Bioimpedance, or the electrical impedance of biological tissues, describes the passive electrical properties of these materials. To simplify bioimpedance datasets, fractional-order equivalent circuit presentations are often used, with the Cole-impedance model being one of the most widely used fractional-order circuits for this purpose. In this work, bioimpedance measurements from 10 kHz to 100 kHz were collected from participants biceps tissues immediately prior and immediately post completion of a fatiguing exercise protocol. The Cole-impedance parameters that best fit these datasets were determined using numerical optimization procedures, with relative errors of within approximately ± 0.5 % and ± 2 % for the simulated resistance and reactance compared to the experimental data. Comparison between the pre and post fatigue Cole-impedance parameters shows that the R , R 1 , and f p components exhibited statistically significant mean differences as a result of the fatigue induced changes in the study participants.

1. Introduction

Bioimpedance, or the electrical impedance of biological tissues, describes the passive electrical properties of these materials. The aim of collecting these measurements is to provide details regarding the electrochemical structures and processes within a tissue or material under study [1]. These measurements are being used in a wide range of applications including: As a method to monitor hydration during hemodialysis [2], to detect changes resulting from muscle injury [3], assessing lympoedema [4], and to assess neuromuscular disorders [5]. There is also a small set of studies that have investigated the bioimpedance of skeletal muscle for changes that result from fatigue [6,7,8].
Regardless of the application, after collecting an impedance dataset (which may contain hundreds of data points at different frequencies) the data needs to be analyzed using either discrete frequencies or electrical equivalent circuit representations. Using equivalent electrical circuits can reduce a dataset from the potentially hundreds of datapoints to a smaller set (dependent on the number of circuit components in the model). This reduction is aimed at decreasing the complexity of tracking changes in the measured tissues. While there are many equivalent circuits that have been used to represent bioimpedance datasets [9], one of the most widely used is referred to as the Cole-impedance model [10], given in Figure 1. This model is composed of three hypothetical circuit elements. A high-frequency resistor R , a resistor R 1 , and a Constant Phase Element (CPE). The impedance of a CPE is Z C P E = 1 / C s α where s α = ( j ω ) α = ω α cos ( α π / 2 ) + j sin ( α π / 2 ) . The value C is often referred to as a pseudo-capacitance with units Farad sec ( α 1 ) . The impedance of the Cole-model is given by:
Z ( s ) = R + R 1 1 + s α R 1 C
Recent studies have utilized the Cole-impedance model in biologic applications including: Characterizing the electrical impedance of blood for potential monitoring of thrombus formation [11], predicting myofiber size in mice [12], modeling the electrical impedance of botanical elements [13]. While this is not an exhaustive list, it does highlight the varied applications of the Cole-impedance model in biology. It is important to note that using a CPE (with the s α term) implies a fractional derivative of order 0 < α < 1 , with the current-voltage relationship for this component defined as:
i ( t ) = C d α v ( t ) d t α
where i ( t ) and v ( t ) are the time-dependent current and voltage, respectively. This therefore places the Cole-impedance model within the domain of fractional-order systems. This field incorporates concepts from fractional calculus [14,15], the branch of mathematics concerning non-integer differentiation and integration, and is being widely explored to model biological phenomena [9,16,17,18]. One definition of a fractional derivative of order α is given by the Grünwald-Letnikov definition [19] as
a D α f ( x ) = lim h 0 1 h α m = 0 x a h 1 m Γ α + 1 m ! Γ α m + 1 f x m h
where Γ ( · ) is the gamma function and n 1 α n . Applying the Laplace transform to the fractional derivative of (3) with zero initial conditions yields
L D t α f ( t ) = s α F ( s )
though other definitions, such as the Riemann-Liouville and Caputo definitions, are also available. While the fractional derivative given by (3) is not often referenced in studies of bioimpedance using the Cole-impedance model, it is still important to understand the underlying fractional-theory behind it.
When calculating (2) using sinusoidal voltages, for it to yield sinusoids requires the fractional derivative to be defined on the whole real line, as is typical of many bioimpedance applications. Therefore, the Grünwald-Letnikov, Riemann-Liouville, and Caputo on R definitions of fractional derivatives must be used [20]. This is important for impedance applications that are collected in experimental setups that apply a sinusoidal excitation signal and measure the corresponding sinusoidal output. For these cases, the correct definition must be used or else the fractional-order impedance model and experimental results will contradict each other. Fractional derivatives, upon which the Cole-impedance model is built, capture phenomena across multiple time scales overcoming the need to define tissue properties at the cellular level. Instead, these models assume that the behaviour is captured in the fractal structure of the tissue [16,17]. Recent research has shown the relationship between fractals and fractional calculus, based on physical and geometric considerations [21]. It was noted in [17] that muscle fibers, tendon, and nerve fibers exhibit patterns that support the dynamics of these multiscale structures expressed by fractional-order models. While these fractional-models fail to describe the underlying physiological mechanisms at the unit or cellular level that contribute to the impedance behavior of a tissue, recent works using fractional-order models to represent three-dimensional resistor-capacitor ( R C ) networks [22] support their application to model biological tissues; which are complex 3D structures of cells with both resistive and capacitive behaviour. Often, the bioimpedance of a tissue is described as being related to the extracellular fluid and intracellular fluid, with measurements at low frequency dependent on the extracellular fluids (attributed to low excitation frequencies failing to penetrate the cellular membranes of cells in the tissue) and high frequency measurements dependent on both intracellular and extracellular fluids (at frequencies where the excitation is able to penetrate the cellular membranes). Based on this interpretation, the model parameters R and R 1 of the Cole-impedance model are associated with the tissue fluids and C, α are associated with the cellular membranes of the tissues.
While it has been previously shown that the bioimpedance at discrete frequencies (10 kHz, 50 kHz, and 100 kHz) of the biceps tissue does change due to exercise-induced fatigue [8] and that the biceps tissue bioimpedance can be well represented by the Cole-impedance model [23]; the changes in the Cole-impedance parameters that result from fatigue have not been deeply investigated. This study evaluates the parameters of the Cole-model equivalent electrical circuit that can represent bioimpedance datasets collected from the biceps tissues of participants immediately prior to and immediately post completion of a fatiguing exercise protocol; expanding on those analyses presented in [8,23]. In this work, the Cole-impedance parameters ( R , R 1 , C, α ) that represent the tissue bioimpedance of the left and right arms of 18 participants, separated into two different exercise intensity groups, are determined using numerical least squares optimization routines. These results of these optimization analyses are presented with statistical tests comparing the pre and post-fatigue measures executed to evaluate those changes that were statistically significant with discussions of these results and their implications.

2. Methods

2.1. Study Participants

This study collected measurements from the left and right biceps of 18 participants (15 males, 3 females, with an average age of 22.2 ± 3.2 years) immediately prior to and immediately after completing an exercise protocol to fatigue the biceps brachii muscles. This study was reviewed and approved by the institutional review board of The University of Alabama (16-OR-234). All study participants were recruited from The University of Alabama and were screened prior to participation using a health questionnaire. Those participants who had any reported muscle or joint problems or who had any recent adverse reactions to exercise were excluded. Informed written consent was obtained from each participant prior to their inclusion in the study.

2.2. Exercise Protocol

Each participant executed a fatiguing exercise protocol using dumbbell biceps curls at either 60 % ( N = 10 ) or 75 % ( N = 8 ) of their previously assessed one-repetition maximum (1-RM) until task failure. The 1-RM value for each participant is the maximum weight that they were able to successfully lift for the dumbbell biceps curl exercise. The grouping of participants by intensity relative to their 1-RM weight was introduced to control for differences between the relative strengths of each participant, which could have an impact on the exercise induced changes. During this protocol, participants completed repetitions of the exercise until failure (this series of repetitions will be referred to as one set). After a two-minute rest the participants completed an additional set. This process continued until the participants had completed ten sets at which point the protocol was terminated.

2.3. Localized Electrical Impedance Measurements

The localized electrical impedance measurements were collected from each participant using a tetrapolar electrode configuration in which two current injection electrodes (I+, I−) were placed on the lateralis side of the biceps, 14 cm apart with the voltage sensing electrodes (V+, V−) placed 4.67  cm apart. This electrode configuration is given in Figure 2. A fixed distance electrode configuration was used in this work over a configuration relative to each participants biceps dimensions to emulate expected configurations if this technique is integrated into wearable systems not tailored to each participants specific dimensions. For all collected measurements, participants were in a standing position and asked to relax their muscles with their arms resting naturally at the sides of their body. The impedance measures immediately prior to completed the protocol and immediately post-completion were collected using a Keysight E4990A impedance analyzer (Keysight Technologies: Santa Rosa, CA, USA) with a custom-interface to adapt the BNC-connectors of the E4990A to a cable-set with the required snap connectors for the Ag/AgCl electrodes. Using this setup, which is detailed in Figure 2, measurements were collected from 10 kHz to 100 kHz at 67 logarithmically spaced frequencies. This frequency range was selected because it is widely employed in bioimpedance applications [6]. It should be noted that before data collection, the Keysight E4990A was calibrated using the open/short/load procedure with the developed interface. Further details of this interface and its accuracy evaluation are detailed in [8]. Examples of the impedance collected from the left and right biceps of two participants are given in Figure 3 as solid lines, with the pre-fatigue values given as black lines and the post-fatigue values as blue lines.

2.4. Nonlinear Least Squares Fitting

The Cole-impedance parameters ( R , R 1 , C, α ) that best fit the model given by (1) for each collected dataset in this study were determined using a nonlinear least squares fitting (NLSF) procedure. This procedure iteratively attempts to numerically solve the problem
min x f 0 ( x ) = j = 1 n Z ( x , ω j ) y j 2 s . t . x > 0
where x is the set of impedance parameters ( R , R 1 , C, α ) to minimize f o ( x ) (the objective function), Z ( x , ω j ) is the impedance of (1) at frequency ω j , y j is the collected impedance at frequency ω j , and n is the total number of data points in the collected response. A constraint is added to the problem to limit the possible solutions to real values; because negative resistance and capacitance values are not physically possible. However, because the impedance is a complex value, many numerical solvers are not able to solve this problem and it requires that the objective function given in (5) be revised to separate and fit the real and imaginary components of the impedance. Applying this separation to (5) yields:
min x f 0 ( x ) = j = 1 n { Z ( x , ω j ) y j } + { Z ( x , ω j ) y j } 2 s . t . x > 0
where { · } and { · } denote the real and imaginary components, respectively. There are a variety of solvers available for this class of problem, which include the MATLAB optimization functions (lsqcurvefit, fmincon, fminsearch). The least squares method is not the only optimization method available for this class of problem; methods including particle swarm optimization and other biologically inspired algorithms having been investigated [24,25,26]. Now while those alternative methods have shown very good results in terms of both accuracy and execution speed [26], the least squares method is applied in this work because of its widely available MATLAB implementation.
Algorithm 1 Sample MATLAB Code to Extract Cole-Impedance Parameters from Experimental Impedance Data
  • % Requires:   
  • %          Experimental impedance (Ohms) stored in variable Z   
  • %          Frequency range (Hz) of Z stored in variable f    
  •       
  • % Step 1: Setup function (ColeZ) to generate Cole-impedance  
  • % P(1) → R_\infty, P(2) → R1, P(3) → C, P(4) → alpha  
  •      w = 2 ∗ pi ∗ f;  
  •      ColeZ = @(P) (P(1) + P(2)./(1 + (1i ∗ w). P(4) ∗ P(2) ∗ P(3)));   
  •       
  • % Step 2: Setup objective function to minimize (Least Squares)  
  •      objfun = @(P) sum((real(Z) − real(ColeZ(P))). 2 + (imag(Z) − imag(ColeModel(P))). 2);   
  •       
  • % Step 3: Setup constraints for fmincon   
  •      A = []; b = []; Aeq = []; beq = []; nonlcon = [];  
  •      lb = [0, 0, 0, 0];                                 % Lower search boundary  
  •      ub = [inf, inf, inf, inf];                    % Upper search boundary  
  •      options = optimoptions(’fmincon’);   
  •       
  • % Step 4: Generate random initial iterates for solver  
  •      InitialIterates = 100;  
  •      x0 = zeros(InitialIterates, 4);  
  •      lb2 = [1 1 −9 0.5];  
  •      for k = 1:1:4  
  •             x0(:,k) = lb2(k) + (ub2(k)-lb2(k)). ∗ rand(InitialIterates,1);  
  •      end   
  •      x0(:,3) = 10. x0(:,3);    
  •       
  • % Step 5: Execution of optimization search for Cole parameters  
  •      MinError = inf  
  •      for m = 1:1:InitialIterates  
  •            [ColeParameters, ObjFunValue]=fmincon(objfun, x0, A, b, Aeq, beq, lb, ub, nonlcon, options);  
  •            if (ObjFunValue < MinError)  
  •                    BestFitCole = ColeParameters;  
  •                    MinError = ObjFunValue;  
  •            end  
  •      end
The NLSF is susceptible to the local minima problem, where a solution that is not the global solution may meet the stopping criteria of the solver. To overcome this challenge, the solver was applied 100 times to the simulated dataset using a different initial iterate for each execution. This method was previously implemented in [27], with the execution that yielded the lowest objective function selected as the global solution. Each of the initial iterates were randomly generated within the ranges 1 Ω < R 0 , 1 < 50 Ω , 0.1 nF < C < 100 μ F , and 0.5 < α < 1 . The intent of this approach is for at least one of the generated initial iterates to be very close to the global solution, which improves the likelihood of returning the global solution and avoiding a local minima when the ending criteria of the solver are satisfied. A sample of MATLAB code (Algorithm 1) is presented to detail this setup of the required functions to use the fmincon function. This implementation takes advantage of the inbuilt MATLAB functions real and imag to separate the real and imaginary components of both experimental measurements and the Cole-impedance model; which eliminates the need to determine the algebraic representations of the real and imaginary components of (1) to implement the objective function to minimize.

3. Results

The complete set of Cole-impedance parameters extracted from the pre- and post-fatigue measurements of participants in the 60 % and 75 % exercise-intensity groups are given in Table 1 and Table 2, respectively. To visualize the agreement between simulations using the extracted parameters and the measured impedance, two representative cases are provided in Figure 3 for participants 0007 and 0018. These particular datasets were selected because they represent the best and worst fittings, as evaluated by the average absolute error of the simulations compared to the experimental data. The simulations using the Cole-impedance parameters are presented in Figure 3 as dashed lines, with the experimental data presented as solid lines. The Cole-impedance model was simulated from 1 mHz to 100 MHz to highlight the impedance-arc which is not fully captured in the experimental data frequency range of 10 kHz to 100 kHz. This visualizes how using the Cole-impedance model parameters supports extrapolating the theoretical low and high frequency impedance values.
Additionally, the frequency at which the reactance reaches its peak value ( f p = ω p / 2 π = 1 / 2 π ( R 1 C ) ( 1 / α ) ) was determined using the extracted Cole-impedance model parameters for the left and right arms of each participant. These values are also given in Table 1 and Table 2. Samples of these calculated peak reactances using the values in Table 1 and Table 2 are given in Figure 3 as (o) symbols.
To fully detail the agreement between the experimental data and simulations using the extracted circuit parameters, relative error distributions of the resistance and reactance of both groups of participants are given in Figure 4. From these distributions, the relative errors of the resistances (given in Figure 4a,c) are predominately grouped between ± 0.5 % , while the reactances (given in Figure 4b,d) show a wider grouping between ± 2 % . This supports that the fitting process achieves better agreement with the resistance data than the reactance data, but both still show very good agreement with the experimental data.

Statistical Tests

Paired-samples t-tests (SPSS, IBM Inc.) were used to determine whether there was a statistically significant ( p < 0.05 ) difference between the mean values of the Cole-impedance model parameters extracted from the pre-fatigue and post-fatigue datasets for both the 60 % and 75 % groups of participants. These tests were applied to those datasets that did not violate the assumption of normality as assessed by the Shapiro-Wilk’s test. Based on these tests, there were statistically significant differences between the pre/post fatigue R ( p < 0.0005 , p = 0.001 ), R 1 ( p < 0.0005 , p < 0.0005 ), and f p ( p < 0.0005 , p = 0.008 ) means for both the 60 % and 75 % groups. There were no statistically significant differences between the pre/post fatigue means of C ( p = 0.603 ) and α ( p = 0.223 ) for the 60 % group.
It should be noted that the C and α datasets for the 75 % 1-RM group violated the assumption of normality as assessed by the Shapiro-Wilk’s test with p = 0.007 and p < 0.0005 , respectively. Due to these violations of normality, the statistical comparisons of the pre- and post-fatigue values were tested using a Wilcoxon signed-rank test. Using this test it was determined that there were no statistically significant differences in the median values of both C ( p = 0.612 ) and α ( p = 0.160 ) in this group, similar to the testing of the 60 % group. All post-fatigue means that show a statistically significant difference are denoted using the ( * ) symbol in Table 1 and Table 2.

4. Discussion

In this work, a total of 72 impedance datasets (18 participants with two biceps each and two measurements per biceps) were fit to the Cole-model given by (Figure 1), expanding on the preliminary fitting of five datasets in [23]. The low errors of both resistance and reactance between the Cole-impedance model using the extracted parameters and the experimental measurements, observed in the relative error distribution of Figure 4, continues to support that this model is a very good choice for representing biological tissue impedance in this frequency range (1 kHz to 100 kHz). This highlights that this fractional-order model can accurately represent the frequency-dependent impedance behaviour of these tissues. The range of extracted values of α , 0.552 α 0.781 support that these tissues are well represented by a fractional-order, as these values are far from the integer order case of α = 1 that would occur otherwise.
From the experimental measurements, there are changes in resistance and reactance comparing the pre and post fatigue measurements of the participants biceps tissues. Samples of these changes were given in Figure 3 for participants 0007 and 0018. These changes result in differences of the Cole-impedance parameters that represent these impedance datasets in the pre and post fatigue states. Based on the statistical testing of the extracted model parameters, only R , R 1 , and f p show a statistically significant ( p < 0.05 ) difference between the pre-fatigue and post-fatigue states for both groups of participants. The resistance parameters of the 60 % group had mean decreases of 5.0 % and 13.2 % for R and R 1 , respectively, with the 75 % group having mean decreases of 5.4 % and 12.3 % . The decrease in resistance parameters is consistent with the previous studies investigating skeletal muscle fatigue using bioimpedance measurements [6,7]. These studies hypothesized that the decreases in resistance may be the result of mechanisms that include: (i) increased blood flow to the muscle due to the hemodynamic response to exercise [7], (ii) heat dissipation in the tissue increasing cutaneous blood flow and vasodilation [7], and (iii) an increase in metabolites in the tissue increasing its electrical conductivity [6]. An increase in muscle edema paired with a decrease in muscle quality has been hypothesized as the reason for decreased muscular strength following exercise [28]. Localized edema is expected to increase the available charge carriers in the region resulting in the decreased tissue resistance observed after the participants completed the exercise protocol. However, a limitation of this study is that the circumference of each participants biceps pre and post fatigue were not measured to quantify the association between the resistance parameters in the Cole-impedance model and swelling.
It is interesting to note that the C and α parameters did not have a statistically significant change between the immediately pre and immediately post exercise data in either of the groups measured in this study. While the discrete 10 kHz, 50 kHz, and 100 kHz reactance measurements did show a statistically significant difference between pre and post-fatigue measurements in the study participants [8], this may be an effect of the interactions between the resistive and capacitive components contributing to the overall real and imaginary components of the impedance. To highlight the complex interactions between components, the real and imaginary components of the Cole-impedance given by (1) have been separated to yield the expression:
Z = R + R 1 1 + ω α R 1 C cos α π 2 1 + 2 ω α R 1 C cos α π 2 + ω 2 α R 1 2 C 2 j R 1 w α R 1 C sin α π 2 1 + 2 ω α R 1 C cos α π 2 + ω 2 α R 1 2 C 2
Note that the imaginary term in (7) has contributions from R 1 , C, and α , so that changes in the measured reactance are not a result of only changes in C and α . Therefore, changes in R 1 will result reactance changes without changes in C or α , which could explain the significant changes in reactance reported in [8] without the significant changes in this analysis. To highlight this, a simulation of (7) with R = 15 Ω , C = 3 μ F sec ( α 1 ) , α = 0.75 and R 1 = 30 Ω , 28 Ω , and 26 Ω are given in Figure 5. Each of these simulations have different reactances even though only the R 1 value is varied. Further, the impedances at 50 kHz are given on each simulation case as a solid (·), which show a decrease with the decreasing R 1 values.
This also raises questions about which component of measured bioimpedance is the most appropriate marker to monitor changes that are a result of muscle damage, discrete reactances or capacitive components of electrical circuit representations. Previous work by Sanchez et al. investigating the bioimpedance changes in mice after injury supports that the capacitive component of the Cole-model is sensitive to changes from injury [29], which could indicate that the exercise protocol in this study, while fatiguing the participants, did not induce detectable levels of skeletal muscle damage. However, since this work did not collect markers to assess the localized tissue damage in participants it cannot be determined and will require follow-up investigations that collect further markers of tissue damage in addition to the participant bioimpedance measures. While the C and α changes in this study were not significant, the trend of decreasing resistance components and increases in the peak reactance frequency are consistent with changes that were observed in [29].
It is important to note that electrode positioning has a significant impact on the measured electrical impedance of localized tissues [30] yet there is no widely adopted process to standardize electrode placement for localized bio-impedance measurements; though recent guidelines have been proposed [31]. A limitation of this work is that the use of only a single-electrode configuration on the study participants fails to provide the data necessary to evaluate the optimum configuration to detect changes in the localized tissues due to exercise and fatigue.
Referring to the distortions that are observed in the impedance measures of the right arm of participant 0018, given in Figure 3d, these distortions may be an artifact of the measurement setup, specifically, resulting from the time required to apply the stepped-sine excitations to collect the measurements from 10 kHz to 100 kHz. For this series of tests, the Keysight E4990A was configured to measure at its highest accuracy setting, which required 0.2 s per measured frequency for a total of 13.6 s for the measurement from 10 kHz to 100 kHz. Even though participants were asked to relax their arms and remain motionless, there could have been variations in arm position and changes in the contraction of muscles which could have altered the electrical impedance of the tissue at each instant. This is a known limitation of the measurement of impedances using stepped sines for time-invariant impedances [32]. To reduce the potential impact of the time variant measures for future studies using the Keysight E4990A for bioimpedance studies, the measurement time could be reduced. However, this will reduce the accuracy of collected measurements and needs to be investigated to determine the correct balance of measurement time/accuracy for these applications. Also, additional sensor data could be collected during bioimpedance measures to assess motion and posture to evaluate windows during which movement artifacts will not be impacting the bioimpedance data; with this approach having been implemented for wearable systems to assess knee joint health [33].

5. Conclusions

We conclude that the Cole-impedance model does provide a very good fit to the measured electrical impedance of the biceps tissues from 10 kHz to 100 kHz of the participants in this study during both rested and fatigued states. This model shows very good agreement with the experimental data with less than ± 0.5 % and ± 2 % relative errors comparing the simulations using the Cole-model and extracted parameters to the experimental data, continuing to support the use of fractional-order circuit models to represent the frequency-dependent behaviour of biological tissues. The R , R 1 , and f p components of the Cole-model each exhibited statistically significant differences comparing pre and post-fatigue data in both groups of participants, supporting that these components may be an effective marker of fatigue-induced changes of the skeletal muscle. The changes in resistance parameters R and R 1 are expected to be a result of the localized tissue edema after the exercise protocol, but further studies are required to quantify this association.

Author Contributions

T.F. and B.F. conceived and designed the experiments; B.F. performed the experiments; T.F. analyzed the data; T.F. and B.F. prepared and reviewed the paper.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Grimnesand, S.; Martinsen, O. Bioimpedance and Bioelectricity Basics, 3rd ed.; Academic Press: London, UK, 2013. [Google Scholar]
  2. Zhu, F.; Kuhlmann, M.K.; Kotanko, P.; Seibert, E.; Leonard, E.G.; Levin, N.W. A method for the estimation of hydration state during hemodialysis using a calf bioimpedance technique. Physiol. Meas. 2008, 29, S503–S516. [Google Scholar] [CrossRef] [PubMed]
  3. Nescolarde, L.; Yanguas, J.; Terricabras, J.; Lukaski, H.; Alomar, X.; Rosell-Ferrer, J.; Rodas, G. Detection of muscle gaps by L-BIA in muscle injuries: clinical prognosis. Physiol. Meas. 2017, 38, L1–L9. [Google Scholar] [CrossRef] [PubMed]
  4. York, S.L.; Ward, L.C.; Czerniec, S.; Lee, M.J.; Refshauge, K.M.; Kilbreath, S.L. Single frequency versus bioimpedance spectroscopy for the assessment of lymphoedema. Breast Cancer Res. Treat. 2009, 117, 117–182. [Google Scholar] [CrossRef] [PubMed]
  5. Sanchez, B.; Rutkove, S.B. Electrical impedance myography and its applications in neuromuscular disorders. Neurotherapeutics 2017, 14, 107–118. [Google Scholar] [CrossRef] [PubMed]
  6. Li, L.; Shin, H.; Li, S.; Zhou, P. Localized electrical impedance myography of the biceps brachii muscle during different levels of isometric contraction and fatigue. Sensors 2016, 16, 581. [Google Scholar] [CrossRef] [PubMed]
  7. Vescio, G.; Rosell, J.; Nescolarde, L.; Giovinazzo, G. Muscle fatigue monitoring using a multifrequency bioimpedance technique. In 5th European Conference of the International Federation for Medical and Biological Engineering; Springer: Berlin/Heidelberg, Germany, 2011; pp. 1257–1260. [Google Scholar] [CrossRef]
  8. Fu, B.; Freeborn, T.J. Biceps tissue bioimpedance changes from isotonic exercise-induced fatigue at different intensities. Biomed. Phys. Eng. Express. 2018, 4, 025037. [Google Scholar] [CrossRef] [Green Version]
  9. Freeborn, T.J. A survey of fractional-order circuit models for biology and biomedicine. IEEE J. Emerging Sel. Top. Circuits Syst. 2013, 3, 416–424. [Google Scholar] [CrossRef]
  10. Cole, K.S. Permeability and impermeability of cell membranes for ions. Proc. Cold Spring Harbor Symp. Quant. Biol. 1940, 8, 110–122. [Google Scholar] [CrossRef]
  11. Nguyen Huu, D.; Kikuchi, D.; Maruyama, O.; Sapkota, A.; Takei, M. Cole-Cole analysis of thrombus formation in an extracoporeal blood flow circulation using electrical measurement. Flow Meas. Instrum. 2017, 53A, 172–179. [Google Scholar] [CrossRef]
  12. Kapur, K.; Taylor, R.S.; Qi, K.; Nagy, J.A.; Li, J.; Sanchez, B.; Rutkove, S.B. Predicting myofiber size with electrical impedance myography: A study in immature mice. Muscle Nerve 2018, 58, 106–113. [Google Scholar] [CrossRef] [PubMed]
  13. Jesus, I.S.; Tenreiro-Machado, J.A.; Cunha, J.B. Fractional electrical impedances in botanical elements. J. Vib. Control 2008, 14, 1389–1402. [Google Scholar] [CrossRef]
  14. Oldham, K.B.; Spanier, J. The Fractional Calculus: Theory and Applications of Differentiation and Integration to Arbitrary Order; Academic Press: New York, NY, USA, 1974. [Google Scholar]
  15. Ortigueira, M.D. Fractional Calculus for Scientists and Engineers; Springer: Heidelberg, Germany, 2011; ISBN 9789400707474. [Google Scholar]
  16. Magin, R.L. Fractional Calculus in Bioengineering; Begell House Connecticut: Paris, France, 2006. [Google Scholar]
  17. Magin, R.L. Fractional calculus models of complex dynamics in biological tissues. Comput. Math. Appl. 2010, 59, 1586–1593. [Google Scholar] [CrossRef]
  18. Ionescu, C.; Lopes, A.; Copot, D.; Tenreiro-Machado, J.A.; Bates, J.H.T. The role of fractional calculus in modeling biological phenomena: A review. Commun. Nonlinear Sci. Numer. Simul. 2017, 51, 141–159. [Google Scholar] [CrossRef]
  19. Das, S. Functional Fractional Calculus for System Identification and Controls; Springer: Berlin/Heidelberg, Germany, 2008. [Google Scholar]
  20. Ortigueira, M.D.; Tenreiro-Machado, J.A.; Trujillo, J.J. Fractional derivatives and periodic functions. Int. J. Dyn. Control 2015, 5, 72–78. [Google Scholar] [CrossRef]
  21. Butera, S.; Di Paola, M. A physically based connection between fractional calculus and fractal geometry. Annals Phys. 2014, 350, 146–154. [Google Scholar] [CrossRef] [Green Version]
  22. Galvao, R.K.H.; Hadjiloucas, S.; Kienitz, K.H.; Paiva, H.M.; Afonso, R.J.M. Fractional order modeling of large three-dimensional RC networks. IEEE Trans. Circuits Syst. I: Regul. Pap. 2013, 60, 624–637. [Google Scholar] [CrossRef]
  23. Freeborn, T.J.; Bohannan, G.W. Changes of fractional-order model parameters in biceps tissue from fatiguing exercise. In Proceedings of the 2018 IEEE International Symposium on Circuits and Systems (ISCAS), Florence, Italy, 27–30 May 2018. [Google Scholar] [CrossRef]
  24. Halter, R.J.; Hartov, A.; Paulsen, K.D.; Schned, A.; Heaney, J. Genetic and least squares algorithms for estimating spectral EIS parameters of prostatic tissues. Physiol. Meas. 2008, 29, S111–S123. [Google Scholar] [CrossRef] [PubMed]
  25. Gholami-Boroujeny, S.; Bolic, M. Extraction of Cole parameters from the electrical bioimpedance spectrum using stochastic optimization algorithms. Med. Biol. Eng. Computer. 2016, 54, 643–651. [Google Scholar] [CrossRef] [PubMed]
  26. Yousri, D.A.; AbdelAty, A.M.; Said, L.A.; AboBakr, A.; Radwan, A.G. Biological inspired optimization algorithms for cole-impedance parameters identification. Int. J. Electr. Commun. 2017, 78, 79–89. [Google Scholar] [CrossRef]
  27. Freeborn, T.J.; Maundy, B.; Elwakil, A.S. Extracting the parameters of the double-dispersion Cole bioimpedance model from magnitude response measurements. Med. Biol. Eng. Comput. 2014, 52, 749–758. [Google Scholar] [CrossRef] [PubMed]
  28. Clarkson, P.M.; Nosaka, K.; Braun, B. Muscle function after exercise-induced muscle damage and rapid adaptation. Med. Sci. Sports Exerc. 1992, 24, 512–520. [Google Scholar] [CrossRef] [PubMed]
  29. Sanchez, B.; Iyer, S.R.; Li, J.; Kapur, K.; Xu, S.; Rutkove, S.B.; Lovering, R.M. Non-invasive assessment of muscle injury in healthy and dystrophic animals with electrical impedance myography. Muscle Nerve 2017, 56, E85–E94. [Google Scholar] [CrossRef] [PubMed]
  30. Fu, B.; Freeborn, T.J. Estimating localized bio-impedance with measures from multiple redundant electrode configurations. In Proceedings of the 40th International Engineering in Medicine and Biology Conference, Honolulu, HI, USA, 17–21 July 2018. [Google Scholar]
  31. Sanchez, B.; Pacheck, A.; Rutkove, S.B. Guidelines to electrode positioning for human and animal electrical impedance myography research. Sci. Rep. 2016, 6, 32615. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  32. Louarroudi, E.; Sanchez, B. On the correct use of stepped-sine excitations for the measurement of time-varying bioimpedance. Physiol. Meas. 2017, 38, N73–N80. [Google Scholar] [CrossRef] [PubMed]
  33. Hersek, S.; Toreyin, H.; Teague, C.N.; Millard-Stafford, M.L.; Jeong, H.K.; Bavare, M.M.; Wolfkoff, P.; Sawka, M.N.; Inan, O.T. Wearable vector electrical bioimpedance system to assess knee joint health. IEEE Trans. Biomed. Eng. 2017, 64, 2353–2360. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Fractional-order equivalent circuit model, also known as the Cole-impedance model, to represent the frequency-dependent impedance of biological tissues.
Figure 1. Fractional-order equivalent circuit model, also known as the Cole-impedance model, to represent the frequency-dependent impedance of biological tissues.
Fractalfract 02 00027 g001
Figure 2. (a) Experimental test-setup to collect measurements from study participants using Keysight E4990A impedance analyzer using (b) tetrapolar electrode configuration.
Figure 2. (a) Experimental test-setup to collect measurements from study participants using Keysight E4990A impedance analyzer using (b) tetrapolar electrode configuration.
Fractalfract 02 00027 g002
Figure 3. Measured (solid) and simulated using Cole-impedance parameters (dashed) impedance of biceps tissue from (a) left and (b) right biceps of study participant 0007 and (c) left and (d) right biceps of study participant 0018. Pre-fatigue measures are presented as black lines with post-fatigue measures presented as blue lines. The peak reactance calculated using the Cole-impedance parameters are presented as a (o) in each figure.
Figure 3. Measured (solid) and simulated using Cole-impedance parameters (dashed) impedance of biceps tissue from (a) left and (b) right biceps of study participant 0007 and (c) left and (d) right biceps of study participant 0018. Pre-fatigue measures are presented as black lines with post-fatigue measures presented as blue lines. The peak reactance calculated using the Cole-impedance parameters are presented as a (o) in each figure.
Fractalfract 02 00027 g003
Figure 4. Relative error distributions of simulated impedances using Cole-impedance parameters compared to experimental values for (a) resistance and (b) reactance of the 60 % 1-RM group; and (c) resistance and (d) reactance of the 75 % 1-RM group (using all right/left and pre/post-fatigue datasets).
Figure 4. Relative error distributions of simulated impedances using Cole-impedance parameters compared to experimental values for (a) resistance and (b) reactance of the 60 % 1-RM group; and (c) resistance and (d) reactance of the 75 % 1-RM group (using all right/left and pre/post-fatigue datasets).
Fractalfract 02 00027 g004
Figure 5. Simulated Cole-impedance with R = 15 Ω , C = 3 μ F sec ( α 1 ) , α = 0.75 and R 1 = 30 Ω (black), 28 Ω (blue), and 26 Ω (red).
Figure 5. Simulated Cole-impedance with R = 15 Ω , C = 3 μ F sec ( α 1 ) , α = 0.75 and R 1 = 30 Ω (black), 28 Ω (blue), and 26 Ω (red).
Fractalfract 02 00027 g005
Table 1. Extracted equivalent circuit parameters from pre- and post-fatigue impedance measured from the left and right biceps of participants in the 60 % one-repetition maximum (1-RM) group.
Table 1. Extracted equivalent circuit parameters from pre- and post-fatigue impedance measured from the left and right biceps of participants in the 60 % one-repetition maximum (1-RM) group.
Participant R ( Ω ) R 1 ( Ω )C (F sec ( α 1 ) ) α f p (kHz)
PrePostPrePostPrePostPrePostPrePost
Left Biceps
0001 14.9 13.3 30.3 27.2 3.41 μ 2.05 μ 0.733 0.741 43.8 51.2
0002 17.4 15.6 40.1 34.4 1.70 μ 1.86 μ 0.768 0.759 42.5 53.4
0003 15.5 14.6 23.4 21.4 2.55 μ 2.34 μ 0.758 0.766 59.1 65.1
0004 28.2 28.5 36.4 29.1 5.06 μ 3.52 μ 0.676 0.719 52.7 55.8
0005 18.7 15.8 32.3 27.1 1.76 μ 1.82 μ 0.756 0.755 65.4 80.6
0006 15.0 15.5 33.6 29.1 3.75 μ 2.49 μ 0.717 0.753 43.4 50.0
0007 22.4 21.8 25.9 22.8 4.07 μ 2.69 μ 0.722 0.737 51.4 53.5
0008 15.4 14.0 34.8 29.2 1.88 μ 1.81 μ 0.759 0.768 51.6 58.5
0009 12.3 9.9 31.5 22.5 2.23 μ 3.36 μ 0.768 0.747 40.5 52.4
0010 28.0 27.3 36.4 36.7 6.39 μ 8.87 μ 0.658 0.627 53.3 58.4
Right Biceps
0001 10.9 9.3 26.3 25.7 3.24 μ 4.27 μ 0.745 0.723 45.7 47.6
0002 15.5 14.8 31.4 24.6 1.98 μ 2.08 μ 0.779 0.779 40.1 51.7
0003 16.5 14.7 24.9 23.2 1.95 μ 1.95 μ 0.781 0.776 53.5 62.6
0004 30.5 30.0 38.7 33.2 8.20 μ 6.58 μ 0.632 0.655 55.0 62.2
0005 16.6 15.4 31.1 24.5 1.75 μ 1.75 μ 0.761 0.774 64.1 69.7
0006 10.5 10.7 26.8 24.4 2.80 μ 2.28 μ 0.741 0.760 59.3 63.0
0007 20.0 19.1 25.5 22.1 4.40 μ 2.67 μ 0.719 0.736 49.3 57.6
0008 14.4 13.8 34.6 30.5 2.64 μ 2.68 μ 0.747 0.748 41.1 46.4
0009 11.4 10.9 30.9 26.6 4.45 μ 5.33 μ 0.727 0.715 32.6 38.4
0010 28.6 28.3 41.9 36.7 11.4 μ 9.83 μ 0.604 0.619 50.7 58.5
Mean 18.1 17.2 * 31.8 27.6 * 3.78 μ 2.47 μ 0.728 0.733 49.8 56.8 *
Std. Dev. 6.21 6.51 5.30 4.76 2.40 μ 2.32 μ 0.049 0.047 8.56 9.15
Note: ( * ) denotes that there is a statistically significant ( p < 0.05 ) difference between the pre-fatigue and post-fatigue means.
Table 2. Extracted equivalent circuit parameters from pre- and post-fatigue impedance measured from the left and right biceps of participants in the 75 % 1-RM group.
Table 2. Extracted equivalent circuit parameters from pre- and post-fatigue impedance measured from the left and right biceps of participants in the 75 % 1-RM group.
Participant R ( Ω ) R 1 ( Ω )C (F sec ( α 1 ) ) α f p (kHz)
PrePostPrePostPrePostPrePostPrePost
Left Biceps
0011 20.5 19.6 28.1 25.4 8.88 μ 10.1 μ 0.666 0.659 40.8 44.7
0012 13.5 14.4 33.3 25.3 6.48 μ 2.98 μ 0.704 0.779 25.5 30.9
0013 12.0 11.2 24.4 22.8 3.38 μ 3.82 μ 0.760 0.750 27.4 41.4
0014 32.3 33.0 50.8 44.8 13.1 μ 14.6 μ 0.552 0.552 90.2 93.8
0015 19.4 15.5 26.5 27.2 16.3 μ 13.2 μ 0.584 0.584 91.5 125.7
0016 13.3 12.7 34.3 31.1 2.70 μ 2.70 μ 0.755 0.756 34.9 39.1
0017 20.2 18.1 34.2 24.0 4.47 μ 3.56 μ 0.697 0.734 45.8 55.9
0018 24.0 23.2 24.7 22.0 6.38 μ 5.67 μ 0.704 0.713 40.0 47.7
Right Biceps
0011 20.7 18.6 25.1 21.3 6.81 μ 7.53 μ 0.706 0.702 34.2 40.7
0012 16.6 14.9 33.3 33.0 3.67 μ 3.41 μ 0.747 0.746 27.5 31.4
0013 9.5 8.0 21.4 18.5 2.83 μ 2.94 μ 0.774 0.775 44.9 51.0
0014 34.4 33.3 38.8 34.3 6.76 μ 7.26 μ 0.624 0.628 86.6 87.6
0015 28.8 29.0 31.5 33.2 6.74 μ 9.09 μ 0.625 0.602 118.9 112.0
0016 15.0 13.4 40.1 32.5 2.39 μ 2.03 μ 0.753 0.779 34.5 37.0
0017 20.3 18.8 33.3 27.1 2.56 μ 2.50 μ 0.731 0.742 59.0 65.9
0018 24.9 23.0 28.9 24.7 6.39 μ 5.42 μ 0.690 0.705 41.3 49.7
Mean 20.3 19.2 * 31.8 27.9 * 6.24 μ 6.05 μ 0.692 0.700 53.3 59.7 *
Std. Dev. 7.18 7.48 7.33 6.55 3.88 μ 3.92 μ 0.066 0.073 27.8 29.3
Note: ( * ) denotes that there is a statistically significant ( p < 0.05 ) difference between the pre-fatigue and post-fatigue means.

Share and Cite

MDPI and ACS Style

Freeborn, T.J.; Fu, B. Fatigue-Induced Cole Electrical Impedance Model Changes of Biceps Tissue Bioimpedance. Fractal Fract. 2018, 2, 27. https://0-doi-org.brum.beds.ac.uk/10.3390/fractalfract2040027

AMA Style

Freeborn TJ, Fu B. Fatigue-Induced Cole Electrical Impedance Model Changes of Biceps Tissue Bioimpedance. Fractal and Fractional. 2018; 2(4):27. https://0-doi-org.brum.beds.ac.uk/10.3390/fractalfract2040027

Chicago/Turabian Style

Freeborn, Todd J., and Bo Fu. 2018. "Fatigue-Induced Cole Electrical Impedance Model Changes of Biceps Tissue Bioimpedance" Fractal and Fractional 2, no. 4: 27. https://0-doi-org.brum.beds.ac.uk/10.3390/fractalfract2040027

Article Metrics