Surface Modification of Additively Manufactured Nitinol by Wet Chemical Etching

1
Institute of Machinery, Materials and Transport, Peter the Great Saint Petersburg Polytechnic University, Polytechnicheskaya, 29, 195221 Saint Petersburg, Russia
2
Research Centre “Innovative Technologies of Composite Nanomaterials”, Saint Petersburg State University, Universitetskaya Nab, 7/9, 199034 Saint Petersburg, Russia
3
Laboratory “Photoactive Nanocomposite Materials”, Saint Petersburg State University, Ulianovskaia Str. 1, Peterhof, 198504 Saint Petersburg, Russia
*
Author to whom correspondence should be addressed.
Submission received: 17 November 2021 / Revised: 7 December 2021 / Accepted: 8 December 2021 / Published: 13 December 2021
(This article belongs to the Special Issue Advances in Materials, Design and Modeling of Additive Manufacturing)

Abstract

:
Three-dimensional printed nitinol (NiTi) alloys have broad prospects for application in medicine due to their unique mechanical properties (shape memory effect and superplasticity) and the possibilities of additive technologies. However, in addition to mechanical properties, specific physicochemical characteristics of the surface are necessary for successful medical applications. In this work, a comparative study of additively manufactured (AM) NiTi samples etched in H2SO4/H2O2, HCl/H2SO4, and NH4OH/H2O2 mixtures was performed. The morphology, topography, wettability, free surface energy, and chemical composition of the surface were studied in detail. It was found that etching in H2SO4/H2O2 practically does not change the surface morphology, while HCl/H2SO4 treatment leads to the formation of a developed morphology and topography. In addition, exposure of nitinol to H2SO4/H2O2 and HCl/H2SO4 contaminated its surface with sulfur and made the surface wettability unstable in air. Etching in NH4OH/H2O2 results in surface cracking and formation of flat plates (10–20 microns) due to the dissolution of titanium, but clearly increases the hydrophilicity of the surface (values of water contact angles are 32–58°). The etch duration (30 min or 120 min) significantly affects the morphology, topography, wettability and free surface energy for the HCl/H2SO4 and NH4OH/H2O2 etched samples, but has almost no effect on surface composition.

Graphical Abstract

1. Introduction

Nearly equiatomic nickel–titanium (NiTi) alloys are used in many biomedical applications (stents, dental braces, surgical instruments, implants, etc.) due to their unique shape memory effect, superelasticity, relatively low modulus of elasticity, as well as good biocompatibility [1,2]. The rapid and successful development of 3D printing in recent years opens up even broader prospects for the use of NiTi in personalized medicine [1,3]. However, in addition to mechanical properties and specialized design, the biomaterial must have a number of specific characteristics such as morphology, topography, charge, and surface chemical composition [4]. Depending on the application, the material must be either bioinert or bioactive. At the same time, there are many requirements for materials resistance to biocorrosion, antibacterial properties, interaction with living cells, tissues, and the immune system [5]. In order for the material to meet such a set of characteristics, it is necessary to modify its surface.
Many surface modification methods are applied to nitinol. Micromachining methods including water jet machining (WJM), laser surface texturing (LST), electrochemical machining, and Micro-Electrical Discharge Machining (Micro-EDM) are actively used to remove material from the surface and produce developed topography [6]. WJM is promising since modified surfaces are not subjected to heat-related processes, and there are no formed cracks or deformed structures [7]. However, there are challenges relating to the depth homogeneity due to the strain hardening and yield strength [6]. LST is a thermal process to remove the material without any mechanical contact between NiTi and the laser beam. Therefore LST does not depend on the mechanical properties of the material. This method is very precise and allows you to adjust the surface topography in a wide range, but at the same time, it is quite expensive for scaling and industrial applications [8,9]. Electrochemical methods including anodization are also independent of the mechanical properties of NiTi and allow produce and precisely regulate topography and morphology of the surface [9,10]. However, these methods, as well as plasma treatment and ion implantation, are energy consuming and additional treatments are usually required to further improving medical properties [11].
Wet chemical etching (CE) is an effective and technically very simple method of forming surfaces with developed micro- and nanoscale morphology and modifying the surface composition by varying the nature of the etching agent, its composition, etching time and temperature [5,12]. CE of pure titanium and many of its medical alloys is very well studied and intensively used in industry [13,14]. For example, chemical etching using an H2SO4/HCl mixture is part of sandblasting and acid etching (SLA) technique to create a developed topography and hydrophilic surface, that accelerates and improves the osseointegration of titanium implants [15]. Despite this, the etching of nitinol is still very poorly studied, and the features of this process for 3D-printed NiTi samples are not known at all. To date, there are data on the CE of NiTi in acidic mixtures HF/HNO3/H2O [16,17], HCl/HNO3 [18] two-stage treatment using HCl/HF/H3PO4 and HNO3/HCl mixtures [19], and its surface modification with NaOH [20] and sequential exposure to NaOH and HNO3 [21]. The use of etchants of complex composition, as well as multi-step processes, allows for creating a complex hierarchical morphology, which is extremely important for nitinol applications as implants [4,22]. However, to the best of our knowledge, no data on the peculiarities of the nitinol etching process exist yet.
In this work, the effect of etchant composition and etching time on the physicochemical properties of the surface of nitinol produced by selective laser melting (SLM) was studied. Three types of etchants were chosen, the H2SO4/HCl mixture used in SLA technology [15], and piranha solutions (H2SO4/H2O2 and NH4OH/H2O2), which were successfully used for etching the titanium surface previously [23,24,25].

2. Materials and Methods

2.1. Samples Preparation

In our study, nitinol rods were obtained by selective laser melting (SLM). Commercial powders of nearly equiatomic NiTi alloys (CNPC Powder, Shanghai, China) were gas atomized. Samples were prepared using an SLM 280HL selective laser melting machine (SLM Solutions GmbH, Luebeck, Germany). The machine is equipped with two lasers with a power of 400 W and 1 kW [26].
Before etching, the NiTi rods (13 mm diameter, 40 cm length) were cut into discs (2–3 mm thick) using a Buehler IsoMet 1000 (Buehler, Lake Bluff, IL, USA). The obtained discs were polished using a semiautomatic Buehler MiniMet 1000 setup (Buehler, Lake Bluff, IL, USA) to the mirror-like surface using 600, 800, and 1200 grit sandpapers and a suspension of silica nanoparticles as described elsewhere [23,24]. Finally, the disks were repeatedly cleaned with acetone and deionized water in an ultrasonic bath for 15 min.
The polished NiTi discs were placed in 100 mL glass vessels with dual-acid etchant (H2SO4/HCl/H2O v/v/v = 9/11/30, Vecton, Saint-Petersburg, Russia), basic piranha solution (NH4OH/H2O2 v/v = 7/3, Vecton, Saint-Petersburg, Russia), or acidic piranha solution (H2SO4/H2O2 v/v = 7/3, Vecton, Saint-Petersburg, Russia) for 30 and 120 min. Immediately after etching, the samples were washed in deionized water.

2.2. Samples Characterization

The surface morphology was studied with a scanning electron microscopy (SEM) (Zeiss Merlin) operated at 20 kV. The magnifications from 100 up to 400,000 at 3–5 random positions were used. The topography of the samples’ surfaces was tested by a Solver P47 Pro atomic force microscope (AFM) (NT-MDT, Moscow, Russia) using the tapping mode with 1 × 1 μm and 10 × 10 μm scan areas. For each sample, 5–7 random positions were probed. The vertical range (Rz), root mean square roughness (RMS), average roughness (Ra), and specific surface area (SA—the ratio of 3D surface area over 2D scan size, in percent) were calculated using the program Gwyddion 2.59.
The composition of the samples’ surfaces was determined by X-ray photoelectron spectroscopy (XPS) with an Escalab 250Xi spectrometer (Thermo Fisher Scientific, Waltham, MA, USA). For XPS measurements, samples were excited with Al Kα (1486.7 eV) X-rays, and high-resolution spectra were charge-compensated by setting the C1s carbon line binding energy at 284.8 eV.
The contact angle (CA) measurements were performed by the sessile drop method using a Theta Lite optical tensiometer (Biolin Scientific, Sweden). The optical tensiometer was equipped with OneAttension Software for obtaining and processing results of tensiometric measurements. The volume of liquid droplets did not exceed 2.2 µL. The mean CA value (Θ) for the same sample was obtained by averaging at least 12 results of repeated measurements. The surface free energy (SFE) was calculated by the Owens–Wendt–Rabel–Kaelble (OWRK)/Fowkes approach using the two-liquid method (water versus diiodomethane contact angles) [27]. Ultrapure water served as a liquid with a dominant polar component of SFE (SFEp = 51.0 mN/m, SFEd= 21.8 mN/m), and diiodomethane stabilized by copper (99%, Sigma-Aldrich) was used as a liquid with a dominant dispersive component of SFE (SFEd = 48.5 mN/m, SFEp = 2.3 mN/m). The Cassie–Baxter equation [28] was used to account for the effect of surface roughness on the value of contact angles:
Cos Θr = r·Cos Θ0,
where Θr and Θ0 are the contact angles for rough and flat surfaces, respectively, and r is the roughness factor (SA ratio), r > 1.0.

3. Results

3.1. Weight Loss, Morphology, and Topography

Regardless of the etchant type, etching for 30 min does not lead to a noticeable change in weight (Table 1), but after the 2-h process, a noticeable weight loss is observed for HCl/H2SO4 and NH4OH/H2O2 treatments. Samples etched in H2SO4/H2O2 for two hours reveal negligible weight loss (0.051%).
The microscale (Figure 1) submicron (Figure 2) and nanoscale (Figure 3) morphology of the etched samples were studied by SEM using different magnification. It was found that the HCl/H2SO4-30 min samples do not have a developed topography: the surface is slightly damaged, the scratches formed during grinding and polishing are still visible (Figure 1a, Figure 2a and Figure 3a). However, chaotically located micron-size pits resulting from surface erosion appeared. In addition, grains with a diameter of several tens of nanometers were observed (Figure 3a). With an increase in the etching time up to 2 h, the aforementioned pits disappeared (Figure 1d and Figure 2d), and a developed microscale topography with particles of a few tens of nanometers (Figure 3d) appeared.
Regardless of the process duration, the samples etched in H2SO4/H2O2 do not have a developed morphology (Figure 1b,e and Figure 2b,e)and did not differ from the unetched polished NiTi (Figure S1). However, nanosized pits (Figure 3b) and nanoparticles (Figure 3e) were observed on the surface.
The morphology of the samples etched in NH4OH/H2O2 changed significantly. After etching for 30 min, a nanoscale porous structure and cracks were formed (Figure 2c and Figure 3c). The surface of the samples etched during 120 min consisted of relatively flat areas (Figure 1f) no more than 20–30 microns in size, separated by deep cracks (Figure 2f and Figure 3f).
The topography of the samples was studied by AFM using 10 × 10 (Figure 2, Insets) and 1 × 1 µm (Figure 3, Insets) scanning areas. AFM 3D surface topographies showed the same surface features (pits, particles, cracks, etc.) as the SEM images.
Surface topography parameters such as average roughness (Ra—Figure 4a), root mean square roughness (RMS—Figure 4b), the vertical range (Rz—Figure 4c), and specific surface area (SSA—ratio of 3D surface area to 2D scan size—Figure 4a) were calculated using AFM data. Etching in H2SO4/H2O2 slightly increased Ra, Rz, RMS, and SA. Etching in HCl/H2SO4 and NH4OH/H2O2-120 min significantly increased roughness, height difference and surface area. Topography parameters for NH4OH/H2O2-120 min and HCl/H2SO4 samples are very close to each other, but the real Ra, RMS, Rz, and surface area for NH4OH/H2O2-120 min sample should be greater because it has very deep cracks which cannot be properly measured by AFM. It is noteworthy that, for the NH4OH/H2O2-30 min sample, the roughness and maximum height difference practically did not increase (Figure 4c), but the specific surface area did increase (Figure 4d).

3.2. Surface Composition and Wettability

The wettability of samples strongly depends on the type of etchant and the duration of etching (Figure 5). The initial surface of nitinol is hydrophilic (CA = 71 ± 7°). Measurement of the water contact angles for the surface of as-prepared samples (immediately after etching within 1 h) showed that the treatment with HCl/H2SO4-30 min increased the water contact angle (WCA), and the surface became hydrophobic (Figure 5a). However, the longer the etching, the more hydrophilic the surface became (28 ± 7°). In general, a long-term etching (120 min) results in a more hydrophilic surface than a short one (30 min) for all etchants used. It is noteworthy that with aging (stored in the air), an increase in WCAs is observed for all samples except for HCl/H2SO4-30 min and NH4OH/H2O2-120 min (Figure 5b). Only NH4OH/H2O2 samples stored for 5 days in the air remained hydrophilic.
Analysis of the calculated surface free energy values showed that the hydrophilicity of the samples is due to the polar component (Figure 6). The hydrophobic HCl/H2SO4-30 min sample has practically no polar component, and all SFE is due to the dispersive component. Hydrophilic freshly prepared HCl/H2SO4-120 min and H2SO4/H2O2-120 min samples had a significant polar component (Figure 6), but they lost it with aging and became more hydrophobic (Figure 5b). Only the hydrophilicity of NH4OH/H2O2 samples was stable during aging: their WCA values (Figure 5b) and SFE values (Figure 6) did not change significantly.
The chemical composition of the surface of the samples was investigated by XPS. In addition to nickel, titanium, and oxygen, carbon was detected on the surface of the samples, as well as sulfur for the samples etched in HCl/H2SO4 and H2SO4/H2O2, and nitrogen for all NH4OH/H2O2 samples. The quantitative analysis showed that the surface layer of unetched NiTi is enriched in titanium with a ratio Ti/Ni = 4.1 (Table 2). Upon etching in acidic piranha solutions, this ratio increased, reaching 32.7 for the H2SO4/H2O2-30 min sample. SEM energy dispersive X-ray (EDX) analysis showed that Ni and Ti were evenly distributed over the material (Figures S2 and S3). When etching NH4OH/H2O2, the ratio decreased significantly because titanium was mostly etched out and the surface was greatly enriched with nickel (Ti/Ni = 0.14–0.16). SEM-EDX measurements (Figures S4 and S5) showed that a more oxidized and nickel-enriched surface is typical for micro-sized plates, and deep cracks contain nickel and titanium in a nearly equiatomic ratio.
In the high-resolution C1s spectra, two intense peaks at 284.8 eV and 286.3 eV were observed for all samples (Figure 7a). These peaks correspond to aliphatic hydrocarbons (C-C, C-H) and hydroxyl (C-OH) groups, respectively. Moreover, a small shoulder at higher binding energies (287–290 eV), assigned to carboxyl and aldehyde groups, was observed for the etched samples. All these carbon-containing species are originated from the adsorption of organic contaminations from the air during the sample storage.
For the samples etched in NH4OH/H2O2, a small amount of nitrogen was observed on the surface. The weak peaks of the N1s’ spectrum (Figure 7b) correspond to the C-N bond, but not to the NH4+ and NO3 or NO2 species. Therefore, we assume that the presence of nitrogen is caused by external contamination and not the result of etching in a nitrogen-containing etchant. On the other hand, sulfur was detected on the surface of the HCl/H2SO4 and H2SO4/H2O samples probably as a result of chemical etching in solutions containing H2SO4. The BE peak maxima in the S2p spectra correspond to SO42− group (Figure 7c).
Analysis of the Ni2p spectra showed (Figure 7b) that the surfaces of the unetched NiTi and NiTi etched in HCl/H2SO4 and H2SO4/H2O2 contain nickel predominantly in the metallic state, and etching in NH4OH/H2O2 leads to nickel oxidation with the formation of oxides and hydroxides.
According to high resolution Ti2p XPS spectra (Figure 8), the Ti 2p3/2 and Ti 2p1/2 peaks are located at 458.6 eV and 464.3 eV respectively for all samples and can be attributed to Ti4+ (Ti 2p1/2–Ti 2p3/2 splitting equal to 5.7) [29]. Low-intensity Ti3+ or Ti2+ shoulders at lower binding energies were detected for all samples. Metallic Ti or Ti–Ni peaks at 454.3 and 460.3 eV are quite evident in the spectrum of polished NiTi (Figure 8a) and HCl/H2SO4 (Figure 8b) samples.
O1s spectra showed the presence of TiO2, Ti–OH, H2O/C-OH groups for all samples (Figure 9). Moreover, for NiTi and NH4OH/H2O2 samples (Figure 9a), Ni–OH peaks are observed, and, for NH4OH/H2O2, a less intense NiO peak is also seen. For HCl/H2SO4 and H2SO4/H2O2 samples, SO42− peaks are present at 532.1 eV (Figure 9b).

4. Discussion

4.1. Surface Morphology and Topography

Wet chemical etching of metals is a very complex process. It includes many physical and chemical stages such as material removal, surface functionalization, oxidation and passivation. These steps depend on the composition of the etchant and etching conditions (time, temperature, mixing, etc.). Depending on the conditions, the etching process obeys either diffusion or kinetic control. In diffusion control, the process rate is determined by the rates of reagent supply and reaction product removal from the surface. As a result, the etching proceeds according to the layer-by-layer mechanism, and the roughness of the resulting surface is smoothed. The limiting step of the kinetic control etching is the chemical reaction on the local areas of the surface. In the case of surface heterogeneity and meaningful difference in the reactivity of local areas, the surface topography changes significantly.
It is important to note that the surfaces of the NiTi samples were mechanically treated (polished) before etching. Therefore, their surfaces were initially modified. We believe that the mirror-like surfaces of polished NiTis were homogenous. In this case, uniform etching of such surfaces using a layer-by-layer material removal mechanism is assumed. In this work, this type of etching is observed for the H2SO4/H2O2 mixture. Slow weight loss was observed (Table 1), but no significant changes of morphology and topography of the surface occurred (Figure 1, Figure 2, Figure 3 and Figure 4). Only nanoscale pits and particles were formed for the H2SO4/H2O2-30 min and H2SO4/H2O2-120 min samples, respectively. It was previously shown [23,30] that a noticeable change in the morphology and topography of the pure Ti surface upon etching in H2SO4/H2O2 begins to occur from 1–2 h from the start of etching. Probably, the NiTi sample requires a longer etching to form a developed morphology and topography than the Ti sample due to differences in thickness and/or composition of the surface layer. For HCl/H2SO4 and NH4OH/H2O2 etching, the weight loss within 30 min is also insignificant. At the same time, the formation of micro-sized pits (pitting etching) and nanosized grains was observed for the HCl/H2SO4-30 min samples, as well as microsized cracks with a “spongy” nanostructure for the NH4OH/H2O2-30 min samples. It can be assumed that these very active etchants can quickly remove the surface layer, and the etching process occurs by the mechanism of kinetic control with an inhomogeneous surface. We can also conclude that etching will proceed similarly in the case of an untreated additive manufactured nitinol surface. At longer etching (2 h), the developed microsized topography and high surface roughness were revealed. On the surface of the NH4OH/H2O2-120 min sample, flat plates of no more than 10–20 µm in size are formed, separated by deep microscale cracks. We believe that cracks are formed in regions with more defects and therefore thermodynamically more active for chemical etching grain boundaries. The formation of pits and different nanoscale and microscale structures can be explained by the selective etching of structure/composition defects of grains, as well as local heterogeneous areas on the solid–liquid interface.
Based on sample thickness and relative weight loss, wet chemical etching removes a very thin layer from several hundred nanometers (H2SO4/H2O2 and HCl/H2SO4-30 min etching) to several micrometers (NH4OH/H2O2 and HCl/H2SO4-120 min). Therefore, it can be assumed that the homogeneous surface layer that appears during polishing is very thin.
It should be noted that AFM images of surface topography converge with SEM images (Figure 2 and Figure 3). However, analysis of AFM data allows quantification of the surface topography. The task of comprehensive quantification of topography using AFM data is rather complicated [4,31], so we calculated the different topography parameters such as arithmetic mean and root means square roughness, maximum height amplitudes, and specific surface area. Roughnesses values are robust parameters of the overall topography of etched surfaces, but maximum height is very sensitive to noises, defects, pits, and spikes; and surface area indicates the degree of the surface topography development [4]. H2SO4/H2O2 etching did not significantly change any of the AFM parameters, but HCl/H2SO4 changed them significantly (Figure 4). It is noteworthy that etching in HCl/H2SO4 for 30 min strongly increased the maximum height and surface area, while the same etching for 120 min increased the surface roughness. This difference is caused by the formation of deep pits during 30 min-etching, and the increase in the roughness of HCl/H2SO4-120 min samples is a result of the formation of a more developed topography. Notably, etching in NH4OH/H2O2 for 30 min practically did not alter the roughness and maximum height difference (Figure 4c) but increased the specific surface area (Figure 4d) due to the nanoscale porous structure (Figure 3c). Etching in NH4OH/H2O2 for 120 min significantly increased all topographic parameters, and the resulting topography is the most developed of all.

4.2. Surface Energy and Composition

In addition to material removal, surface passivation and oxidation can occur during wet chemical etching. Based on the XPS data, the oxide layer on the surface of polished NiTi is very thin, since metallic Ti is seen in the Ti2p spectra (Figure 8a). After etching in HCl/H2SO4, this peak is also clearly seen, but after H2SO4/H2O2 and NH4OH/H2O2 etching, it is absent, that indicates the surface oxidation, i.e., the surface oxide layer becomes so thick that the basic NiTi alloy is not detected by the XPS method. Moreover, it should be noted that the oxidation process continues even after etching when the samples are stored in air. This is evidenced by the change in surface energy with aging. The surfaces of the as-prepared samples have a high polar component of the surface free energy, which can be explained by the presence of a very thin oxide layer. However, when samples are stored in the air, the SFE diminished due to the disappearance of its polar component. In the case of NH4OH/H2O2, the oxide-hydroxide surface layer is rather thick for as-prepared samples, and the polar component is due to different surface composition, which remains unchanged when stored in air.
There are currently insufficient data on the chemical etching of NiTi. Previously, chemical etching has only been used as a tool to solve various purposes but has not been studied in detail [16,17,18,19,20,21]. Possible reactions and mechanisms of etching and the formation of various micro- and nanoscale structures and surface composition variations were not discussed there. Nevertheless, the data obtained in this work and available from the literature allow us to make some assumptions and conclusions.
The chemistry of NiTi etching is quite complex. Various chemical reactions can take place in which both titanium and nickel are involved. In the case of H2O2-containing solutions, both simple oxidation to TiO2, Ti2O3, TiO [32] and the formation of hydroxo and peroxo complexes of titanium (TiO(OH)2, Ti(H2O2)O2) can occur [33,34]. Etching with H2SO4/H2O2 and HCl/H2SO4 also produces Ti(SO4)2 and TiOSO4 surface compounds. It should be noted that no chlorides were detected when HCl/H2SO4 was used. The NH4OH/H2O2 etching probably results in fewer titanium peroxocomplexes, but more soluble and easily formed ammonia complexes such as [Ti(NH3·H2O)2]4+ [25]. As a result, the titanium can be removed into solution and excess nickel remains on the surface as an oxide-hydroxide (NiO, Ni2O3, Ni(OH)2). According to XPS data, the surfaces of all samples contain Ti–OH species, but it is unlikely that it is titanium hydroxide, since these groups were found even after etching in strong acidic mixtures—HCl/H2SO4. The titanium is probably hydroxylated when the samples are washed in water after etching.
One of the most interesting obtained results is the change in ratio Ni/Ti after etching. We assume that the enrichment of the surface with titanium is caused by the dissolution of nickel with the simultaneous oxidation of titanium and the growth of the surface oxide layer (TiO2 or TiOx). Titanium is more readily oxidized than a nickel from a thermodynamic point of view:
Ti(s) + O2(g) → TiO2(s)  ΔHf = −956 kJ/mol
Ni(s) + 1/2O2(g) → NiO(s)  ΔHf = −241 kJ/mol
The large disparity between the enthalpy of the TiO2 and NiO may reflect the stronger affinity of Ti than that of Ni for oxidation. Such a type of oxidation could be realized when H2O2 is used. According to XPS data, these processes are observed during etching in H2SO4/H2O2 and partly in HCl/H2SO4. Under these conditions, the thickness of the titanium oxide layer increases, and the nickel is present on the surface in a small amount in the metallic state. After etching in NH4OH/H2O2, the surface is enriched with nickel. Apparently, in this case, nickel is oxidized under the action of H2O2 to an insoluble oxide NiOx (Figure 7d), and titanium dissolves and goes into solution in the form of ammine complexes such as Ti(NH3)x4+ or TiO(NH3)x2+. We believe that titanium dissolves precisely under the action of NH4OH, since according to literature when exposed to H2O2, the nitinol surface should be enriched with titanium due to titania formation [35].

4.3. Prospects for NiTi Chemical Etching in Medicine

As mentioned in the Introduction section, the HCl/H2SO4 mixture is actively used in SLA technology to modify the surface of titanium implants. The HCl/H2SO4 etching agent improves surface topography and successfully removes SiO2 and Al2O3 particles that remain on the implant surface after sandblasting. Based on a large number of in vitro and in vivo studies, researchers agree that developed morphology and topography at the macro-, micro- and nanoscale are vital for the most successful biointegration of the material [4,12,31,36]. From a biomechanical point of view, macro-roughness improves the implant surface area as well as increases the friction coefficient and the implant stability. Microscale structures, especially micropores and micropits, significantly enhance the adhesion of bone tissue cells. Specific nano-roughness and nanostructures are very important parameters to improve circulation, adsorption and increase the number of sites for various biomolecules (signaling proteins, cytokines, integrins, hormones, and growth factors, etc.) and nutrients. Moreover, the nanostructures have an antibacterial effect [37].
The results of this study showed that the HCl/H2SO4 treatment forms a developed morphology and topography of the nitinol surface. The NH4OH/H2O2 mixture is an equally successful etchant for NiTi, and the 2 h-treatment results in deep cracks on the surface, separating flat plates no larger than 10–20 µm in size. Such surface morphology has great prospects for use in medicine, because the size of the plates is close to the size of cells such as osteoblasts, fibroblasts, mesenchymal cells, etc., and deep cracks can promote the formation and growth of cell appendages [36].
The surface composition is also very important for successful medical applications. The most valuable parameter is the presence of a large concentration of hydrophilic surface species [38]. This leads to faster adsorption of different biomolecules and affects cell activity [39]. From this point of view, the advantages of the NH4OH/H2O2 etching are obvious. Another important parameter is the presence of a rather thick layer of biocompatible titanium oxide, which prevents biocorrosion and diffusion of metallic Ni. Although nitinol is considered a corrosion-resistant alloy, the nickel in it is extremely toxic and carcinogenic [40,41]. From this point of view, the samples etched with NH4OH/H2O2 are much worse than those etched with HCl/H2SO4 and even H2SO4/H2O2, since their surfaces are nickel-enriched. Nevertheless, the disadvantages of the surface composition produced by chemical etching can be compensated by additional surface modification allowing to preserve the morphology and topography of the surface.
To date, the in vitro results on chemically etched NiTi have been presented only in two works [10,19], while the in vivo results are completely absent. Two-stage chemical etching in HCl/HF/H3PO4 and HNO3/HCl mixtures was shown to be promising for significantly improving the adhesion and proliferation of MG-63 preosteoblast cells. Yamasaki et al. showed that NiTi anodized in HNO3/H3PO4 solutions enhanced the attachment and spreading MC3T3-E1 osteoblast-like cells regardless of the etching mixture ratio [10]. In this regard, our future plans are to study the effect of chemical etching of nitinol in HCl/H2SO4, H2SO4/H2O2, and NH4OH/H2O2 on the in vitro as well as in vivo response of various cells.

5. Conclusions

A comparative study of NiTi etching showed that H2SO4/H2O2 does not significantly change the surface morphology and roughness, whereas etching in HCl/H2SO4 leads to the formation of a developed microsized topography and nanograins. Prolonged etching (2 h) in an H2SO4/H2O2 or HCl/H2SO4 mixture significantly increases the hydrophilicity, as indicated by an increase in the polar component of surface free energy, but hydrophilicity drops when samples are stored in air for 5 days. All samples etched in mixtures containing H2SO4 showed the presence of a small amount of SO42− groups on the surface. Etching in NH4OH/H2O2 dissolves titanium and oxidizes nickel to form its oxide and hydroxide, and steadily increases hydrophilicity over time, as indicated by the increase in surface energy mainly due to its polar component. The surface morphology and topography of the samples etched in NH4OH/H2O2 strongly depend on the duration of etching. At a 30-min etching exposure, the surface is characterized by a spongy nanostructure with low roughness but a large surface. Etching for 2 h leads to the formation of deep microscale cracks separating flat plates not more than 10–20 µm in size. The results presented will allow us to develop an approach for creating surfaces with physicochemical properties that can be varied over a wide range, including for optimization for specific applications.

Supplementary Materials

The following are available online at https://0-www-mdpi-com.brum.beds.ac.uk/article/10.3390/ma14247683/s1, Figure S1: SEM images of polished NiTi at different magnifications—200,000× (a), 100,000× (b), 10,000× (c) 3000× (d). In the insets—AFM surface topographies. Figure S2. SEM images of NiTi etched in H2SO4/H2O2-30 min and the results of EDX chemical analysis. Figure S3. SEM images of NiTi etched in H2SO4/H2O2-120 min and the results of EDX chemical analysis. Figure S4. SEM images of NiTi etched in NH4OH/H2O2-30 min and the results of EDX chemical analysis. Figure S5. SEM images of NiTi etched in NH4OH/H2O2-120 min and the results of EDX chemical analysis.

Author Contributions

Conceptualization, D.N.; methodology, D.N. and E.B.; validation, A.R. and D.N.; formal analysis, D.N. and A.R.; investigation, D.N., E.B. and A.R.; resources, E.B. and A.P.; data curation, D.N. and A.R.; writing—original draft preparation, D.N.; writing—review and editing, A.R. and A.P.; visualization, D.N.; supervision, D.N. and A.P.; project administration, D.N. and A.P.; funding acquisition, D.N. All authors have read and agreed to the published version of the manuscript.

Funding

The study was carried out with the financial support of a scholarship from the President of the Russian Federation to young scientists and graduate students carrying out promising research and development in priority areas of modernization of the Russian economy for 2021–2023 (Project SP-1934.2021.4). The research is partially funded by the Ministry of Science and Higher Education of the Russian Federation under the strategic academic leadership program “Priority 2030” (Agreement 075-15-2021-1333 dated 30 September 2021).

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The main data had been provided in the paper and Supplementary Material. Any other raw/processed data required to reproduce the findings of this study are available from the corresponding author upon request.

Acknowledgments

This research was conducted using the equipment of the resource centers of the Research Park of the St. Petersburg State University «Innovative Technologies of Composite Nanomaterials», «Physical Methods of Surface Investigation» (personally to Alexandra Koroleva), «Nanotechnology» (personally to Vladimir Kalganov) and «Nanophotonics» (personally to Maxim Lozhkin). A.R. acknowledges financial support from Saint Petersburg State University (Pure ID 73032813).

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Wadood, A. Brief Overview on Nitinol as Biomaterial. Adv. Mater. Sci. Eng. 2016, 2016, 4173138. [Google Scholar] [CrossRef] [Green Version]
  2. Hang, R.; Zhao, F.; Yao, X.; Tang, B.; Chu, P.K. Self-assembled anodization of NiTi alloys for biomedical applications. Appl. Surf. Sci. 2020, 517, 146118. [Google Scholar] [CrossRef]
  3. Safaei, K.; Abedi, H.; Nematollahi, M.; Kordizadeh, F.; Dabbaghi, H.; Bayati, P.; Javanbakht, R.; Jahadakbar, A.; Elahinia, M.; Poorganji, B. Additive Manufacturing of NiTi Shape Memory Alloy for Biomedical Applications: Review of the LPBF Process Ecosystem. JOM 2021, 1–16, in press. [Google Scholar] [CrossRef]
  4. Wennerberg, A.; Albrektsson, T.; Jimbo, R. Implant Surfaces and Their Biological and Clinical Impact; Springer: Heidelberg, Germany; New York, NY, USA; Dordrecht, The Netherlands; London, UK, 2015; p. 182. [Google Scholar]
  5. John, A.A.; Subramanian, A.P.; Vellayappan, M.V.; Balaji, A.; Jaganathan, S.K.; Mohandas, H.; Paramalinggam, T.; Supriyanto, E.; Yusof, M. Review: Physico-chemical modification as a versatile strategy for the biocompatibility enhancement of biomaterials. RSC Adv. 2015, 5, 39232–39244. [Google Scholar] [CrossRef]
  6. Mwangi, J.W.; Nguyen, L.T.; Bui, V.D.; Berger, T.; Zeidler, H.; Schubert, A. Nitinol manufacturing and micromachining: A review of processes and their suitability in processing medical-grade nitinol. J. Manuf. Process. 2019, 38, 355–369. [Google Scholar] [CrossRef]
  7. Kong, M.C.; Srinivasu, D.; Axinte, D.; Voice, W.; McGourlay, J.; Hon, B. On geometrical accuracy and integrity of surfaces in multi-mode abrasive waterjet machining of NiTi shape memory alloys. CIRP Ann. 2013, 62, 555–558. [Google Scholar] [CrossRef]
  8. Wang, M.; Yang, Z.; Yang, C.; Zhang, D.; Tian, Y.; Liu, X. The investigation of mechanical and thermal properties of super-hydrophobic nitinol surfaces fabricated by hybrid methods of laser irradiation and carbon ion implantation. Appl. Surf. Sci. 2020, 527, 146889. [Google Scholar] [CrossRef]
  9. Ma, L.; Li, W.; Yang, Y.; Ma, Y.; Luo, K.; Jia, B.; Xu, Z.; Yu, Z. Corrosion Behavior of NiTi Alloys Fabricate by Selective Laser Melting Subjected to Femtosecond Laser Shock Peening. Coatings 2021, 11, 1078. [Google Scholar] [CrossRef]
  10. Yamasaki, K.; Hirano, M.; Komai, S.; Taniho, H.; Ohtsu, N. Cell responses on Ni-free anodized layer of NiTi alloy with various surface morphologies. Appl. Surf. Sci. 2020, 531, 147351. [Google Scholar] [CrossRef]
  11. Gao, A.; Hang, R.; Bai, L.; Tang, B.; Chu, P.K. Electrochemical surface engineering of titanium-based alloys for biomedical application. Electrochim. Acta 2018, 271, 699–718. [Google Scholar] [CrossRef]
  12. Liu, X.; Chu, P.; Ding, C. Surface modification of titanium, titanium alloys, and related materials for biomedical applications. Mater. Sci. Eng. R Rep. 2004, 47, 49–121. [Google Scholar] [CrossRef] [Green Version]
  13. Souza, J.C.M.; Sordi, M.B.; Kanazawa, M.; Ravindran, S.; Henriques, B.; Silva, F.S.; Aparicio, C.; Cooper, L.F. Nano-scale modification of titanium implant surfaces to enhance osseointegration. Acta Biomater. 2019, 94, 112–131. [Google Scholar] [CrossRef]
  14. Janeczek, M.; Szymczyk, P.; Dobrzynski, M.; Parulska, O.; Szymonowicz, M.; Kuropka, P.; Rybak, Z.; Zywicka, B.; Ziolkowski, G.; Marycz, K.; et al. Influence of surface modifications of a nanostructured implant on osseointegration capacity—Preliminary in vivo study. RSC Adv. 2018, 8, 15533–15546. [Google Scholar] [CrossRef] [Green Version]
  15. Masrouri, M.; Faraji, G.; Pedram, M.S.; Sadrkhah, M. In-vivo study of ultrafine-grained CP-Ti dental implants surface modified by SLActive with excellent wettability. Int. J. Adhes. Adhes. 2020, 102, 102684. [Google Scholar] [CrossRef]
  16. Shabalovskaya, S.A.; Anderegg, J.; Laab, F.; Thiel, P.A.; Rondelli, G. Surface Conditions of Nitinol Wires, Tubing, and As-Cast Alloys. The Effect of Chemical Etching, Aging in Boiling Water, and Heat Treatment. J. Biomed. Mater. Res.-Part B Appl. Biomater. 2003, 65, 193–203. [Google Scholar] [CrossRef]
  17. Khalili, V.; Naji, H. Developing a mechanochemical surface pretreatment to increase the adhesion strength of hydroxyapatite electrophoretic coating on the NiTi alloy as a bone implant. Surf. Coat. Technol. 2020, 397, 125985. [Google Scholar] [CrossRef]
  18. Pace, B.; Bendavid, A.; Ahsan, M.; Dargusch, M.; Bhatia, V.; Byrnes, J.; Cairney, J. Tuning Ta coating properties through chemical and plasma etching pre-treatment of NiTi wire substrates. Surf. Coat. Technol. 2021, 418, 127214. [Google Scholar] [CrossRef]
  19. Norouzi, N.; Nouri, Z. The Effect of Two-Stage Acid Treatment on Surface Behavior and Improvement of Bioactivity of Nitinol Alloy. Biointerface Res. Appl. Chem. 2020, 11, 10690–10702. [Google Scholar] [CrossRef]
  20. Fathyunes, L.; Sheykholeslami, S.O.R. The surface modification of Nitinol superelastic alloy with alkaline-heat treatment and hydroxyapatite/chitosan composite coating for biomedical applications. J. Ultrafine Grained Nanostruct. Mater. 2020, 53, 166–176. [Google Scholar] [CrossRef]
  21. Rossi, S.; Deflorian, F.; Pegoretti, A.; D’Orazio, D.; Gialanella, S. Chemical and mechanical treatments to improve the surface properties of shape memory NiTi wires. Surf. Coat. Technol. 2008, 202, 2214–2222. [Google Scholar] [CrossRef]
  22. Ren, B.; Wan, Y.; Liu, C.; Wang, H.; Yu, M.; Zhang, X.; Huang, Y. Improved osseointegration of 3D printed Ti-6Al-4V implant with a hierarchical micro/nano surface topography: An in vitro and in vivo study. Mater. Sci. Eng. C Mater. Biol. Appl. 2021, 118, 111505. [Google Scholar] [CrossRef] [PubMed]
  23. Nazarov, D.V.; Smirnov, V.M.; Zemtsova, E.G.; Yudintceva, N.M.; Shevtsov, M.A.; Valiev, R.Z. Enhanced Osseointegrative Properties of Ultra-Fine-Grained Titanium Implants Modified by Chemical Etching and Atomic Layer Deposition. ACS Biomater. Sci. Eng. 2018, 4, 3268–3281. [Google Scholar] [CrossRef] [PubMed]
  24. Nazarov, D.; Zemtsova, E.; Smirnov, V.; Mitrofanov, I.; Maximov, M.; Yudintceva, N.; Shevtsov, M. The Effects of Chemical Etching and Ultra-Fine Grain Structure of Titanium on MG-63 Cells Response. Metals 2021, 11, 510. [Google Scholar] [CrossRef]
  25. Yuan, X.; Kang, Y.; Zuo, J.; Xie, Y.; Ma, L.; Ren, X.; Bian, Z.; Wei, Q.; Zhou, K.; Wang, X.; et al. Micro/nano hierarchical structured titanium treated by NH4OH/H2O2 for enhancing cell response. PLoS ONE 2018, 13, e0196366. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  26. Borisov, E.; Starikov, K.; Popovich, A.; Tihonovskaya, T. Investigation of the Possibility of Tailoring the Chemical Com-Position of the NiTi Alloy by Selective Laser Melting. Metals 2021, 11, 1470. [Google Scholar] [CrossRef]
  27. Owens, D.K.; Wendt, R.C. Estimation of the surface free energy of polymers. J. Appl. Polym. Sci. 1969, 13, 1741–1747. [Google Scholar] [CrossRef]
  28. Cassie, A.B.D.; Baxter, S. Wettability of porous surfaces. Trans. Faraday Soc. 1944, 40, 546–551. [Google Scholar] [CrossRef]
  29. Biesinger, M.C.; Lau, L.W.M.; Gerson, A.R.; Smart, R.S.C. Resolving surface chemical states in XPS analysis of first row transition metals, oxides and hydroxides: Sc, Ti, V, Cu and Zn. Appl. Surf. Sci. 2010, 257, 887–898. [Google Scholar] [CrossRef]
  30. Nazarov, D.V.; Zemtsova, E.G.; Solokhin, A.Y.; Valiev, R.Z.; Smirnov, V.M. Modification of the Surface Topography and Composition of Ultrafine and Coarse Grained Titanium by Chemical Etching. Nanomaterials 2017, 7, 15. [Google Scholar] [CrossRef] [Green Version]
  31. Wennerberg, A.; Albrektsson, T. Suggested Guidelines for the Topographic Evaluation of Implant Surfaces. Int. J. Oral Maxillofac. Implant. 2000, 15, 331–344. [Google Scholar]
  32. Cotton, F.A.; Wilkinson, G. Advanced Inorganic Chemistry; John Willey & Sons, Inc.: London, UK; New York, NY, USA, 1972; pp. 807–818. [Google Scholar]
  33. Tengvall, P.; Lundström, I.; Sjöqvist, L.; Elwing, H.; Bjursten, L.M. Titanium-hydrogen peroxide interaction: Model studies of the influence of the inflammatory response on titanium implants. Biomaterials 1989, 10, 166–175. [Google Scholar] [CrossRef]
  34. Mühlebach, J.; Müller, K.; Schwarzenbach, G. The peroxo complexes of titanium. Inorg. Chem. 1970, 9, 2381–2390. [Google Scholar] [CrossRef]
  35. Chu, C.L.; Chung, C.Y.; Chu, P.K. Surface oxidation of NiTi shape memory alloy in a boiling aqueous solution containing hydrogen peroxide. Mater. Sci. Eng. A 2006, 417, 104–109. [Google Scholar] [CrossRef]
  36. Bacakova, L.; Filova, E.; Parizek, M.; Ruml, T.; Svorcik, V. Modulation of cell adhesion, proliferation and differentiation on materials designed for body implants. Biotechnol. Adv. 2011, 29, 739–767. [Google Scholar] [CrossRef]
  37. Lorenzetti, M.; Dogsa, I.; Stosicki, T.; Stopar, D.; Kalin, M.; Kobe, S.; Novak, S. The influence of surface modification on bacterial adhesion to titanium-based substrates. ACS Appl. Mater. Interfaces 2015, 7, 1644–1651. [Google Scholar] [CrossRef]
  38. Geetha, M.; Singh, A.K.; Asokamani, R.; Gogia, A.K. Ti based biomaterials, the ultimate choice for orthopaedic implants–A review. Prog. Mater. Sci. 2009, 54, 397–425. [Google Scholar] [CrossRef]
  39. von der Mark, K.; Park, J. Engineering biocompatible implant surfaces. Prog. Mater. Sci. 2013, 58, 327–381. [Google Scholar] [CrossRef]
  40. Matusiewicz, H. Potential release of in vivo trace metals from metallic medical implants in the human body: From ions to nanoparticles--A systematic analytical review. Acta Biomater. 2014, 10, 2379–2403. [Google Scholar] [CrossRef]
  41. Veverkova, J.; Bartkova, D.; Weiser, A.; Dlouhy, A.; Babula, P.; Stepka, P.; Pavkova Goldbergova, M. Effect of Ni ion release on the cells in contact with NiTi alloys. Environ. Sci. Pollut. Res. Int. 2020, 27, 7934–7942. [Google Scholar] [CrossRef]
Figure 1. SEM images (10 µm scale bars) of nitinol surface for samples: (a) HCl/H2SO4-30 min, (b) H2SO4/H2O2-30 min, (c) NH4OH/H2O2-30 min, (d) HCl/H2SO4-120 min, (e) H2SO4/H2O2-120 min, (f) NH4OH/H2O2-120 min.
Figure 1. SEM images (10 µm scale bars) of nitinol surface for samples: (a) HCl/H2SO4-30 min, (b) H2SO4/H2O2-30 min, (c) NH4OH/H2O2-30 min, (d) HCl/H2SO4-120 min, (e) H2SO4/H2O2-120 min, (f) NH4OH/H2O2-120 min.
Materials 14 07683 g001
Figure 2. SEM images (3 µm scale bars) of nitinol surface for samples: (a) HCl/H2SO4-30 min, (b) H2SO4/H2O2-30 min, (c) NH4OH/H2O2-30 min, (d) HCl/H2SO4-120 min, (e) H2SO4/H2O2-120 min, (f) NH4OH/H2O2-120 min. The insets show the AFM surface topographies of the corresponding samples.
Figure 2. SEM images (3 µm scale bars) of nitinol surface for samples: (a) HCl/H2SO4-30 min, (b) H2SO4/H2O2-30 min, (c) NH4OH/H2O2-30 min, (d) HCl/H2SO4-120 min, (e) H2SO4/H2O2-120 min, (f) NH4OH/H2O2-120 min. The insets show the AFM surface topographies of the corresponding samples.
Materials 14 07683 g002
Figure 3. SEM images (300 nm scale bars) of nitinol surface for samples: (a) HCl/H2SO4-30 min, (b) H2SO4/H2O2-30 min, (c) NH4OH/H2O2-30 min, (d) HCl/H2SO4-120 min, (e) H2SO4/H2O2-120 min, (f) NH4OH/H2O2-120 min. The insets show the AFM surface topographies of the corresponding samples.
Figure 3. SEM images (300 nm scale bars) of nitinol surface for samples: (a) HCl/H2SO4-30 min, (b) H2SO4/H2O2-30 min, (c) NH4OH/H2O2-30 min, (d) HCl/H2SO4-120 min, (e) H2SO4/H2O2-120 min, (f) NH4OH/H2O2-120 min. The insets show the AFM surface topographies of the corresponding samples.
Materials 14 07683 g003
Figure 4. AFM topography parameters for etched nitinol samples: average roughness (Ra)—(a), root mean square roughness (RMS)—(b), the vertical range (Rz)—(c), and specific surface area—(d).
Figure 4. AFM topography parameters for etched nitinol samples: average roughness (Ra)—(a), root mean square roughness (RMS)—(b), the vertical range (Rz)—(c), and specific surface area—(d).
Materials 14 07683 g004
Figure 5. Results of water and diiodomethane contact angle measurements of as-prepared (a) and aged in the air (b) samples.
Figure 5. Results of water and diiodomethane contact angle measurements of as-prepared (a) and aged in the air (b) samples.
Materials 14 07683 g005
Figure 6. Results of the surface free energy calculation for unetched and etched nitinol.
Figure 6. Results of the surface free energy calculation for unetched and etched nitinol.
Materials 14 07683 g006
Figure 7. High-resolution XPS spectra of polished NiTi and etched samples: C1s (a), N1s (b), S2p (c), Ni2p (d).
Figure 7. High-resolution XPS spectra of polished NiTi and etched samples: C1s (a), N1s (b), S2p (c), Ni2p (d).
Materials 14 07683 g007
Figure 8. High-resolution Ti2p XPS spectra of polished NiTi and etched in NH4OH/H2O2-120 min samples (a) and etched in HCl/H2SO4-120 min and H2SO4/H2O2-120 min (b).
Figure 8. High-resolution Ti2p XPS spectra of polished NiTi and etched in NH4OH/H2O2-120 min samples (a) and etched in HCl/H2SO4-120 min and H2SO4/H2O2-120 min (b).
Materials 14 07683 g008
Figure 9. High-resolution O1s XPS spectra of polished NiTi and etched in NH4OH/H2O2-120 min samples (a) and etched in HCl/H2SO4-120 min and H2SO4/H2O2-120 min (b).
Figure 9. High-resolution O1s XPS spectra of polished NiTi and etched in NH4OH/H2O2-120 min samples (a) and etched in HCl/H2SO4-120 min and H2SO4/H2O2-120 min (b).
Materials 14 07683 g009
Table 1. Weight loss of chemically etched nitinol.
Table 1. Weight loss of chemically etched nitinol.
Type of Chemical
Etching
Weight
before CE, g
Weight
after CE, g
Weight Loss,
g
Weight Loss,
%
HCl/H2SO4-30 min1.713601.713240.000360.021
HCl/H2SO4-120 min1.787141.783890.003250.182
H2SO4/H2O2-30 min1.528441.527870.000570.037
H2SO4/H2O2-120 min1.623491.622670.000830.051
NH4OH/H2O2-30 min1.321861.321250.000610.046
NH4OH/H2O2-120 min1.431721.427710.004020.281
Table 2. Data of XPS quantitative analysis.
Table 2. Data of XPS quantitative analysis.
SampleTi2pNi2pO1sC1sN1sS2p
NiTi8.292.0129.7559.95--
HCl/H2SO4-30 min6.180.4637.7253.15-2.49
HCl/H2SO4-120 min10.200.4744.8941.70-2.74
H2SO4/H2O2-30 min8.830.2740.6248.56-1.72
H2SO4/H2O2-120 min6.560.5236.5054.79-1.64
NH4OH/H2O2-30 min2.0514.2443.9537.682.08-
NH4OH/H2O2-120 min2.9517.8936.9240.471.76-
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Nazarov, D.; Rudakova, A.; Borisov, E.; Popovich, A. Surface Modification of Additively Manufactured Nitinol by Wet Chemical Etching. Materials 2021, 14, 7683. https://0-doi-org.brum.beds.ac.uk/10.3390/ma14247683

AMA Style

Nazarov D, Rudakova A, Borisov E, Popovich A. Surface Modification of Additively Manufactured Nitinol by Wet Chemical Etching. Materials. 2021; 14(24):7683. https://0-doi-org.brum.beds.ac.uk/10.3390/ma14247683