High-Resolution Multi-Channel Frequency Standard Comparator Using Digital Frequency Measurement

1
National Time Service Center, Chinese Academy of Sciences, East Shu Yuan Road, Xi’an 710600, China
2
Key Laboratory of Time and Frequency Primary Standards, Chinese Academy of Sciences, Xi’an 710600, China
3
School of Astronomy and Space Science, University of Chinese Academy of Sciences, Yu Quan Road, Beijing 100049, China
*
Author to whom correspondence should be addressed.
Submission received: 9 July 2021 / Revised: 10 August 2021 / Accepted: 16 August 2021 / Published: 20 August 2021
(This article belongs to the Section Physical Sensors)

Abstract

:
The rapid improvement accuracy of the atomic frequency standard puts forward higher requirements for the measurement resolution of the atomic frequency standard comparison system. To overcome the defect that the single zero-crossing point detection is sensitive to noise in the traditional dual mixer time difference measurement method, a digital frequency measurement method is proposed. This method combines sinusoidal beat technology, multi-channel synchronous acquisition technology, and digital frequency measurement technology, and uses differential compensation of system error to realize the precision measurement of atomic frequency standard. The frequency measurement accuracy is less than 2.5 × 10−14 and the noise floor is better than 6.5 × 10−15 = 1 s. The system has a high frequency measurement accuracy and a low noise floor, which can realize the precise measurement of a highly stable frequency source.

1. Introduction

Precise atomic frequency standard has provided time-frequency reference for modern communications, deep space exploration, satellite navigation [1,2]. However, due to the factors such as the aging of components and changes in the operating environment, the atomic frequency standard has a long-term drift during operation [3,4,5]. To achieve the state monitoring and performance evaluation of the atomic frequency standard, the development of the atomic frequency standard measurement and analysis system is of great significance. Besides, with the development of science and technology, the performance of atomic frequency standards is constantly improving [6]. For example, the accuracy of a new type of atomic frequency standard can reach 1 × 10−15, the uncertainty of the optical clock can even reach 1 × 10−18/d, and the stability of an ultra-stable crystal oscillator can reach 1 × 10−14/τ = 1 s [7,8,9], which puts forward higher requirements for the measurement resolution of atomic frequency standard measurement and analysis equipment.
High-precision frequency standard comparison methods usually use phase comparison technology. Among them, the dual mixer time difference (DMTD) method is widely used for atomic frequency standard comparison, and many precision frequency measurement equipments have been developed based on this method. For example, to provide time-frequency reference laboratories with real-time monitoring and performance testing of atomic clocks, TimeTech has developed a Phase-Comp [10]. The Phase-Comp uses two time interval counters to measure the time interval between zero-crossing points through coarse and fine measurements. To meet the needs of deep space exploration, NASA Jet Propulsion Experiment (JPL) has developed a multi-channel Frequency Standard Stability Analyzer (FSSA) [11,12]. When the beat signal triggers the zero-crossing event, the FSSA uses the event counter to mark the time stamp to obtain the phase difference between the zero-crossing points. In addition, there is a Software Defined Radio (SDR) frequency standard comparator, which is a fully digital DMTD [13]. There is the TSC-5125A phase noise tester from MICROSEMI of the United States and the VCH-323 phase comparison analyzer from VREMYA-CH of Russia [14,15], which use a fully digital dual mixer time difference method to perform analog-to-digital conversion on the input signal directly. Through digital signal processing, the phase difference between the signal under test and the reference signal is measured. The performance parameters of each frequency standard comparison equipment are shown in Table 1.
To overcome the defect that the single zero-crossing point detection is sensitive to noise in the traditional dual mixer time difference measurement method, this paper proposes a digital frequency measurement method. Different from fully digital DMTD in Software Defined Radio such as TSC-5125A or VCH-323, the method performs A/D sampling on the analog sinusoidal beat signal and uses a digital signal processing method to measure the frequency of the sinusoidal beat signal. This article proposes two frequency measurement methods: the cross-correlation decimal frequency measurement method and the three-point integer frequency measurement method. The paper [16] also mentioned the same method, but it uses the acquisition card of National Instrument (NI) to sample the sinusoidal beat signal. On the one hand, the data acquisition card requires initial calibration, and the calibration results may affect the accuracy of the signal measurement results. On the other hand, the data acquisition card is difficult to expand the channel. This paper presents a hardware implementation scheme of data acquisition and analysis based on ZYNQ-7020. The scheme has a low design cost and is easy to expand the channel. This scheme also has lower design costs and better frequency comparison performance than the data acquisition card.
To study the high-resolution atomic frequency standard comparison method, this paper first analyzes the basic principle of the DMTD method and factors affecting the measurement resolution. Then the frequency measurement algorithm is introduced in detail. Further, a hardware implementation is given. Simultaneously, an experimental test platform is created, the frequency measurement accuracy and noise floor of the measurement system based on the two methods are analyzed. Ultimately, we analyze the test results, and the discussion is given.

2. The Principle of DMTD Method

The DMTD method is a high-precision phase comparison technique that can be used to evaluate signal stability [17,18,19]. As can be seen from Figure 1, the test signal v 1 ( t ) with the nominal frequency f 0 and reference signal v 2 ( t ) with the same nominal frequency shown in the following formula is connected to the input port of the DMTD measurement system for time difference measurement.
v i ( t ) = A i ( t ) sin { 2 π ( f 0 + Δ f i ( t ) ) t + θ i } , i = 1 , 2
where A i ( t ) is the amplitude of each signal. θ i is the initial phase. Δ f i ( t ) is a smooth real function, which reflects the phase noise of the signal.
In the DMTD method, a frequency offset generator (FOG) is used as shown in Figure 1. The FOG outputs the frequency offset signal v 3 ( t ) with the frequency f c , which has a frequency offset f b e a t compared to f 0 . The frequency offset signal v 3 ( t ) can be expressed by the formula as below:
v 3 ( t ) = A 3 ( t ) sin { 2 π ( f c + Δ f c ( t ) ) t + θ 3 }
The test signal and reference signal are mixed with the frequency offset signal through a double-balanced mixer. The signals output from the two double-balanced mixers are processed by a low-pass filter to generate the sinusoidal beat signals v b e a t _ i ( t ) .
v b e a t _ i = B i ( t ) sin { 2 π ( f b e a t t + [ Δ f i ( t ) Δ f c ( t ) ] t ) }
where B i ( t ) > 0 is the amplitude of each beat signal. f b e a t is a nominal frequency of the sinusoidal beat signal.
After two beat signals are detected at the zero-crossing detector, it is provided as the start signal and stop signal to the time interval counter (TIC) for time difference measurement.
Assuming that t k ,   t k respectively denotes the time when the beat signals v b e a t _ 1 ( t ) and v b e a t _ 2 ( t ) cross zero for the k-th time, then the following formulas are obtained:
k = f b e a t t k + [ Δ f 1 ( t k ) Δ f c ( t k ) ] t k
k = f b e a t t k + [ Δ f 2 ( t k ) Δ f c ( t k ) ] t k
Therefore, the time interval between two beating signals measured by TIC is Δ t k = t k t k .
Δ t k = Δ f 1 ( t k ) t k Δ f 2 ( t k ) t k f b e a t + Δ f c ( t k ) t k Δ f c ( t k ) t k f b e a t
The first term of the above formula denotes the effect of frequency fluctuation of the test signal and reference signal on the time interval between two beat signals. The second term shows that although the symmetric structure of DMTD can eliminate most of the noise from the frequency offset signal, it still contains the residual amount of the noise from the frequency offset signal [20,21]. To reduce the influence of noise as much as possible, a phase shifter is added to the reference signal channel to reduce the phase difference between the reference signal and the test signal. Then, the Δ t k can be approximately expressed as:
Δ t k Δ f 1 ( t k ) t k Δ f 2 ( t k ) t k f b e a t f 0 f b e a t Δ f 1 ( t k ) t k Δ f 2 ( t k ) t k f 0
The second term of the multiplier factor in Equation (7) denotes the time difference Δ T between the test signal and the reference signal. It also shows that the measurement resolution Δ T is enhanced by a factor of f 0 / f b e a t .
In addition to noise from the frequency offset signal, there is also random noise generated by each device in the DMTD system [22]. Since the zero-crossing point detection is sensitive to noise, it is easy to lead to misjudgment of the zero-crossing point and cause measurement errors, which limit the improvement of the measurement resolution of DMTD method.

3. Method of Digital Frequency Measurement

To reduce the influence of noise superimposed on a single zero-crossing point, a multi-channel digital frequency measurement method is proposed [16]. Compared with the DMTD method, the proposed method performs analog-to-digital conversion on the sinusoidal beat signal output by the low-pass filter instead of zero-crossing detection. Multi-point sampling (including zero-crossing point) of the beat signal can reduce the influence of noise on the measurement result more than single-point zero-crossing detection [23]. In addition, the design of multi-channel ensures that the signals of each channel can be measured synchronously, to calibrate the error of the measurement results. On the other hand, it can also meet the needs of simultaneous measurement of multiple test signals. The multi-channel digital frequency measurement system is displayed in Figure 2.
The multi-channel digital frequency measurement system consists of one calibration channel and seven measuring channels, and the number of measuring channels can be arbitrarily extended. The circuit structure of the calibration channel is consistent with each measurement channel. In the process of data acquisition, the Analog-Digital Converter of each channel should use a common sampling clock to ensure that the frequency measurement results are simultaneous, it also ensures the strong correlation of the noise of each channel.
Different from the measurement channel, the input signal of the calibration channel is also a reference input of the frequency offset generator. Due to the consistency of circuit structure and strong correlation of noise, the measurement error of the beat signal frequency of the calibration channel can represent the systematic error. Then the frequency measurement results of each measurement channel can be subtracted from that of the calibration channel to eliminate the influence of additional measurement noise.
The calculation of frequency adopts digital signal processing methods, including cross-correlation decimal frequency measurement and three-point integer frequency measurement.
The discrete sinusoidal beat signal of each channel is shown in the following formula:
v i ( n ) = V i sin ( 2 π f b e a t + Δ f i f s n + φ i ) + N i ( n )
where V i is the amplitude of the beat signal of channel i. f b e a t is the nominal frequency of the digital beat signal. Δ f i is the frequency deviation of the input signal of channel i. The frequency deviation includes the integer component Δ f i n t and fractional component Δ f d e c , i.e., Δ f i = Δ f i n t _ i + Δ f d e c _ i . f s is the sampling clock rate. φ i is the initial phase of the beat signal of channel i . n is the discrete time. φ i is the total noise of channel i.

3.1. Cross-Correlation Decimal Frequency Measurement

To calculate the decimal frequency deviation of the beat signal, the cross-correlation frequency measurement method extracts two groups of continuous sampling data with the length of f s from each channel for cross-correlation operation. It takes two seconds of sampled data to calculate Δ f d e c _ i , therefore the calculated Δ f d e c _ i is the average decimal frequency deviation within the time interval of 2 s. The signals at the j and j + 1 second of channel i are expressed as follows:
v i , j ( n ) = V i , j sin ( ω i , j n + φ i , j ) + N i , j ( n )
v i , j + 1 ( n ) = V i , j + 1 sin ( ω i , j n + φ i , j + 1 ) + N i , j + 1 ( n )
where, ω i , j = 2 π f b e a t + Δ f i , j f s , Δ f i , j denotes the average frequency deviation between the j and j+1 second of channel i. V i , j and V i , j + 1 are the amplitude of the beat signal.
The cross-correlation result of v i , j ( n ) and v i , j + 1 ( n ) can be expressed by:
R v i , j , v i , j + 1 ( τ ) = 1 f s n = 0 f s 1 v i , j ( n ) v i , j + 1 ( n + τ ) = 1 2 V i , j V i , j + 1 cos ( 2 π ω i , j τ + Φ i , j ) + R v i , j , N i , j + 1 + R v i , j + 1 , N i , j + R N i , j , N i , j + 1
where τ is the delay time between v i , j ( n ) and v i , j + 1 ( n ) . Φ i , j is the initial phase difference between v i , j ( n ) and v i , j + 1 ( n ) in the range of [ π , π ] , and Φ i , j = φ i , j + 1 φ i , j . R v i , j , N i , j + 1 and R v i , j + 1 , N i , j denote the cross-correlation value between digital sinusoidal beat signal and noise. R N i , j , N i , j + 1 denotes the cross-correlation value between noises. Statistically, the signal is not correlated with noise, i.e., R v i , j , N i , j + 1 = 0 , R v i , j + 1 , N i , j = 0 . There is also no correlation between the noises or the correlation is very small enough to be ignored, i.e., R N i , j , N i , j + 1 0 .
When τ = 0 Equation (11) can be expressed as:
R v i , j , v i , j + 1 ( 0 ) = 1 2 V i , j V i , j + 1 cos ( Φ i , j )
If v i , j ( n ) and v i , j + 1 ( n ) performs cross-correlation operation with themselves respectively, V i , j and V i , j + 1 can be expressed by:
V i , j = 2 R v i , j , v i , j ( 0 )
V i , j + 1 = 2 R v i , j + 1 , v i , j + 1 ( 0 )
According to the Formula (12)–(14), the Φ i , j can be expressed as follows:
Φ i , j = ± arccos ( R v i , j , v i , j + 1 ( 0 ) R v i , j , v i , j ( 0 ) R v i , j + 1 , v i , j + 1 ( 0 ) )
Since v i , j ( n ) and v i , j + 1 ( n ) are two groups of continuous sampling data, the initial phase v i , j + 1 ( n ) is equal to the phase of v i , j ( n ) at n = f s . As shown in the follows:
φ i , j + 1 = φ i , j + 2 π ( f b e a t + Δ f i , j f s f s )
f b e a t is an integer, and Δ f i , j = Δ f i n t _ i , j + Δ f d e c _ i , j , the initial phase difference Φ i , j can be expressed as:
Φ i , j = φ i , j + 1 φ i , j = 2 π Δ f d e c _ i , j
Combining Formulas (15) and (17), Δ f d e c _ i , j can be expressed as:
| Δ f d e c _ i , j | = 1 2 π arccos ( R v i , j , v i , j + 1 ( 0 ) R v i , j , v i , j ( 0 ) R v i , j + 1 , v i , j + 1 ( 0 ) )
The fractional part of frequency deviation can be measured by the cross-correlation decimal frequency measurement method. There is only multiplication and addition in the cross-correlation algorithm, so the method is simple and easy to implement with Field Programmable Gate Array (FPGA), which not only meets the requirements of algorithm implementation, but it also supports multi-channel parallel measurement. The numerical range of Δ f d e c _ i , j is [ 0.5 , 0.5 ] . As for the determination of the sign Δ f d e c _ i , j , a detailed description will be given later.

3.2. Three-Point Integer Frequency Measurement

Three-point integer frequency measurement can measure the integer part of beat signal frequency, including nominal frequency f b e a t and integer frequency deviation Δ f i n t . According to the trigonometric identities, the beating signal of channel i at j seconds expressed by Formula (9) has the following relationship:
v i , j ( n ) = 2 cos ( ω i , j ) v i , j ( n 1 ) v i , j ( n 2 )
where ω i , j = 2 π f b e a t + Δ f i , j f s , Δ f i , j = Δ f i n t _ i + Δ f d e c _ i denotes the average frequency deviation at j second.
Therefore, the frequency of the beat signal can be expressed as follows:
f b e a t + Δ f i , j = f s 2 π arccos ( v i , j ( n ) + v i , j ( n 2 ) 2 v i , j ( n 1 ) )
Generally, the frequency value of the digital beat signal can be calculated as long as three consecutive sampling points are obtained. However, in the process of sampling and digitizing, there will be cases where the sampling data is zero. The above methods need to be improved.
If the sampling data v i , j ( n 1 ) and v i , j ( n 2 ) are used to estimate the data v i , j ( n ) according to Formula (2), the estimated value v ^ i , j ( n ) can be expressed as:
v ^ i , j ( n ) = c v i , j ( n 1 ) v i , j ( n 2 )
where c = 2 cos ( 2 π f ^ b e a t + Δ f ^ i , j f s ) , f ^ b e a t + Δ f ^ i , j is the estimated frequency of the beat signal. Then the estimation error can be expressed as:
e i , j ( n ) = v i , j ( n ) v ^ i , j ( n ) = v i , j ( n ) c v i , j ( n 1 ) v i , j ( n 2 )
When the estimation error e i , j ( n ) is minimum, it means that f ^ b e a t + Δ f ^ i , j is the optimal estimated frequency of the beat signal. Then the following equation holds:
E ( | e i , j ( n ) | 2 ) c = 0
According to Formula (23), the estimated frequency can be calculated as follows:
c = 3 N [ v i , j ( n ) + v i , j ( n 2 ) ] v i , j ( n 1 ) 3 N v i , j ( n 1 ) v i , j ( n 1 )
f ^ b e a t + Δ f ^ i , j = f s 2 π arccos ( c 2 )
where N denotes the number of sample data participating in the operation. The integer frequency of the beat signal can be expressed as follows:
f i n t _ i , j = I N T [ f ^ b e a t + Δ f ^ i , j ]
where I N T [ ] is the rounding function.
The integral part of frequency deviation can be measured by the three-point integer frequency measurement method. The multiplication and addition operation in this method is easy to implement with FPGA. The numerical range of f i n t _ i , j is [ 0 , f s / 2 ] . Since the frequency measurement resolution is related to f s . Care should be taken to select the appropriate f s to ensure the accuracy of integer measurement.

3.3. Frequency Measurement of the Digital Beat Signal

In practice, the digital beat signal has a positive or negative decimal frequency deviation, but the cross-correlation decimal frequency measurement can only calculate the absolute value of decimal frequency deviation. The sign of the value of Δ f d e c _ i , j can be judged with the help of a three-point integer frequency measurement result, and the frequency of the digital beat signal is finally expressed below as:
case (i):
if I N T [ f ^ b e a t + Δ f ^ i , j ] < I N T [ f ^ b e a t + Δ f ^ i , j + 0.5 ] ,
then f b e a t _ i , j = f i n t _ i , j | Δ f d e c _ i , j | + 1 .
case (ii):
if I N T [ f ^ b e a t + Δ f ^ i , j ] = I N T [ f ^ b e a t + Δ f ^ i , j + 0.5 ] ,
then f b e a t _ i , j = f i n t _ i , j + | Δ f d e c _ i , j | .
According to the measured frequency value of the digital beat signal of each channel, the frequency deviation of the sinusoidal beat signal between each measurement channel and reference channel can be obtained. The result of dividing the frequency deviation of the sinusoidal beat signal by the factor f 0 f b e a t is the relative frequency deviation between the measured signal and the reference signal.

4. Hardware Implementation

A multi-channel digital frequency measurement method is proposed in Section 3. The specific hardware design of this method will be introduced in this section. We adopt a modular design to divide the hardware circuit into three parts: frequency offset generator, sinusoidal beat device, and multi-channel data acquisition and analysis module. A detailed description of each module is given below.

4.1. Frequency Offset Generator Module

As shown in Figure 3, the frequency offset generation module uses direct digital synthesis technology to generate frequency offset signals through AD9956. The reference signal with the frequency of 10 MHz passes through the ×10 frequency multiplier, and the 100 MHz sinusoidal signal generated by the frequency multiplier provides the system clock to AD9956. Then the AD9956 generates a frequency offset signal with a frequency 10 M H z f b e a t , Usually the value of f b e a t is 1 Hz, 10 Hz, or 100 Hz. The crystal filter is a bandpass filter with a center frequency of 10 MHz and a passband range of 1 kHz. The in-band fluctuation of the filter is less than 1 dB. The amplifier uses the LMH6702 operational amplifier circuit to amplify the power of the frequency offset signal and divides one signal into 8 channels for output.

4.2. Sinusoidal Beat Device Module

In the sinusoidal beat device, an SYPD-1+ phase detector with high sensitivity to phase fluctuation is selected as the double-balanced mixer, The measured signal is mixed with the frequency offset signal by SYPD-1+, and the primary amplification circuit composed of OP1177 amplifies the mixed signal. The bipolar multiple feedback active filter circuit composed of OP117 filters the mixed signal to select the low-frequency sinusoidal beat signal. The circuit structure of the sinusoidal beat device module is shown in Figure 4.

4.3. Multi-Channel Data Acquisition and Analysis Module

As shown in Figure 5, the analog-to-digital data acquisition system (DAS) is realized by AD7606, which supports 8 synchronous sampling inputs. The resolution of AD7606 is 16 bits, and the sampling rate of all channels is 200kSPS. The processor adopts Xilinx ZYNQ-7020 System-on-Chip (SoC) with a dual-core ARM Cortex-A9 processor and FPGA, which is divided into programmable logic (PL) and processor system (PS). The PL of ZYNQ-7020 drives AD7606 to realize data acquisition, and the collected data is transmitted to DDR3 SDRAM through direct memory access (DMA). In addition, the PL also performs multiplication and addition operations in the frequency measurement algorithm on the collected data. The PS of ZYNQ-7020. The PS of ZYNQ-7020 performs division and inverse cosine operation on the multiplication and addition result, and finally calculates the frequency deviation of sinusoidal beat signal between each measurement channel and reference channel.

5. Experimental Testing

To evaluate the performance of the multi-channel digital frequency measurement method proposed in this article, a multi-channel digital frequency standard comparison prototype has been developed. The prototype consists of one calibration channel and seven measurement channels. The nominal frequency of the sinusoidal beat signal is 1 Hz, and the sampling clock of ADC is 200 kHz. Figure 6 presents the structure of the multi-channel digital frequency standard comparison prototype.
We have also established an experimental test platform to evaluate the frequency measurement accuracy and noise floor of the prototype. Figure 7 presents the experimental test platform diagram. A sinusoidal signal generated by an active hydrogen maser with a nominal frequency of 10 MHz is distributed to four channels by the frequency distribution amplifier. Then four sinusoidal signals are connected to the input port of one calibration channel and three measurement channels respectively. A Personal Computer (PC) is used to send commands to the prototype and receive the data of frequency calculation results.

5.1. Measurement Accuracy

The accuracy can be used to evaluate the measurement ability of the multi-channel digital frequency standard comparison prototype, which can be characterized by relative error. The relative error can be measured by the quotient of the measurement error and the nominal value. Table 2 lists the frequency measurement average and relative error of CH1 in different time intervals. Figure 8 shows the relative error of the frequency during the continuous measurement of 10,000 s by Stable32 software. Table 2 and Figure 8 show that the relative error of frequency measurement of the prototype is less than 2.5 × 10−14, the error is very small, and the measurement accuracy is high, which can meet the needs of high-precision measurement.

5.2. Noise Floor

Noise Floor refers to the additional noise of the measuring equipment itself except the useful signal, which is usually characterized by Allan variance at 1 s in the field of time and frequency. Figure 9 shows the frequency stability of three measurement channels with systematic error elimination. For an input signal with the frequency 10 MHz, the noise floor test results of the three measurement channels of the prototype are 6.26 × 10−15/τ = 1 s, 6.19 × 10−15/τ = 1 s, and 6.10 × 10−15/τ = 1 s. Compared with Figure 10, the measurement results of the measurement channel are subtracted from the calibration channel, which eliminates the influence of noise from the device and improves the performance of frequency standard comparison. The noise floor of the prototype is low enough to be used to evaluate the ultra-stable frequency source with frequency stability of 1 × 10−14/τ = 1 s. Simultaneously, the Allan variance curves of the three channels are approximately coincident, which also shows that there is a high consistency between the channels.

6. Conclusions and Discussion

The fast-developing high-precision clock sources are an important part of the time-frequency system. The Atomic clock usually provides time-frequency references for timekeeping systems in the form of clock groups, and it is of great practical significance to evaluate their frequency stability. In this study, a digital frequency measurement method is proposed. The performance of this method has been rigorously assessed.
In the prototype test, the relative error can reach 10−14 Hz. This is due to the multi-channel synchronous sampling of the prototype, which ensures that the system error of each channel has a strong correlation. Therefore, the system error of the calibration channel, whose structure is consistent with the measurement channel, can be used to compensate for the measurement results from the measurement channel.
The noise floor of the prototype is equivalent to that of existing products, and it can also support multi-channel simultaneous testing. Since the prototype only supports the signal with a nominal frequency of 10 MHz, it is necessary to increase the frequency measurement bandwidth of the system to expand the application range in subsequent work.

Author Contributions

Conceptualization, B.X. and Y.L.; methodology, B.X. and Y.L.; software, Z.D.; validation, Y.L. and X.L.; formal analysis, Y.L.; investigation, B.X.; data curation, B.X.; writing—original draft preparation, B.X.; review and editing, Y.L. and X.L.; supervision, Y.L. and X.L.; project administration, Y.X.; funding acquisition, Y.L. All authors have read and agreed to the published version of the manuscript.

Funding

This work is funded by the National Natural Science Funds of China (Grant No: U2031125).

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Not applicable.

Acknowledgments

Many thanks are due to the Time-frequency Reference Laboratory of the National Time Service Center (NTSC) for providing the test signal.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Prestage, J.D.; Weaver, G.L. Atomic Clocks and Oscillators for Deep-Space Navigation and Radio Science. Proc. IEEE 2007, 95, 2235–2247. [Google Scholar] [CrossRef]
  2. Stein, S.R. The design of atomic frequency standards and their performance in specific applications. Prog. Quantum Electron. 1984, 8, 129–137. [Google Scholar] [CrossRef]
  3. Parker, T.E. Environmental factors and hydrogen maser frequency stability. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 1999, 46, 745–751. [Google Scholar] [CrossRef] [PubMed]
  4. Shuzhou, L.I.; Wang, M.; Xiao, S. Frequency Stability of Hydrogen Maser Affected by Environment Temperature. Mod. Navig. 2017, 118–121. [Google Scholar]
  5. Zhai, Z.C. Systematic effects analysis of affecting hydrogen maser long-term frequency stability. J. Astronaut. Metrol. Meas. 2008, 28, 9–15. [Google Scholar]
  6. Zhou, W.; Bai, L.; Su, X.; Yu, J.; Ma, J.; Yu, G. The high stability of device resolution based precise phase difference measurement. In Proceedings of the 2014 IEEE International Frequency Control Symposium (FCS), Taipei, Taiwan, 19–22 May 2014. [Google Scholar]
  7. Hong, F.L. Optical frequency standards for time and length applications. Meas. Sci. Technol. 2016, 28, 012002. [Google Scholar] [CrossRef]
  8. Cermak, J.; Kuna, A.; Sojdr, L.; Salzenstein, P. Short-Term Frequency Stability Measurement of BVA Oscillators. In Proceedings of the 2007 IEEE International Frequency Control Symposium Joint with the 21st European Frequency and Time Forum, Geneva, Switzerland, 29 May–1 June 2007. [Google Scholar]
  9. Salzenstein, P.; Kuna, A.; Sojdr, L.; Sthal, F.; Cholley, N.; Lefebvre, F. Frequency stability measurements of ultrastable BVA resonators and oscillators. Electron. Lett. 2010, 46, 686–688. [Google Scholar] [CrossRef]
  10. TimeTech GmbH. Phase Comparator 6 Channels, 100 MHz with 5/10 MHz Option; TimeTech GmbH Corporation: Stuttgart, Germany, 2006. [Google Scholar]
  11. Greenhall, C.A.; Kirk, A.; Stevens, G.L. A multichannel dual-mixer stability analayzer: Progress report. In Proceedings of the 33th Annual Precise Time and Time Interval Systems and Applications Meeting, Long Beach, CA, USA, 27–29 November 2001. [Google Scholar]
  12. Greenhall, C.A.; Kirk, A.; Tjoelker, R.L. Frequency standards stability analyzer for the deep space network. Interplanet. Netw. Prog. Rep. 2007, 42–169. Available online: https://ipnpr.jpl.nasa.gov/progress_report/42-169/169B.pdf (accessed on 15 August 2021).
  13. Sherman, J.A.; Jrdens, R. Oscillator metrology with software defined radio. Rev. Sci. Instrum. 2016, 87, 470–475. [Google Scholar] [CrossRef] [PubMed]
  14. Microsemi. High-Performance Extend-Range Phase Noise and Allan Deviation Test Set; Microsemi Corporation: Culver City, CA, USA.
  15. VREMYA-CH. PHASE COMPARATOR-ANALYZER VCH-323 Operational Manual; VREMYA-CH Joint Stock Company: Nizhniy Novgorod, Russia, 2018. [Google Scholar]
  16. Liu, Y.; Li, X.H.; Wang, Y.L. Multi-channel frequency measurement system based on digital signal processing. Chin. J. Sci. Instrum. 2009, 30, 1963–1968. [Google Scholar]
  17. Nakagawa, F.; Imae, M.; Hanado, Y.; Aida, M. Development of Multi Channel Dual Mixier Time Difference System. In Proceedings of the 2004 Conference on Precision Electromagnetic Measurements, London, UK, 27 June–2 July 2004. [Google Scholar]
  18. Mochizuki, K.; Uchino, M.; Morikawa, T. Frequency-stability measurement system using high-speed adcs and digital signal processing. IEEE Trans. Instrum. Meas. 2007, 56, 1887–1893. [Google Scholar] [CrossRef]
  19. Uchino, M.; Mochizuki, K. Frequency stability measuring technique using digital signal processing. Electron. Commun. Jpn. Part I Commun. 2004, 87, 21–33. [Google Scholar] [CrossRef]
  20. Greenhall, C.A. Common-Source Phase Error of a Dual-Mixer Stability Analyzer. Interplanet. Netw. Prog. Rep. 2000, 42, 42–143. [Google Scholar]
  21. Sze-Ming, L. Influence of noise of common oscillator in dual-mixer time-difference measurement system. Instrum. Meas. 1986, 4, 648–651. [Google Scholar] [CrossRef]
  22. Sojdr, L.; Cermak, J.; Barillet, R. Optimization of Dual-Mixer Time-Difference Multiplier; Frequency & Time Forum, Eftf European IET: London, UK, 2004. [Google Scholar]
  23. Liu, Y.; Li, X.H.; Wang, D.N.; Wang, W.L.; Li, J.X. Analysis and comparison of performance of frequency standard measurement systems based on beat-frequency method. In Proceedings of the IEEE International Frequency Control Symposium, Honolulu, HI, USA, 19–21 May 2008. [Google Scholar]
Figure 1. A schematic diagram of the DMTD method.
Figure 1. A schematic diagram of the DMTD method.
Sensors 21 05626 g001
Figure 2. A schematic diagram of the multi-channel digital frequency measurement system.
Figure 2. A schematic diagram of the multi-channel digital frequency measurement system.
Sensors 21 05626 g002
Figure 3. A block diagram of the frequency offset generator.
Figure 3. A block diagram of the frequency offset generator.
Sensors 21 05626 g003
Figure 4. The circuit structure of the sinusoidal beat device module.
Figure 4. The circuit structure of the sinusoidal beat device module.
Sensors 21 05626 g004
Figure 5. A block diagram of multi-channel data acquisition and analysis module.
Figure 5. A block diagram of multi-channel data acquisition and analysis module.
Sensors 21 05626 g005
Figure 6. The structure of the multi-channel digital frequency standard comparison.
Figure 6. The structure of the multi-channel digital frequency standard comparison.
Sensors 21 05626 g006
Figure 7. A diagram of the experimental test platform.
Figure 7. A diagram of the experimental test platform.
Sensors 21 05626 g007
Figure 8. The frequency relative error during the continuous measurement of 10,000 s.
Figure 8. The frequency relative error during the continuous measurement of 10,000 s.
Sensors 21 05626 g008
Figure 9. The frequency stability of three measurement channels with systematic error elimination.
Figure 9. The frequency stability of three measurement channels with systematic error elimination.
Sensors 21 05626 g009
Figure 10. The frequency stability of three measurement channels without systematic error elimination.
Figure 10. The frequency stability of three measurement channels without systematic error elimination.
Sensors 21 05626 g010
Table 1. The performance parameters of each frequency standard comparison equipment.
Table 1. The performance parameters of each frequency standard comparison equipment.
EquipmentNumber of ChannelsNoise Floor (ADEV τ = 1 s)
Phase-Comp6~162.5 × 10−14(@10 MHz)
FSSA82.0 × 10−15(@100 MHz)
TSC-5125A23.0 × 10−15(@10 MHz)
VCH-32337.0 × 10−15
Table 2. The requency measurement average and relative error.
Table 2. The requency measurement average and relative error.
Time (s)Frequency (Hz)Relative Error
19,999,999.999999756−2.44 × 10−14
39,999,999.999999785−2.15 × 10−14
109,999,999.999999798−2.02 × 10−14
309,999,999.999999803−1.97 × 10−14
1009,999,999.999999809−1.91 × 10−14
3009,999,999.999999803−1.97 × 10−14
10009,999,999.999999806−1.94 × 10−14
30009,999,999.999999805−1.95 × 10−14
10,0009,999,999.999999804−1.96 × 10−14
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Xiao, B.; Liu, Y.; Li, X.; Deng, Z.; Xue, Y. High-Resolution Multi-Channel Frequency Standard Comparator Using Digital Frequency Measurement. Sensors 2021, 21, 5626. https://0-doi-org.brum.beds.ac.uk/10.3390/s21165626

AMA Style

Xiao B, Liu Y, Li X, Deng Z, Xue Y. High-Resolution Multi-Channel Frequency Standard Comparator Using Digital Frequency Measurement. Sensors. 2021; 21(16):5626. https://0-doi-org.brum.beds.ac.uk/10.3390/s21165626

Chicago/Turabian Style

Xiao, Bo, Ya Liu, Xiaohui Li, Zhifeng Deng, and Yanrong Xue. 2021. "High-Resolution Multi-Channel Frequency Standard Comparator Using Digital Frequency Measurement" Sensors 21, no. 16: 5626. https://0-doi-org.brum.beds.ac.uk/10.3390/s21165626

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop